BMC INH2 Bedienungsanleitung

English
InH2®Heated Humidifier
User Manual
0123

InH2®Heated Humidifier User Manual V2.7
English
The InH2®Heated Humidifier is designed only for use with specific BMC
RESmart®CPAP/Auto CPAP or BPAP devices. Do not use InH2®with any other
devices.
The humidifier moistens the air delivered by the BMC RESmart®CPAP/BPAP
devices.
The InH2®Heated Humidifier is only used for single patient and must not be
re-used on another person. This is to avoid the risk of cross-infection.
The InH2®Heated Humidifier is not intended for use with a patient whose
upper airways have been bypassed.
Table of Contents
Warnings & Cautions ····································································· 1
Symbols ·····················································································2
Features·····················································································2
Set Up ·······················································································3
Daily Use ····················································································4
Cleaning ····················································································· 4
Service······················································································· 5
Specifications··············································································· 6
Disposal ·····················································································7
Traveling With the System·······························································7
EMC Requirements ········································································ 8
Warranty ·················································································· 13
RESmart®, InH2®are registered trademarks of BMC Medical Co., Ltd.
© BMC Medical Co., Ltd. 2013

InH2®Heated Humidifier User Manual V2.7
1/13
English
Warnings & Cautions
IMPORTANT!
• Read all instructions before using the humidifier.
• Use only with BMC RESmart®devices whose instructions specify the use of
this humidifier.
CAUTIONS!
• Indicates the possibility of damage to the device.
• US federal law restricts this device to sale by or on the order of a physician.
• If fluids are spilled onto the humidifier platform, unplug the power cord from
the AC wall outlet and allow the humidifier platform to drain and dry before
using.
• Take precautions to protect furniture from water damage.
WARNINGS!
• Indicates the possibility for injury to the user or the operator.
• Use the humidifier only for its intended use as described in this manual.
• Use only accessories recommended by BMC.
• Never operate the humidifier if any of the parts are damaged, if it is not
working properly, or if the humidifier has been dropped or mishandled. Do not
use the humidifier if the water chamber is leaking or damaged in any way.
Have any damaged parts replaced before continuing use.
• Never touch the heater plate unless the humidifier is unplugged and the plate
has cooled down.
• This equipment is not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
• Periodically inspect the power cord for signs of wear or damage. Replace if
necessary.
•Do not operate the device in direct sunlight or near a heating appliance
because these conditions can increase the temperature of the air coming out
of the device.
• When humidifier is used outside the specified ambient temperature range or
humidity range, the performance of humidifier will be compromised.

InH2®Heated Humidifier User Manual V2.7
2/13
English
Symbols
Attention, Consult Accompanying Documents
Type BF Applied part
Class II (Double Insulated)
DC Power
IPX1 Drip-Proof, Vertical
European CE Declaration of Conformity
Features
Fill Line:
This indicates the maximum water level for safe operation.
Outlet Port:
Connect the flexible tubing (coaxial 22 mm) here.
Water Chamber:
The removable water chamber holds the water for
humidification. It has a silica gel cover to outflow the water in cleaning.
Heater Plate:
Warms the water in the water chamber.
Indicator Light:
When lit, this indicates that the heater plate has been
turned on.
Heater
Plate
Indicator
Li
g
ht
Water Chamber
Fill Line
Outlet Port

InH2®Heated Humidifier User Manual V2.7
3/13
English
Sensor:
Do not block this sensor. The humidifier will not operate unless it can
detect that the RESmart®device is connected correctly.
Air Inlet with Rubber Connector:
Connect to the outlet port on the
RESmart®device.
Power Connector:
Plug this connector into the power outlet on the BMC
RESmart®Device.
Set Up
1. Connect the Humidifier to the RESmart®device. Make sure they are
connected completely.
2. Place the RESmart®device and humidifier on a firm, flat surface at a level
lower than your sleeping position.
CAUTION!
Do not turn the humidifier on without the water chamber installed.
CAUTION!
Take precautions to protect furniture from water damage.
1
2
3
Senso
r
Air Inlet with Rubber
Connector
Power Connecto
r

