BEAMEX POC8 Bedienungsanleitung

Beamex®POC8
AUTOMATIC PRESSURE CONTROLLER
User Manual
Dear user,
We have made every effort to ensure the accuracy of the contents of this manual.
Should any errors be detected, we would greatly appreciate to receive suggestions
to improve the quality of the contents of this manual.
The above notwithstanding, we can assume no responsibility for any errors in this
manual or their eventual consequences.
We reserve rights to make modifications to this manual without any further notice.
For more detailed technical data about the Beamex® POC8 Automatic Pressure
Controller, please contact the manufacturer.
© 2017
BEAMEX OY AB
Ristisuonraitti 10
FIN-68600 Pietarsaari
FINLAND
Tel +358 - 10 - 5505000
Fax +358 - 10 - 5505404
Internet: https://www.beamex.com
8888500 / POC8uEng / Version 1.0a

This Warning symbol indicates that danger of injury for persons and the
environment and/or considerable damage (mortal danger, danger of injury)
will occur if the respective safety precautions are not taken.
This Caution symbol indicates danger for the system and material if the re-
spective safety precautions are not taken.
This Notice symbol does not indicate safety notices but information for a
better understanding of the facts.

Contents
Beamex® POC8 Automatic Pressure Controller
Contents
General Information................................................................1
FCC Radio Frequency Emission Notice .................................... 1
Trademarks................................................................................ 1
Packaging for Shipment............................................................. 1
Safety Notices.........................................................................2
User Responsibilities ................................................................. 2
General Safety Notices.............................................................. 2
Warnings and Caution Notices .................................................. 3
General Description................................................................4
Features..................................................................................... 4
Turning On................................................................................. 5
Front Panel ................................................................................ 6
Power Switch ..................................................................... 6
USB Port............................................................................ 6
Display ....................................................................................... 7
Chassis Assembly...................................................................... 8
Specifications .........................................................................9
Measure Specification................................................................ 9
Control Specifications................................................................ 9
General Specifications............................................................. 10

Contents
Beamex® POC8 Automatic Pressure Controller
Installation ............................................................................ 11
Unpacking the Instrument........................................................ 11
Dimensions in mm / inches...................................................... 12
Mounting.................................................................................. 13
Rear Panel............................................................................... 13
Pressure Connections...................................................... 13
Supply Port ...................................................................... 14
Exhaust Port .................................................................... 14
Vent Port.......................................................................... 14
Measure/Control Port....................................................... 14
Reference Port................................................................. 14
Barometric Reference Port .............................................. 14
Remote Communication Connections ............................. 14
Power Up ......................................................................... 14
Local Operation and Setup .................................................. 15
General Operation ................................................................... 15
Setup ............................................................................. 15
Display Screen Features.................................................. 15
Initial Setup.............................................................................. 16
Contact and Version Information View............................. 16
Language Selection......................................................... 16
Settings View's' Parameter Inputs ........................................... 17
Main Functions......................................................................... 18
Home View....................................................................... 18
Settings View ................................................................... 26
Control Settings ............................................................... 30
Display Settings ............................................................... 35
Programs ......................................................................... 37
Favorites View ................................................................. 38
Information View .............................................................. 39
Troubleshooting View ...................................................... 40
Remote Settings View...................................................... 41
Step Settings View........................................................... 43
Leak Test ......................................................................... 45
Burst Test......................................................................... 46
Service View .................................................................... 47
Unlocked Service View.................................................... 48

Contents
Beamex® POC8 Automatic Pressure Controller
Remote Operation.................................................................49
Command Set.......................................................................... 49
Operation with Beamex calibrators.................................. 49
IEEE-488.................................................................................. 50
IEEE-488.2 Commands ................................................... 50
Ethernet ................................................................................... 51
USB device (Virtual Serial Port)............................................... 51
Serial........................................................................................ 52
Serial Cable Requirements.............................................. 52
Command and Query Format.................................................. 53
Command Set Definitions........................................................ 53
Output Formats........................................................................ 54
Commands and Queries.......................................................... 55
Beamex default command set ......................................... 55
Units Command Syntax for Measurement Units.............. 62
Error Codes...................................................................... 63
SCPI Commands and Queries......................................... 63
GPIB Capability Codes .................................................... 65
GPIB Capability Codes .................................................... 65
Interface Functions .......................................................... 66
USB Software Upgrade............................................................ 67
Options..................................................................................68
Barometric Module................................................................... 69
Gauge Pressure Emulation.............................................. 69
Absolute Pressure Emulation........................................... 69
Emulation Mode Accuracy ............................................... 69
Calibration of the Barometric Module............................... 69
Specifications of the Barometric Module.......................... 69
Additional measurement modules ........................................... 70
Secondary measurement module installation.................. 70
Bench Mounting Kit.................................................................. 70
Fittings ..................................................................................... 70
Re-calibrating........................................................................... 71
Dirt and Moisture Trap............................................................. 71

