Buchi CH-9230 Bedienungsanleitung

Ultrasonic Package
Operation Manual
11594107 | A en

Imprint
Product Identification:
Operation Manual (Original) Ultrasonic Package
11594107
Publication date: 01.2020
BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
E-Mail: [email protected]
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experi-
ence, especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for com-
petitive purposes, nor made available to third parties. The manufacture of any component with the aid of
this manual without prior written agreement is also prohibited.

BÜCHI Labortechnik AG Table of contents
Operation Manual Ultrasonic Package iii
Table of contents
1 About this document...........................................................................................................5
1.1 Warning notices in this document........................................................................................................ 5
1.2 Connected devices .............................................................................................................................. 5
1.3 Symbols............................................................................................................................................... 5
1.3.1 Warning symbols .....................................................................................................................5
1.3.2 Mark-ups and symbols.............................................................................................................6
1.4 Trademarks.......................................................................................................................................... 6
2 Safety....................................................................................................................................7
2.1 Proper use ........................................................................................................................................... 7
2.2 Use other than that intended ............................................................................................................... 7
2.3 Staff qualification ................................................................................................................................. 7
2.4 Location of warning signs on the product ............................................................................................ 8
2.5 Residual risks ...................................................................................................................................... 8
2.5.1 Faults during operation ............................................................................................................8
2.5.2 Hot surfaces.............................................................................................................................9
2.5.3 Dangerous vapors.................................................................................................................... 9
2.6 Personal protective equipment ............................................................................................................ 9
2.7 Modifications........................................................................................................................................ 9
3 Product description...........................................................................................................10
3.1 Description of function ....................................................................................................................... 10
3.2 Configuration ..................................................................................................................................... 10
3.2.1 Front view ultrasonic controller ..............................................................................................10
3.2.2 Rear view ultrasonic controller...............................................................................................11
3.2.3 Ultrasonic nozzle.................................................................................................................... 11
3.3 Type plate.......................................................................................................................................... 12
3.4 Scope of delivery ............................................................................................................................... 12
3.5 Technical data ................................................................................................................................... 12
3.5.1 Ultrasonic controller ...............................................................................................................12
3.5.2 Ultrasonic nozzle.................................................................................................................... 12
3.5.3 Ambient conditions.................................................................................................................13
3.5.4 Materials ................................................................................................................................13
4 Transport and storage ......................................................................................................14
4.1 Transport ........................................................................................................................................... 14
4.2 Storage .............................................................................................................................................. 14
5 Installation..........................................................................................................................15
5.1 Before installation .............................................................................................................................. 15
5.2 Installation site................................................................................................................................... 15
5.3 Establishing electrical connections.................................................................................................... 15
5.4 Installing the ultrasonic controller ...................................................................................................... 16
5.5 Installing the ultrasonic nozzle in open mode.................................................................................... 16
5.6 Installing the ultrasonic nozzle in closed mode.................................................................................. 18

Table of contents BÜCHI Labortechnik AG
iv Operation Manual Ultrasonic Package
6 Operation............................................................................................................................22
6.1 Operating the display......................................................................................................................... 22
6.1.1 Layout of the display.............................................................................................................. 22
6.1.2 Entering settings ....................................................................................................................22
6.1.3 Selecting the operation mode ................................................................................................22
6.2 Finding the optimal spray cone.......................................................................................................... 23
6.2.1 Finding the optimal spray cone (Power-leveling mode)......................................................... 23
6.2.2 Finding the optimal spray cone (Load-leveling mode) ...........................................................23
6.3 Carrying out a spray drying process in open mode ........................................................................... 23
6.3.1 Tasks before starting spray drying in open mode.................................................................. 23
6.3.2 Preparing the instrument for open mode ...............................................................................24
6.3.3 Starting a spray drying process in open mode....................................................................... 24
6.3.4 Tasks during spray drying......................................................................................................24
6.3.5 Ending a spray drying process in open mode........................................................................24
6.3.6 Shutting down the instrument ................................................................................................25
6.4 Carrying out a spray drying precess in closed loop mode................................................................. 25
6.4.1 Tasks before starting spray drying in closed mode................................................................ 25
6.4.2 Preparing the instrument for closed mode............................................................................. 25
6.4.3 Starting a spray drying process in closed mode ....................................................................26
6.4.4 Tasks during spray drying......................................................................................................26
6.4.5 Ending a spray drying process in closed mode .....................................................................26
6.4.6 Shutting down the instrument ................................................................................................27
7 Cleaning and servicing .....................................................................................................28
7.1 Regular maintenance work................................................................................................................ 28
7.2 Cleaning the ultrasonic nozzle........................................................................................................... 28
8 Help with faults ..................................................................................................................30
8.1 No liquid delivery ............................................................................................................................... 30
8.2 Liquid emerges unatomized............................................................................................................... 30
8.3 Error messages ................................................................................................................................. 30
9 Taking out of service and disposal..................................................................................31
9.1 Taking out of service.......................................................................................................................... 31
9.2 Disposal............................................................................................................................................. 31
9.3 Returning the instrument ................................................................................................................... 31
10 Appendix ............................................................................................................................32
10.1 Spare parts and accessories ............................................................................................................. 32
10.1.1 Accessories............................................................................................................................32
10.1.2 Spare parts ............................................................................................................................32