InH2®Heated Humidifier User Manual V2.7
4/13
English
Daily Use
1. Fill the chamber to the fill line with water (approx. 350 ml) from the outlet
port. Distilled water is recommended. Do not overfill the water chamber.
CAUTION!
Always remove the chamber from the humidifier before filling with water.
2. Press down the spring loaded heater plate with the water chamber and slide
the chamber into place. Make sure the rubber connector on the inlet port fits
securely over the RESmart®device’s air outlet.
3. Connect the flexible tubing to the outlet port on the water chamber.
CAUTION!
Avoid moving or tilting the humidifier when the water chamber has water in it.
4. When the RESmart®device begins blowing, the humidifier will work
automatically. The yellow indicator light on the humidifier will turn on. Press
the humidifier button can turn off or restart the humidifier.
5. The ideal humidity setting depends on room temperature and humidity.
Initially, a setting of 3 is recommended. You can adjust this setting at any time.
Please adjust the humidifier setting according to the CPAP User Manual.
IMPORTANT!
When the air flow is turned off, the humidifier will automatically shut off.
Cleaning
The water chamber should be cleaned daily or after each use.
WARNING!
Emptying and cleaning the water chamber daily will help to prevent mold and
bacteria growth.
WARNING!
Allow the water in the chamber to cool down to room temperature before
removing it from the humidifier.
WARNING!
To avoid electrical shock, disconnect the power cord of RESmart®device
before cleaning the humidifier. DO NOT immerse the humidifier into any fluids.

InH2®Heated Humidifier User Manual V2.7
5/13
English
Water Chamber
1. Turn the RESmart®device off and allow approximately 15 minutes for the
heater plate and water to cool.
2. Disconnect the tubing from the water chamber. Press down on the water
chamber and slide it out of the humidifier platform.
3. Open the silica gel cover of water chamber and discard any remaining water.
Fill a solution of warm water and a mild dishwashing detergent into the
chamber, cover the silica gel cover, rock the chamber a few minutes, and then
outflow the solution. Rinse the chamber several times with clean water and
allow to air dry.
4. Fill the water chamber and close the silica gel cover. Inspect the water
chamber for any leak or damage. Replace the water chamber if any damage is
present.
Humidifier Platform
1. Clean the humidifier platform by wiping with a damp cloth. Allow to air dry.
2. Inspect the humidifier platform for any damage and replace if necessary.
Silicone Tube of Humidifier
Clean the Silicone Tube by wiping with an alcohol cotton stick. Allow to air dry.
Service
The humidifier does not require routine servicing.
If the humidifier malfunctions, contact your home care provider immediately.
Never attempt to open the humidifier’s enclosure. If necessary, contact your
local authorized dealer or BMC Medical Co., Ltd. for technical support and
documents.

InH2®Heated Humidifier User Manual V2.7
6/13
English
Specifications
Size
Dimensions:120 × 194 × 112 mm
Weight:< 0.8 kg
Water Capacity:> 350 ml at recommended water level
Product Use, Transport and Storage
Operation Transport and Storage
Temperature: 5 to 30℃-20 to 55℃
Humidity: ≤80% Non-condensing ≤93% Non-condensing
Atmospheric Pressure: 860 to 1060 hPa 500 to 1060 hPa
Power Requirements
24V DC 1.0A max
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IPX1 –Drip-Proof, Vertical
Heater Settings
1 to 5 (104 to 149℉/ 40 to 65℃)
Maximum Operating Pressure
30hPa
Pressure Drop w/ Humidifier
<0.5 hPa at 60 LPM flow
Maximum Delivered Gas Temperature
<40℃
Humidity Range
10 to 45 mg H2O/L

InH2®Heated Humidifier User Manual V2.7
7/13
English
Disposal
When necessary, dispose of the device and accessories in accordance with
local regulations.
Traveling With the System
Packing the System
When traveling, the optional CPAP carrying case is for carry-on luggage only.
The carrying case will not protect the humidifier if it is put through checked
baggage.
Security Stations
For ease at security stations, there is a note on the bottom of the humidifier
stating that it is medical equipment. It may be helpful to bring this manual
along with you for security personnel.

InH2®Heated Humidifier User Manual V2.7
8/13
English
EMC Requirements
Guidance and manufacturer's declaration - electromagnetic emissions
The RESmart®is intended for use in the electromagnetic environment specified
below. The user of the RESmart®should ensure that it is used in such an
environment.
Emissions Test Compliance Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1,
Class B
The RESmart®uses RF energy only for its
internal function. Therefore its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Group 1,
Class B The RESmart®is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2 Class A
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Class D
Inhaltsverzeichnis