Contents
Beamex® POC8 Automatic Pressure Controller
Maintenance.......................................................................... 72
Removing the measurement module....................................... 72
Troubleshooting....................................................................... 73
Calibration............................................................................. 74
Environment............................................................................. 74
Pressure Standards................................................................. 74
Media....................................................................................... 74
Service View (unlocked) .......................................................... 75
Calibration Data....................................................................... 76
One Point Calibration............................................................... 76
Two Point Calibration............................................................... 77
Linearization............................................................................. 78
Head Pressure......................................................................... 80
Appendix............................................................................... 81
Measurement Units.................................................................. 81
Conversion Factors, PSI.......................................................... 82
Conversion Factors, millitorr.................................................... 83
Conversion Factors, Pascal..................................................... 84

Feedback
Beamex® POC8 Automatic Pressure Controller
Feedback
We want to improve our products and services constantly. Therefore we’d
like to know Your opinion of the product You use. Please spend a moment of
Your valuable time in filling this form. All respondents will receive a surprise
gift in return.
Certain questions can be answered immediately after receiving the product.
Others require some use of the product before You are able to answer them.
The best way to fill the form is to answer the items as it applies, and send the
form to us when all items are answered. There are however no definite re-
strictions; fill in the form when you feel like it (all items need not be an-
swered). Then send it to Beamex using one of the possibilities listed below.
Mail: Beamex Oy, Ab
Quality Feedback
P.O. Box 5
68601 Pietarsaari
FINLAND
Fax +358 - 10 - 5505404
Only the next page need to be faxed to us.
Internet: https://www.beamex.com
A similar form is available as a web page
E-mail: support@beamex.com
Refer to the numbered items on the next
page in Your e-mail message.

Feedback
Beamex® POC8 Automatic Pressure Controller
1. Name of the product you give feedback of: _____________
2. Serial number and software version number (if applicable) _____________
_____________
3. Any comments when receiving the product. Did the package contain all required items and
was it as expected?
___________________________________________________________
___________________________________________________________
4. For how long have you been using the product? _____________
5. How helpful was the manual in using the product?
(Tick a box in the percentage scale below)
6. How well did the product suit your needs?
7. How satisfied are you with the product?
8. Did anything in the product exceed your expectations? In that case, what was it?
___________________________________________________________
___________________________________________________________
9. Did anything in the product disappoint you? In that case, please specify.
___________________________________________________________
___________________________________________________________
10. Any ideas You want to propose to Beamex so that we can improve our products, operations
and/or services.
___________________________________________________________
___________________________________________________________
Title & Name: _________________________
_________________________
Address: _____________________________
_____________________________
_____________________________
_____________________________
Please contact me concerning the
Feedback I have given.
I want to receive more information on
Beamex products.

General Information 1
Beamex® POC8 Automatic Pressure Controller
General Information
FCC Radio Frequency Emission Notice
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protec-
tion against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful in-
terference in which case the user will be required to correct the interference at his or her own
expense.
Use shielded cables to connect external devices to this instrument to minimize RF radiation.
Trademarks
Beamex is a registered trade mark of Beamex Oy Ab.
Packaging for Shipment
If the product must be shipped to a different location or returned to Beamex for any reason
through a common carrier it must be packaged properly to minimize the risk of damage.
The recommended method of packing is to place the instrument in a container, surrounded on
all sides with at least 100 mm / four inches of shock attenuation material such as styrofoam
peanuts.

2 Safety Notices
Beamex® POC8 Automatic Pressure Controller
Safety Notices
User Responsibilities
To ensure safety, the user must make sure that:
The system is used properly, no dangerous media are used and that all technical specifi-
cations are observed.
The system is operated in perfect operating condition.
This operation manual is legible and accessible to the user at the system’s location.
The system is operated, serviced and repaired only by authorized and qualified person-
nel.
The operator receives instruction on industrial safety and environmental protection, and
is knowledgeable of the operating instructions and the safety notices contained therein.
General Safety Notices
The system should only be operated by trained personnel who are familiar
with this manual and the operation of the instrument.
A condition for trouble-free and safe operation of this system is proper
transport, proper storage, installation, assembly and proper use as well as
careful operation and maintenance.
Any operation not described in the following instructions should be prohib-
ited. The system must be handled with care required for an electronic preci-
sion instrument (protect from humidity, impacts, strong magnetic fields,
static electricity and extreme temperatures). Do not insert any objects into
the instrument.
The system is powered via the power cable with a voltage that can cause
physical injury. Even after disconnecting the system from the power supply,
dangerous volt- ages can temporarily occur due to capacitance.
Extreme care must be taken with pressure connections when using hazard-
ous or toxic media.
Repairs must only be performed by authorized service personnel.
Additional safety notices are found throughout this manual.
Inhaltsverzeichnis
Andere BEAMEX Controller Handbücher