Dataq DI-4108 Bedienungsanleitung

The way PC-based instrumentation should be
DI-4108 and DI-4208
8-Channel High Speed Data Acquisition Systems with Programmable Gain,
Expandability with ChannelStretch™, and Stand-alone Capability
User's Manual
Manual Revision G
Copyright © 2020 by DATAQ Instruments, Inc. The Information contained herein is the exclusive property of
DATAQ Instruments, Inc., except as otherwise indicated and shall not be reproduced, transmitted, transcribed, stored
in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic,
mechanical, magnetic, optical, chemical, manual, or otherwise without expressed written authorization from the com-
pany. The distribution of this material outside the company may occur only as authorized by the company in writing.
DATAQ Instruments' hardware and software products are not designed to be used in the diagnosis and treatment of
humans, nor are they to be used as critical components in any life-support systems whose failure to perform can rea-
sonably be expected to cause significant injury to humans.
DATAQ, the DATAQ logo, and WINDAQ are registered trademarks of DATAQ Instruments, Inc. All rights reserved.
DATAQ Instruments, Inc.
241 Springside Drive
Akron, Ohio 44333 U.S.A.
Telephone: 330-668-1444
Fax: 330-666-5434
Designed and manufactured in the
United States of America


iii
Warranty and Service Policy
Product Warranty
DATAQ Instruments, Inc. warrants that this hardware will be free from defects in materials and workmanship under
normal use and service for a period of 1 year from the date of shipment. DATAQ Instruments' obligations under this
warranty shall not arise until the defective material is shipped freight prepaid to DATAQ Instruments. The only
responsibility of DATAQ Instruments under this warranty is to repair or replace, at its discretion and on a free of
charge basis, the defective material.
This warranty does not extend to products that have been repaired or altered by persons other than DATAQ Instru-
ments employees, or products that have been subjected to misuse, neglect, improper installation, or accident.
DATAQ Instruments shall have no liability for incidental or consequential damages of any kind arising out of the
sale, installation, or use of its products.
Service Policy
1. All products returned to DATAQ Instruments for service, regardless of warranty status, must be on a freight-pre-
paid basis.
2. DATAQ Instruments will repair or replace any defective product within 5 days of its receipt.
3. For in-warranty repairs, DATAQ Instruments will return repaired items to the buyer freight prepaid. Out of war-
ranty repairs will be returned with freight prepaid and added to the service invoice.


DI-4108 and DI-4208 Hardware Manual
Table of Contents
v
Table of Contents
Warranty and Service Policy ................................................................................................................ iii
1. Introduction ........................................................................................................................................ 1
Features .............................................................................................................................................. 1
Analog Inputs .................................................................................................................................... 1
Digital I/O .......................................................................................................................................... 2
Software ............................................................................................................................................. 2
WinDaq® Recording and Playback Software ............................................................................ 2
Help ............................................................................................................................................. 2
2. Specifications ...................................................................................................................................... 3
3. Installation .......................................................................................................................................... 7
Install WinDaq Software ................................................................................................................... 7
Connect the Instrument to Your Computer ....................................................................................... 8
Install an Ethernet Device on Your Local Area Network ................................................................. 8
4. Controls, Indicators, and Connections ............................................................................................. 9
Mini-B USB Connection ................................................................................................................... 9
Screw Terminals ................................................................................................................................ 9
DI-4108 Signal Connections ....................................................................................................... 10
DI-4208 Signal Connections ....................................................................................................... 10
GnD: Ground. .............................................................................................................................. 11
Connecting Signal Leads ............................................................................................................ 11
Analog Inputs .............................................................................................................................. 11
Low-pass Filter ..................................................................................................................... 13
Resolution as a Function of Sample Rate ............................................................................. 13
Digital Ports ................................................................................................................................ 14
WinDaq Remote Events (D0 Event) .................................................................................... 15
WinDaq Remote Storage (D1 Record) ................................................................................. 17
WinDaq Rate (D2 Rate) ....................................................................................................... 17
WinDaq Counter (D3 Count) ............................................................................................... 19
General Purpose Digital Inputs ............................................................................................ 21
General Purpose Digital Outputs .......................................................................................... 22
4-20mA Current Loop Measurements ............................................................................................... 23
Mode LED Indicator .......................................................................................................................... 24
Control Button Operations ................................................................................................................. 25
Record Data to USB Thumb Drive ............................................................................................. 25
Stop Recording to USB Thumb Drive ........................................................................................ 26
Fix Errors .................................................................................................................................... 26
Switching between LibUSB and CDC Modes ............................................................................ 26
Save Ethernet and Stand-alone configuration to Thumb Drive .................................................. 26
Apply Ethernet and Stand-alone Configuration to Device ......................................................... 27
WinDaq Dashboard ........................................................................................................................... 27
ChannelStretch™ ............................................................................................................................... 28
Connection .................................................................................................................................. 29
Channel Selection ....................................................................................................................... 30
5. Unlock WinDaq or Stand-alone ........................................................................................................ 31
6. Dimensional Drawing ......................................................................................................................... 33


