NeuroPlexTM E
PN 10906, 10907, 10908, & 10909
Digital Data Cable: PN 11058 & 11404
Instructions for Use

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 2
Table of Contents
What This Manual Covers ..................................................... 4
Intended Use and Indications for Use ................................... 4
Warnings and Precautions .................................................... 5
Warnings ........................................................................................ 5
Precautions ..................................................................................... 6
Symbols ................................................................................ 7
Specifications ...................................................................... 10
Hardware ............................................................................. 11
Blackrock Microsystems NeuroPlex E Setup ................................ 11
Blackrock Microsystems NeuroPlex E Pin-out .............................. 12
Instructions for Assembly .............................................................. 13
Description of Switches on the NeuroPlex E ................................. 15
Impedance and Recording Selector Switch ................................ 15
Frequency Range Selector: Wide & Norm .................................. 15
Reference and Ground Selector ................................................. 15
R1 and R2 Reference Selector ................................................... 15
Cleaning, Disinfection, Sterilization, Maintenance, and
Disposal .............................................................................. 16
Sterilization ................................................................................... 16
Cleaning, Disinfection, and Maintenance ...................................... 16
Filament Film Disassembly and Replacement .............................. 17
Disposal ........................................................................................ 18
Magnetic Resonance .......................................................... 18
Troubleshooting .................................................................. 19
Non-Green Indicator Light on NeuroPlex E ................................... 19
Poor Signal to Noise Ratio / Noisy Signal ..................................... 19

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 3
Non-Green Light on Digital Hub Bank ........................................... 19
Return Merchandise Authorization ...................................... 20
Warranty .............................................................................. 21
Support ................................................................................ 21
Complaints .................................................................................... 21

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 4
What This Manual Covers
Blackrock Microsystems’ NeuroPlexTM E provides a compact interface from a
NeuroPortTM Electrode to Blackrock’s NeuroPort Bio-Potential Signal Processing
Systems for up to 128 channels on the E128 or 96 channels on the E96. The NeuroPlex
E is designed for high fidelity transmission and recording of extracellular spikes and local
field potentials from the brain. Blackrock Microsystems’ NeuroPlex E converts analog
signals to digital format at the recording site which can reduce noise introduced to the
signal during transmission.
This manual covers the intended use, warnings, precautions, technical specifications,
pinout, cleaning, disinfection, sterilization, and maintenance instructions, and warranty of
the NeuroPlex E digital headstage manufactured by Blackrock Microsystems.
Intended Use and Indications for Use
The NeuroPlex E supports recording, processing, and display of biopotential signals
from user-supplied electrodes. Biopotential signals include, but are not limited to:
Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG),
electrocardiography (ECG), electrooculography (EOG), and Evoked Potential (EP).

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 5
Warnings and Precautions
The NeuroPlex E, as part of the NeuroPort BioPotential Signal Processing System, has
the following warnings and precautions:
Warnings
Only connect NeuroPort System components to properly tested,
grounded and dedicated AC outlets to reduce the risk of electrical shock.
Do not use an adapter for ungrounded wall outlets.
Do not connect the NeuroPort System to an outlet controlled by a wall
switch.
Do not use the NeuroPort System in the presence of flammable
anesthetic agents.
Avoid strong static discharges from sources like televisions or computer
monitors because it can damage the electrical parts of the system.
Keep the NeuroPort System away from liquids. Contact with water,
shower spray, or wet surfaces can lead to the patient receiving an
electrical shock.
Connection of external instruments may compromise electrical safety
compliance with IEC 60601-1.
The Digital Data Cable is long and may snag on equipment in the room.
Secure equipment in place and avoid placing in walkways or other traffic
areas.
The NeuroPlex E is attached to the NeuroPort Electrode percutaneous
connector, force applied to the NeuroPlex E will be applied to this
connector. Avoid large impacts to the device when it is attached to the
patient. Consequences can include tearing of bone screws from skull.
Do not use the NeuroPlex E (PN 10907, 10909) or Digital Data Cable (PN
11058) if the sterile barrier packaging is damaged.
Only sterile NeuroPlex Es and Digital Data Cables should be used in the
intraoperative environment. Non-sterile NeuroPlex Es and Digital Data
Cables should only be used in the postoperative environment.
Do not allow the Digital Data Cable to be pulled in such a way that it can
apply force to the NeuroPlex E.
Avoid spilling fluid onto or into the NeuroPlex E as doing so may cause
electrical damage to the device and loss of function or electrical shock.

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 6
Precautions
Read this entire manual prior to using the device.
Do not allow any object to impact the NeuroPlex E while in use.
Do not drop the NeuroPlex E when handling.
Do not use the NeuroPlex E with damaged Digital Data Cables or Digital
Data Cables that are not supplied by Blackrock Microsystems. Doing so
may damage the device.
When replacing the filament film of the device, avoid touching circuit
board components as electrostatic discharge can cause damage to the
device.
When replacing the filament film, do not overtighten the screws.
Overtightening the screws may cause poor connection between the
device and the pedestal.
Do not attempt to use the NeuroPlex E with connectors other than the
NeuroPort pedestal connector.
Do not plug the Digital Data Cable into devices other than the Digital Hub
and Blackrock Microsystems approved headstage accessories.
Avoid overtightening the NeuroPlex E when attaching to the NeuroPort
Electrode. Device damage may result.
Using the ‘Wide-Band’ filter mode on the device may cause a reduced
signal to noise ratio.
Swollen or degraded filament film may result in loss of signals.
Do not immerse devices in liquid.
Do not use a washing machine or an automated washing process (such
as an automated washer/disinfector). Doing so may damage the devices.
Do not allow contaminated devices to dry. All subsequent cleaning and
steps are facilitated by not allowing blood, body fluids, bone and tissue
debris, saline, or disinfectants to dry on used devices.

