GasLab Pro Multi Gas Sampling Data Logger
Operating Instructions
CM-1000
Product Overview
Thank you for selecting the CM-1000 multi gas sampling data logger. The CM-1000is designed to
simultaneously measure multiple gas concentrations through sampling methods.With long-termdata storage,
the user can assess the previous records.When the CM-1000measures the gas concentration reaching the
alarm setting or higher, the data display and alarm functions areactivated. The CM-1000is equipped with
an RS485 interface for connecting to a computer for remote monitoring. The Multi Gas Sampling Data
Logger can be used in a wide-range of industries and applications.
Features:
V□Options for CO2measurement range: 1%, 5%, 10%, 20% 30%, and 100%
V□Measure CO2,CO, O2, CH4, RH, AMB, DP, ALTI
V□Large LCD for easy reading of gas concentration and temperature and humidity.
V□Audible alarm
V□Data logging with SD card
V□Through the RS485 interface connection, multiple CM-1000 units can be connected to the back-end
computer system as a security monitoring application.
V□Built-in LCD back light for easy reading in the dark.
V□This device is supplied with Li- ion 18650 3.7V rechargeable batteries,can be used for along time.
A. Power Key B. Temperature & Humidity Sensor C. LCD
D.Rubber Boot E.Up Key F.Left Key G.Right Key
H.Down Key I. Mode Key J.EnterKey K.Pump Key
L.USB Socket M.SD Card Slot N.RS485 Jack O.Analog / Voltage Output Port
Key description:
I
E
H
F
G
Change parameter values or select options.
J
K
Start the pump or data logger.
LCD display symbol description:
CO2concentration,
million (ppm)The current CO2concentration
CO concentration, partsper
million (ppm)
The current CO concentration
O2concentration,% (percent)
The current O2concentration
4
concentration, parts per
million (ppm)
The current CH4concentration
Ambient temperature Ambient temperature
Dew point The current dew point
Atmospheric pressure
Compensate the pressure changes with
appropriate altitude of location when
Time weighted average (8 hours) CO2, 8-hr time weighted average
‐
TermExposure Limit
(15 min. weightedaverage)
CO2, 15 min. weighted average
Maximum value Maximum value of gas measurement
Minimum value Minimum value of gas measurement
High/low alarm Customize the range of color Bar
Battery capacity Battery capacity without segment display.
Warning
Need to calibrate the sensor or exceed the
pump life time.
Pump
The pump icon rotates to indicate that the
pump has been started.
Micro SD card
Micro SD card working status (When the
icon flashes, it means that the Micro SD
card is broken or full; when the icon remains
stationary, it indicates normal operation.)
Operation Instructions
1. Power on:
Switch the Power Key (○
A) on.
2. Measurement:
After power on, the device starts to measure and update the data every second.
Press Enterkey (○
J) to select zone. The selection order is: Zone 1Zone 2Zone 3. The selected zone will
display a red frame. Press Left (○
F) / Right (○
G) key to select the parameter or sensor to be adjusted. The
selected parameter orsensor displays the red background color, press Right key (○
G) to proceed the setting
and thenpress Up (○
E) / Down (○
H) key to edit the parameter value.
Zone 1 sensor display: CO2, O2, CO, CH4
Zone 2 sensor display: O2, CO, CH4, RH
Zone 3 sensor display: CO, CH4, AMB, DP,ALTI
Zone 1 parameter display and setting
Item
AL L / AL H
○
E
○
H
) key to
edit the parameter value.
Note:AL LandAL H are expressed as a percentage. Example: 0.04 = 400ppm
Zone 3 parameter unit selection:
3. Display allsensor readings
Press and hold Enter key (○
J) todisplay all sensorreadings. Press Mode key (○
I) to return to the main
screen.
4. Setting
Press and hold Mode key (○
I) to enter the setup menu, then press Mode key (○
I) to continue to the next
page. Press Up (○
E) / Down(○
H) key to select options or change parameter values.And then press Enter key
(○
J) to confirm it. Press and hold Mode key (○
I) to return to the main screen.
The first page:
Yes (The buzzer sounds continuously.) / No (The buzzer sounds once.)
1~32 (The larger the number, the brighter the screen)
□□□□
(There are a total of 4 background colors.)
10~999 sec (Backlight display time)
CO2/ O2/CO/ CH4 (The second trendline shows the gas measurements.)
CO2Meter – www.co2meter.com
The second page:
0-10V/4~20mA (Zone 1 voltage or current output)
CO2/O2/CO/CH4 (Another analog output option)
Data logging interval time
1~5 (The largerthe number, the greater the pump output)
Sensor times info
Sensor recalibrated time and pump life time (The time is displayed in the
countdown mode. If you need to calibrate the sensor or exceed the pump
life time, the time will be displayed in red.)
CO2: 99999999 days
O2: 99999999 days
CO: 99999999 days
CH4: 99999999 days
Yes / No (Recover the factory setting to cancel customize setting.)
Please refer to the calibration section.
5. Power off:
Switch the Power Key (○
A) off.
Note: When the device is charged with USB, it cannot be turned off.
6. Calibration:
(1) Single-point calibration
After entering the setup menu, then press Mode key (○
I)to continueto thenextpage. Press Down key (○
H) to
select"Sensor calibration (▲+▼)"and press Enter key (○
J)to confirmit. Press Up(○
E)and Down (○
H)keys
at the same time to proceed the calibration. Then, press Enter key (○
J) to enter the sensor settings.
(2) Two-point calibration (calibrate by professional)
After entering the setup menu, then press Mode key (○
I) to continue to the next page. Firstly, press and hold
Up key (○
E), then press Down key (○
H). ”1” will appear next to "Sensor calibration (▲+▼)". Press Down
key (○
H) to select "Sensor calibration (▲+▼) 1" and press Enter key (○J) to confirm it. Press Up (○
E) and
Down (○
H) keys at the same time to proceed the calibration. Then, press Enter key (○J) to enter the sensor
settings.
Low concentration value of standard gas
High concentration value of standard gas
Connect the low concentration gas to the device, and
press Pump key (○K) to start the pump. (Set the data
log time to less than 1 min). After the reading is
stable, press Enter key (
○J
Level 2 catch Connect the high concentration gas to the device,
and press Pump key (○
K) to start the pump. (Set the
data log time to less than 1 min). After the reading is
stable, press Enter key (
○J
When "Start" turns red, press Enterkey (
○J
) to
confirm the calibration.
Gain The amount of Gain.
Perform reading Adjusted reading
(Display Features and Modes)