DI-4108 and DI-4208 Hardware Manual
Introduction
1
1. Introduction
This manual contains information designed to familiarize you with the features and functions of the DI-4108 and DI-
4208 USB data acquisition systems.These systems are identical excluding gain ranges.
Features
Both the DI-4108 and DI-4208 data acquisition instruments are portable data recording modules that communicate
through your computer's USB and/or Ethernet port with stand-alone capability. Features include:
• 8 fixed differential analog inputs protected to 120 V rms.*
• USB, Ethernet, and Power over Ethernet (PoE) interface options.
• 7 digital ports protected to +25V.*
• Up to 16-bit measurement resolution (dependent on sample rate).
• The DI-4108 has programmable measurement ranges of ±200, ±500 mV, ±1, ±2, ±5, ±10 Volts.
• The DI-4208 has programmable measurement ranges of ±2, ±5, ±10, ±20, ±50, ±100 Volts.
• Up to 160 kHz maximum throughput sampling rate.
• Adaptable low-pass filter per channel - corner frequency set as a function of sampling rate
• A multi-purpose push-button for WINDAQ remote events and start/stop recording.
• Multi-color LED indication for easy notification of device status.
• Free WinDaq/Lite data acquisition recording software. Records up to four channels of data. Purchase an unlock
code to record more than four channels.
• Stand-alone option allows the user to configure a recording session through their PC then record data in the field
to removable flash memory via the USB drive port (without being connected to their PC). Stand-alone data log-
ging requires Unlock code to record more than 1 channel to removable flash memory.
• Counter channel.
• Frequency channel.
• ChannelStretch™ Technology allows you to sync up to 16 DI-1120, -4108, or -4208 instruments in any combina-
tion with a total throughput of 480kHz or more (requires unlock code for each device).*
• Included .Net Class supports programming under any .Net programming language.*
• Fully documented instrument protocol for programming the device in operating systems other than Windows.*
*WinDaq/Unlock required per device to use ChannelStretch™ and to record more than four total channels when
using WinDaq software. These restrictions do not apply when programming the device yourself.
Analog Inputs
Both the DI-4108 and DI-4208 feature eight differential channel inputs located on a single sixteen-position screw ter-
minal block for easy connection and operation. Gain measurement ranges of ±200, ±500 mV, ±1, ±2, ±5, ±10 Volts
for the DI-4108 or ±2, ±5, ±10, ±20, ±50, ±100 Volts for the DI-4208 are programmable on a per channel basis. Ana-
log channels are protected to 120 V rms.
Utilize the functionality of WINDAQ software to experience all the features encased in these small, inexpensive
instruments. An Unlock Code is required to record more than four channels of any type on either device or to syn-

DI-4108 and DI-4208 Hardware Manual
Introduction
2
chronize multiple units when using WinDaq software. There are no restrictions when programming the instruments
yourself.
Digital I/O
The DI-4208 contains seven digital ports, which may be used as a general-purpose digital input or for a specific func-
tion as designated on the device. Digital port D0 may be used for WinDaq Events; Digital port D1 may be used for
WinDaq remote Start/Stop; Digital port D2 may be used for rate measurements; Digital port D3 may be used for
counts; Digital ports D4 and D5 are general purpose digital input ports. Please Note: Digital outputs are not supported
in WinDaq software. Each port may be configured as a switch using either the provided protocol or .Net class.
Software
All software required to record and playback waveforms is included with the purchase of any DI-4108 or 4208 data
acquisition system via download. Use the WinDaq Dashboard to access WinDaq Acquisition software or to setup the
stand-alone configuration.
WINDAQ®Recording and Playback Software
WINDAQ Acquisition and WINDAQ Waveform Browser allow you to record and playback data acquired through your
instrument. WINDAQ software is an invaluable resource to record and analyze your data and is available for free from
our web site (www.dataq.com).
WINDAQ Lite Data Acquisition software (free) can be used to record waveforms directly and continuously to disk
while monitoring a real time display of the waveforms on-screen. It operates, displays, and records up to four chan-
nels in real time. An optional unlock code is available to record the rest of the channels on the device or to allow you
to sync multiple units.
WINDAQ Waveform Browser playback software (also known as “WWB”) offers an easy way to review and analyze
acquired waveforms. A built-in data file translator allows the user to display multiple waveforms acquired by
WINDAQ Acquisition software or any of a wide range of data acquisition packages. The software’s disk-streaming
design allows data files of any length to be graphically displayed rapidly, in normal or reverse time directions. Seven
standard cursor-based measurements, frequency domain, and statistical analysis functions help simplify waveform
analysis and interpretation. WINDAQ Waveform Browser is free and installed when installing WINDAQ Software.
Help
All WINDAQ software utilizes context-sensitive help. Help may be accessed through the Help menu or by pressing the
F1 key with any pull-down menu item selected. This will take you directly to the Help topic most relevant to that par-
ticular function or feature. Help topics discuss in detail each function available in the software.