BÜCHI Labortechnik AG About this document | 1
Operation Manual Ultrasonic Package 5/34
1 About this document
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the instruc-
tions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or
owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions re-
sulting from not following this operation manual.
If you have any questions after reading this operation manual:
uContact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
1.1 Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device. There
are four danger levels, each identifiable by the signal word used.
Signal word Meaning
DANGER Indicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNING Indicates a danger with a medium level of risk which could result in
death or serious injury if not prevented.
CAUTION Indicates a danger with a low level of risk which could result in mi-
nor or medium-severity injury if not prevented.
NOTICE Indicates a danger that could result in damage to property.
1.2 Connected devices
In addition to these operating instructions, you should also follow the instructions and
specifications in the documentation for the connected devices.
1.3 Symbols
The following symbols are displayed in this operation manual or on the device:
1.3.1 Warning symbols
Symbol Meaning
General warning
Hot surface
Dangerous electrical voltage
Device damage

1 | About this document BÜCHI Labortechnik AG
6/34 Operation Manual Ultrasonic Package
1.3.2 Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
RThis character draws attention to a requirement that must be met before the in-
structions below are carried out.
uThis character indicates an instruction that must be carried out by the user.
ðThis character indicates the result of a correctly carried out instruction.
Mark-up Explanation
Window Software Windows are marked-up like this.
Tab Tabs are marked-up like this.
Dialog Dialogs are marked-up like this.
[Button] Buttons are marked-up like this.
[Field names] Field names are marked-up like this.
[Menu / Menu item] Menus or menu items are marked-up like this.
Status Status is marked-up like this.
Signal Signals are marked-up like this.
1.4 Trademarks
Product names and registered or unregistered trademarks that are used in this docu-
ment are used only for identification and remain the property of the owner in each
case.

BÜCHI Labortechnik AG Safety | 2
Operation Manual Ultrasonic Package 7/34
2 Safety
2.1 Proper use
The device has been designed and built for laboratories.
The ultrasonic nozzle together with the ultrasonic controller are designed for spray
drying purposes using the BUCHI Mini Spray Dryer B-290 and its subsystems.
The ultrasonic nozzle together with the ultrasonic controller are designed to atom-
ize liquids with viscosities lower or equal to 50 cps.
2.2 Use other than that intended
Use of any kind other than that described in the section Proper use Chapter2.1
"Proper use", page7 and any application that does not comply with the technical
specifications (see Chapter3.5 "Technical data", page12) constitutes use other
than that intended.
In particular, the following applications are not permissible:
Use of the instrument with non-BUCHI products nor with any other BUCHI model
than the Mini Spray Dryer B-290.
Use of the instrument in an environment with a potential risk of explosion or areas
which require explosion-safe apparatus.
Use of the instrument for processing substances outside of research and develop-
ment.
Use of the instrument for food, pharmacy and cosmetic products without appropri-
ate cleaning.
Use of the instrument without an appropriate leading away exhaust gas from the
working area.
Use of gases with unknown chemical composition.
Use of the instrument with organic solvents (> 20 %) without safety lamella curtain
and Inert Loop B-295.
Use of the instrument with organic solvents (> 20 %) in open mode.
Production and processing of substances that can lead to spontaneous reactions,
such as explosives, metal hydrides or solvents that can form peroxides.
Use of the instrument with samples which produce oxygen during the processing.
Use of the instrument with toxic substances without appropriate safety measures.
Use of the instrument with biohazardous materials such as viruses or bacteria.
Use of the instrument with samples which can block the feed channel of the ultra-
sonic nozzle.
Use of the instrument with substances which might explode or ignite due to the
processing and the selected parameters.
Use of the instrument with corrosive samples other than the Mini Spray Dryer
B-290 ACID.
Use of the instrument with corrosive samples in closed mode.
2.3 Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater
dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:

2 | Safety BÜCHI Labortechnik AG
8/34 Operation Manual Ultrasonic Package
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the applica-
ble safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator (generally the laboratory manager) is responsible for the following as-
pects:
The device must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the operations
described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to the
manufacturer ([email protected]).
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and
are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair
measures.
2.4 Location of warning signs on the product
1
5
Fig.1: Location of the warning signs
General warning Electricity
2.5 Residual risks
The device has been developed and manufactured using the latest technological ad-
vances. Nevertheless, risks to persons, property or the environment can arise if the
device is used incorrectly.
Appropriate warnings in this manual serve to alert the user to these residual dangers.
2.5.1 Faults during operation
If a device is damaged, sharp edges, moving parts or exposed electrical wires can
cause injuries.
uRegularly check device for visible damage.

BÜCHI Labortechnik AG Safety | 2
Operation Manual Ultrasonic Package 9/34
uIf faults occur, switch off the device immediately, unplug the power cord and in-
form the operator.
uDo not continue to use devices that are damaged.
2.5.2 Hot surfaces
The surfaces of the device can become very hot. If touched they can cause skin
burns.
uDo not touch hot surfaces or else wear suitable protective gloves.
2.5.3 Dangerous vapors
The use of the instrument can produce dangerous vapors that are capable of causing
life-threatening toxic effects.
uDo not inhale any vapors produced during processing.
uEnsure that vapors are removed by a suitable fume hood.
uOnly use the device in well ventilated areas.
uIf vapors escape from connections, check the seals concerned and replace them if
necessary.
uDo not process any unknown fluids.
uObserve the safety data sheets for all substances used.
2.6 Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may
arise.
uAlways wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
uMake sure that the personal protective equipment meets the requirements of the
safety data sheets for all chemicals used.
2.7 Modifications
Unauthorized modifications may impair safety and lead to accidents.
uUse only genuine BUCHI accessories, spare parts and consumables.
uTechnical modifications to the device or accessories should only be carried out
with the prior written approval of BÜCHI Labortechnik AG and only by authorized
BUCHI technicians.
BUCHI accepts no liability whatsoever for damage arising as a result of unauthorized
modifications.

3 | Product description BÜCHI Labortechnik AG
10/34 Operation Manual Ultrasonic Package
3 Product description
3.1 Description of function
The ultrasonic nozzle with the ultrasonic controller are built to be used as liquid atom-
izer in the spray drying process with the Mini Spray Dryer B-290 and its subsystems.
An ultrasonic atomizing nozzle is a device that vibrates at frequencies beyond human
hearing (>20 kHz). The vibration on the atomizing surface of the nozzle produces a
low velocity spray by breaking up a liquid feed into a spray of droplets. This phenome-
non is the result of the decay of an unstable capillary wave that develops in the liquid.
The spray velocity is typically low with 18-36 cm/s as compared to 10-20 m/s for pres-
sure atomizing nozzles. The ultrasonic controller provides the high frequency electri-
cal energy required to operate the ultrasonic atomizer nozzle.
Three modes are available:
Mode Solvent composition
Open mode up to 20 % organic solvent
Closed mode with
B-295
between 80 % - 100 % organic solvent
Closed mode with
B-295 and B-296
between 20 % - 80 % organic solvent
3.2 Configuration
3.2.1 Front view ultrasonic controller
4
3
2
1
Fig.2: Front view
1 On/Off master switch 2 Display
3 Navigation control 4 LED