BioLAB BTST-102 Bedienungsanleitung

Operation Manual
Stability Test Chamber
BTST-102
Thank you very much for Choosing Biolab products. Please read the“Operating Instructions” and “Warranty”
before operating this unit to assure proper operation.

www.biolabscientific.com
Stability Test Chamber BTST-102
2
Index
1. Safety Tips..................................................................................................................................................05
2. Operation Precaution............................................................................................................................06
3. Structure.....................................................................................................................................................08
4. Control Panel............................................................................................................................................10
5. Introduction..............................................................................................................................................12
6. Parameters.................................................................................................................................................13
7. Failure Reason and Countermeasures.............................................................................................19
8. Warranty Commitment.........................................................................................................................27

Stability Test Chamber BTST-102
www.biolabscientific.com 3
The Company has made the thorough research and analysis on possible dangers in the use and
operation of the product. In spite of this, it is still very hard to predict all dangers. Therefore, the
Technical Instruction does not necessarily include all dangers and precautions. However, operations
in accordance with the methods and requirements stated in the Instruction can enhance the safety
of operation to the maximum. In the process of using the product, please pay due attention to the
precautions so as to avoid accidents or product failures.
Biolab is not liable for any damage of the instrument caused by the fact that users do not comply with
the “Environment for the Use of the Instrument” described in the Technical Instruction.
Biolab is entitled to modify the Technical Instruction at any time, without making any prior notice.
Users must comply with important safety stipulations stated in the Technical Instruction.
Proceedings and stipulations stated thereafter can make you correctly use the instrument. Please
follow the precaution signs and tips so as to prevent users and other people from being injured.
Explanation for warning signs:
Attention: please carefully read the Technical Instruction before using the instrument
Warning: please do not open the door in the status of high-temperature and high-humidity
operation
Attention: please do not touch the recycling fans with hands when they are rotating.

www.biolabscientific.com
Stability Test Chamber BTST-102
4
Signs on the instrument
There are some warnings and precautions on the instrument, shown as follows:
Warning Attention
Please do not open the door in high-
temperature and high-humidity
operation. If the door is opened in the
operation, high-temperature steam
will blow out from the incubator, thus
burning operators, although internal
fans do stop.
In addition, when the instrument
breaks down, please rst turn it
o (stop) and open the door after
conrming the temperature reduction.
Easily hurt
Please do not touch cooling fans with bare
hands, which might cut your ngers.
Warning Warning
If the instrument isn’t used for long
time, the water in it should be drained
completely. (When the instrument
has been used for long time, there will
appear dirt such as water stain, stone,
and algae, thus aecting the function
of water level sensor, and further
leading to problems such as water
leakage.)
Before running, make sure that the internal
water tank is lled with pure water (tap
water is strictly prohibited). In order to
ensure normal operation of equipment, it
should be cleaned and maintained in use of
two months.
Danger
Protective earthing sign
Explosion danger
Please do not place any tinder such as
organic solvent into it.

Stability Test Chamber BTST-102
www.biolabscientific.com 5
01Safety tips
Please do not use the instrument outdoors.
The instrument should be installed on the solid ground. If the ground is not solid, or the installation
site is unsuitable, the instrument is likely to overturn, thus hurting operators and other people.
Please do not install the instrument in humid areas or areas close to water. Otherwise, the insulation
performance will decline, thus leading to accident of electricity leakage or electric shock.
Please use power sockets connected with earth wire in case of electric shock. If power sockets are not
connected with earth wire, please ask qualied technicians to connect them with earth wire.
The instrument can not be earthed through gas supply pipe, water supply pipe, telephone line, or
lightning arrester, which is likely to lead to electric shock at their openings.
Please do not place volatile or ammable articles in the instrument. Otherwise, explosion or re may
occur.
Please do not insert metal articles such as iron nail or wire into any hole or clearance of the instrument,
or into air outlet of internal air recycling ring. Otherwise, electric shock or injury may happen due to
the blocking and accidental contact of the above articles.
The working site of the instrument should have the good ventilation. If the ventilation is restricted,
other measures should be adopted to guarantee the safe working environment.
When placing harmful, toxic, and radioactive articles, please use the instrument within the safe area.
Improper use will do harm to human health and the environment.
Before repairing or maintaining the instrument, please cut o the power supply to avoid electric
shock or other injuries.
Without permission, users can not disassemble, repair, or ret the equipment on their own. Otherwise,
re or personnel injury may take place due to improper operation.
If the instrument does not function well, please turn o the instrument and pull down the power
source. Operation under abnormal situation may lead to electric shock or re.
Please do not damage power sockets or power wires. Please do not touch any electric component
and switch when your hands are wet, which is likely to result in electric shock.
Please do not place vessels or articles containing water on the instrument. If the articles fall down,
they will hurt operators, and water contained will give rise to electricity leakage or electric shock.
When moved, the instrument can not overturn, in order to prevent damage of the equipment or hurt
of operators.
When the instrument is not used for long time, power pins should be pulled out. Deterioration of
insulators will lead to electric shock, electricity leakage, or re.