DI-4108 and DI-4208 Hardware Manual
Specifications
3
2. Specifications
Analog Inputs
Number of Channels: 8
Channel Configuration: Differential
Measurement range per channel: DI-4108: Programmable ±200, ±500 mV, ±1, ±2, ±5, ±10 V
DI-4208: Programmable ±2, ±5, ±10, ±20, ±50, ±100 V
Input impedance: DI-4108: 100 k differential
DI-4208: 795 k differential
Absolute accuracy: 0.25% of full scale range @ 25°C, excluding common mode error
Absolute maximum input without damage:
120 V rms (normal mode + common mode)
Maximum common mode voltage:
Common mode range: ±20 V dc or peak ac
Common mode rejection ratio: 50 dB (dc - 60 Hz)
Channel-to-channel crosstalk rejection: -80 dB
ADC Characteristics
Resolution: 12- to 16-bit
Above zero ADC counts: 2,047 to 32,767
Below zero ADC counts: 2,048 to 32,768
Max. throughput sample rate: Standard: 160 kHz
Stand-alone: 20 kHz to 160 kHz (dependent on flash memory)
Min. throughput sample rate: Standard: 20 Hz (2.2 samples per hour with WinDaq)
Stand-alone: 0.0017 Hz
Sample rate timing accuracy: 50 ppm
Digital Ports
Number of Ports: 7
Type: MOSFET switch
Configuration: Programmable as digital input or switch
Threshold levels TTL compatible
Absolute maximum applied voltage (V): 25 V dc
Reserved Digital Inputs
Port 1: WINDAQ remote events
Port 2: WINDAQ remote start/stop
Port 3: Rate input
Port 4: Count input
Count/Rate/Digital Inputs
Internal pull-up value: 4.7 k
Input high voltage threshold: 2.4 V

DI-4108 and DI-4208 Hardware Manual
Specifications
4
Input low voltage threshold: 0.8 V
Terminal count: 65,535
Maximum rate frequency: 50 KHz with one enabled channel
20 KHz with 2-4 enabled channels
10 KHz with 5-7 enabled channels
Minimum rate frequency: 0.5 Hz
Maximum count frequency: 50 kHz
Digital Ports Programmed as Switch
Maximum drain voltage: 25V
Maximum sink current: 100 mA
Ethernet Interface
Type: 10/100Base-T
Connector: RJ-45
Protocol: TCP/IP
Server Type: DHCP or Fixed IP
Removable Memory
Type: MLC, pSLC, or SLC flash memory (with USB SD card reader); USB
thumb drive
Required Format: FAT32
Number of Channels: 1; Unlock code required for 2 or more
Indicators and Connections
USB Interface: USB 2.0 (mini-B style connector)
Button: Multi-function control (Start, Stop, Save Configuration, Apply Configu-
ration)
USB A Connector (Drive): USB drive for Stand-alone data logging
Cannot be used concurrently with the USB interface
Input connections: Two 16-position screw terminal strips
Status LED: One multicolor LED for status indication
USB ChannelStretch™ Operation
Total Max Units: 16
Max Channel Count: 128 Analog, 112 Digital
Max Throughput: ≥480 kHz (higher speeds may result in data loss)
Channel skew between any 2 units: 10 µS, typical
Synchronization Conditions: Syncs with any combination of DI-1120, DI-4108, DI-4208, and DI-
4718B USB instruments.
All synced instruments must operate at the same sample rate per channel,
regardless of channel number, type, or range.
All instruments must be connected to the same USB controller. The use
of USB hubs are recommended to meet this requirement.
Dieses Handbuch passt für folgende Modelle
1
Inhaltsverzeichnis
Andere Dataq Datenlogger Handbücher

Dataq
Dataq DI-1100 Bedienungsanleitung

Dataq
Dataq DI-718B Series Bedienungsanleitung

Dataq
Dataq DI-245 Bedienungsanleitung

Dataq
Dataq DI-2008 Bedienungsanleitung

Dataq
Dataq DI-4718B Bedienungsanleitung

Dataq
Dataq DI-710 Series Bedienungsanleitung

Dataq
Dataq DI-1110 Bedienungsanleitung

Dataq
Dataq DI-2108-P Bedienungsanleitung

Dataq
Dataq DI-4730 Bedienungsanleitung