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 7
Symbols
The following symbols may apply to the NeuroPlex E even if they do not appear on the
physical product.
ISO 15223-1:2016 Medical Devices – Symbols to Be Used with Medical Device Labels,
Labeling, and Information to Be Supplied
Reference Symbol Title Meaning
5.1.1 Manufacturer Indicates the medical device
manufacturer.
5.1.4 Use-by Date Indicates the date after which the
medical device is not to be used.
5.1.5
Batch Code Indicates the manufacturer’s batch
code so that the batch or lot can be
identified.
5.1.6 Catalog
Number
Indicates the manufacturer’s catalog
number so that the device may be
identified. For Blackrock Microsystems
it is called the Part Number (PN).
5.1.7 Serial Number Indicates the manufacturer’s serial
number so that a specific medical
device can be identified.
5.2.3
Sterilized Using
Ethylene Oxide
Indicates that the device has been
sterilized using ethylene oxide.
5.2.6 Do Not
Resterilize
Indicates a medical device that is not
to be resterilized.

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 8
ISO 15223-1:2016 Medical Devices – Symbols to Be Used with Medical Device Labels,
Labeling, and Information to Be Supplied
Reference Symbol Title Meaning
5.2.7 Non-Sterile To indicate that the device that is
normally provided sterile in the same
or similar packaging has not been
sterilized.
5.2.8 Do Not Use if
Package is
Damaged
Indicates that a medical device should
not be used if the package has been
damaged or opened.
5.4.2 Do Not Reuse Indicates a medical device that is
intended for one use, or for use on a
single patient during a single
procedure.
5.4.3
Consult
Instructions for
Use
Indicates the need for the user to
consult the instructions for use, which
you are currently reading.
5.4.4
Caution Indicates the need for the user to
consult the instructions for use for
important cautionary information such
as warning and precautions that
cannot, for a variety of reasons, be
presented on the medical device itself.

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 9
IEC 60601-1, Edition 3.1 Medical Electrical Equipment – Part 1: General Requirements for
Basic Safety and Essential Performance
Reference Symbol Title Meaning
ISO 7010-
M002
Follow
Instructions for
Use
Indicates the need for the user to
consult the instructions for use for
important cautionary information such
as warning and precautions that
cannot, for a variety of reasons, be
presented on the medical device itself.
IEC 60417:2002 DB Graphical Symbols for Use on Equipment
Reference Symbol Title Meaning
5333
Type BF
Applied Part
To identify a type BF applied part
complying with IEC 60601-1.
5334 Electrostatic
Sensitive
Devices
To indicate packages containing
electrostatic sensitive devices, or to
identify a device or a connector that
has not been tested for immunity to
electrostatic discharge.
21 CFR 801. 109 (b) (1), 81 FR 38911 2016-09-13, U.S.A. FDA Guidance: Alternative to
Certain Prescription Device Labeling Requirements 2000-01-21
Symbol Title Meaning
Prescription
Only
Caution: Federal (U.S.A.) law restricts
this device to sale by or on the order
of a physician.

Revision 1.00 / LB-0996 – NeuroPlex E IFU
© 2021 Blackrock Microsystems, LLC 10
Specifications
Model Name
Blackrock NeuroPlex
E128 Non-Sterile
Blackrock NeuroPlex
E128 Sterile
Blackrock NeuroPlex
E96 Non-Sterile
Blackrock NeuroPlex
E96 Sterile
Model Number
10906 10907 10908 10909
Sterility Non-Sterile
Ethylene Oxide
Sterilized
Non-Sterile
Ethylene Oxide
Sterilized
Number of Channels
128 128 96 96
Power Requirements
0 to 5 VDC, 150 mA maximum load
Resolution
16-bit, at 250 nV per bit resolution
Sampling Frequency
30 ksps
Mode of Operation
Continuous
Input Hardware Filter
(User selectable)
Normal mode:
0.3 Hz (1st order) – 7.5 kHz (3rd order)
Wide-band mode:
0.02 Hz (1st order) – 10 kHz (3rd order)
Input Filter Type
Butterworth
Input Impedance
1300 MΩ @ 10 Hz, 13 MΩ @ 1 kHz
Auto
Impedance
Measurement Range
5 kΩ – 1 MΩ
Maximum Input
Range
± 8.192 mV with respect to reference
Ingress Protection
Ordinary Equipment, not fluid resistant, IP20
Dimensions
L 24.9 mm x W 17.7 mm x H 17.9 mm
Operating
Environment
10˚ C to 40˚ C, 5 to 95% R.H. (non-condensing)
Storage Environment
-20˚ C to 50˚ C, 5 to 100% R.H. (non-condensing)
Dieses Handbuch passt für folgende Modelle
4
Inhaltsverzeichnis
Andere Blackrock Microsystems Aufnahmegeräte Handbücher
Beliebte Aufnahmegeräte Handbücher anderer Marken

Strymon
Strymon NIGHTSKY Bedienungsanleitung

Mitsubishi Electric
Mitsubishi Electric 16CH DIGITAL RECORDER DX-TL5000U Bedienungsanleitung

Tews Technologies
Tews Technologies TPMC465 Bedienungsanleitung

Honeywell
Honeywell Excel 50 Bedienungsanleitung

SeaLevel
SeaLevel COMM+8.LPCI Bedienungsanleitung

Arturia
Arturia AUDIOFUSE STUDIO Bedienungsanleitung