www.biolabscientific.com
Stability Test Chamber BTST-102
6
02Operating Precautions
1. The instrument should be installed on level and solid ground.
2. The instrument should not be installed in areas exposed to direct sunshine or heat source.
3. Keep smooth air ow indoor
4. Before the start of operation, please make sure that the water tank is lled with puried water (tap
water is forbidden).
5. Clean and maintain the instrument every three months to ensure the long-term operation.
6. If the instrument isn’t used for long time, the water in it should be drained completely. (When the
instrument has been used for long time, there will appear dirt such as water stain, stone, and algae,
thus aecting the function of water level sensor, and further leading to problems such as water
leakage.)
Environmental Conditions:
1. Indoor use
2. There are no ammable solid, liquid, and gas nearby.
3. The temperature of the working environment is 5°C-30°C.
4. Power transient voltage is Grade II.
5. Power voltage uctuation does not exceed ±10% of the rated voltage.
6. Pollution grade is Grade II.
7. Level and solid ground.
8. Areas without dew.
9. Areas not exposed to direct sunshine.
10. Areas with good ventilation.

Stability Test Chamber BTST-102
www.biolabscientific.com 7
Simplied operating instruction for BTST-102, BTST-103, BTST-104 constant temperature &
humidity
Delivery setting: °C——37°C %——75%
Please comply with the following procedures:
1. Please switch on electricity leakage protector.
2. Set the over-temperature protector. The temperature should be 20°C higher than the set
temperature.
3. Connect external water tank with the water inlet of the instrument and turn on the water faucet
(the tank cover should be opened).
4. Press the power button on the operating manual, and enter the display interface ve seconds later.
The panel will display the water level is too low, and the buzzer will buzz. The buzzer can stop buzzing
after users press any key, but the alarm indicator will not go o. When the water reaches a given level,
the indicator will go o.
5. Press MODE key→ enter program selection.
Program I: ordinary mode
Program II: delay and timing mode
Program III—Program V: program mode (ten steps for each program running in turn)
Program VI: combined mode
6. Press up/down keys→ select program
7. Press ENTER key→ conrm program
8. Press left/right keys to move the cursor and press up/down keys to adjust the value.
9. Run the menu and select YES/NO
(Set 1 for YES, and run the program; Set 0 for NO, and not run the program)
10. Press ENTER key to conrm after the setting, save the setting, and press MODE key to return to
the main interface.
11. Press RUN/STOP key → instrument on/o.
12. Turn o the instrument after the completion of the test, and exhaust all the water in the instrument.

www.biolabscientific.com
Stability Test Chamber BTST-102
8
03Structure

Stability Test Chamber BTST-102
www.biolabscientific.com 9
1. Operating panel: setting and display value.
2. Leakage protector: the main switch will cut o the power in the case of over-current or short circuit.
3. RS232 interface: used for connecting computer test and printer.
4. Over-temperature protector: The temperature should be 20°C higher than the set temperature.
When the temperature exceeds the set value of the over-temperature protector, it will cut o the
power automatically.
5. Water outlet: internal water outlet of the instrument.
6. Water inlet: internal water inlet of the instrument.
7. Test hole: be for the convenience of user observation, and congured with silica gel plug internally.
8.Tube.
9.Glass door handle.
10. Internal glass
11. Shelf support: the support should be taken down after the shelf is removed.
12. Side support: supports at the two sides can be taken down.
13. Shelf: be adjusted and removed easily.
14.Gate lock.
15. Door switch: when the door is opened, the fan stops operation, and all programs stop working.

www.biolabscientific.com
Stability Test Chamber BTST-102
10
04Control panel
1. Liquid display area
The area is used for the display of temperature and humidity measuring values, and also the
corresponding setting process;
When “ALARM” indicator lights up, the alarm content will be displayed in the area in scroll.
2. “TANKWATERLEVEL” indicator
Display the water level of internal water tank in terms of four sections.
3. “REFRIGERATOR” indicator
The indicator lights up, which means that the system is refrigerating so as to reach the set value of
the temperature.
4. “NORMAL MODE” indicator
The indicator lights up, which means that the system enters the normal mode.
5. “TEMP HEATER” indicator
The indicator lights up, which means that the system is heating.
6. “HUMI HEATER” indicator
The indicator lights up, which means that the system is humidifying.
7. “PROGRAM MODE” indicator
Dieses Handbuch passt für folgende Modelle
2
Inhaltsverzeichnis
Beliebte Prüfgeräte Handbücher anderer Marken

SMART
SMART KANAAD SBT XTREME 3G Series Bedienungsanleitung

Agilent Technologies
Agilent Technologies BERT Serial Bedienungsanleitung

Agilent Technologies
Agilent Technologies N3280A Bedienungsanleitung

Vernier
Vernier Go Direct Voltage Bedienungsanleitung

Lifeloc
Lifeloc R.A.D.A.R. Bedienungsanleitung

Fluke
Fluke T5-600 Bedienungs- und Montageanleitung

PCB Piezotronics
PCB Piezotronics 8159-0112A Bedienungsanleitung

BW Technologies
BW Technologies MicroDock II Bedienungsanleitung

Sun Nuclear
Sun Nuclear 1027 Bedienungsanleitung

Biomark
Biomark HPR LITE READER Bedienungsanleitung

Hioki
Hioki 3169-20 Bedienungsanleitung

Ashcroft
Ashcroft ATE-100 Bedienungsanleitung