hatch A50B Bedienungsanleitung

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
HATCH GROUP INC.
Pressure Steam Sterilizer
Autoclave A50B 3L Operation Manual V1.0

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
2
Catalogue
Catalogue.............................................................................................................................................................................2
Scope of Use........................................................................................................................................................................3
User Guide...........................................................................................................................................................................3
1. Brief introduction............................................................................................................................................................7
2. Specifications..................................................................................................................................................................9
3. Installation.....................................................................................................................................................................11
4. Adjusting.......................................................................................................................................................................15
5. Introduction of preset programs ....................................................................................................................................15
6. User interface description..............................................................................................................................................17
7. Instructions of operation................................................................................................................................................23
8. Errors list.......................................................................................................................................................................28
9. Maintenance..................................................................................................................................................................31
10. Quality assurance ........................................................................................................................................................37
11. Accessories..................................................................................................................................................................37
12. appendix......................................................................................................................................................................38
Appendix 1 : Inspection items and results.........................................................................................................................38
Appendix 2: Print record sample and instructions.............................................................................................................38
Appendix 3:Explanation of each stage...........................................................................................................................39

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
3
Scope of Use
This manual is applicable to the tabletop pressure steam sterilizers.
Model:A50B
Please read this manual carefully before use to ensure that the equipment is safe and reliable for your
service!
Please keep this manual during the life of the equipment. If the manufacturer has any necessary
updates, ensure that all updates received are kept as attachments with this manual.
When the place of use or the unit of use of the equipment changes, the manual must be transferred or
handed over as part of the whole equipment.
Please read this manual carefully before using the equipment,be familiar with the
operati
on and safety instructions of the equipment. Do not use instruction that is not from the
manufacturer for reference.
We reserve the right to change the design without prior notice.The information contained in this
manual is current at the time of publication.
User Guide
For safe and reliable use of the equipment, please be sure to pay attention to these instructions.
Safety Caution:
The machine is equipped with some necessary safety measures, in order to avoid causing harm, it is
strictly prohibited to terminate or destroy these safety measures.
Important Notice
Please read this instruction carefully before use.
The equipment must be operated and maintained by professional personnel who must have received
good training.

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
4
For normal operation of the equipment, please keep the equipment clean, do not rinse or water the
equipment;
The maintenance of the equipment must be carried out by the personnel authorized by the
manufacturer, who have passed the corresponding training of the manufacturer;
Equipment accessories or attachments can only be obtained from the manufacturer, otherwise the safe
and effective operation of the equipment cannot be guaranteed.
Emergency handling:
If an emergency is found, it shall be handled as follows:
⚫Turn off the power switch; ⚫
Unplug.
Product responsibilities:
Without the written permission of the manufacturer, it shall not be modified or used beyond the scope.
The manufacturer shall not be liable for any damage caused thereby.
Symbol description:
Special attention should be paid to the danger, warning and attention in this manual
:
Indicates potential hazard to personnel and equipment and must be strictly observed.
:
Indicates potential hazard to personnel and must be strictly observed.
:
The potential harm to the equipment must be taken seriously.
Protective grounding of TypeⅠequipment
Watch for burns on hot surfaces
Refer to manual
Face up

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
5
No rolling
Stack cannot exceed 2 layers
Temperature range 5℃~40℃
Relative humidity is no greater than 80%
keep dry
Safety Notices:
This sterilizer is only suitable to sterilize moisture- and heat-resistant medical devices, not okay
for oils and powders, such as petroleum jelly and agar.
This sterilizer cannot be used for sterilization of liquids or liquids sealed with closed containers
(especially glassware), which may cause the containers to burst, endangering the safety of
persons and equipment.
Chloride ion is an important factor causing corrosion of stainless steel. If you use this sterilizer
to sterilize items which contains chloride ions, you must rinse the sterilization chamber with
clean water after each cycle to prevent the deposited chloride ions from corroding the pot and
extending the service life of the equipment.
When you see the symbol anywhere on the device, it means that the surface temperature
is high, avoid touching to avoid burns.
This equipment complies with the emission and immunity requirements of Class A equipment
specified in GB / T 18268. It may cause interference when used in home or similar environment,
and protective measures are required.
It is recommended to evaluate the electromagnetic environment before using the equipment. It
is forbidden to use this device near strong radiation sources (such as unshielded radio frequency
sources), otherwise it may interfere with the normal operation of the device.

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
6
When an unexpected situation occurs during the use of the product, if the device alarms or other
abnormal conditions occur,please immediately cut off the power supply ofthe device, and check
and correct according to the "abnormal conditions" chapter of the manual.
Necessary monitoring shall be carried out in accordance with relevant national and regional
regulations. After putting the monitoring substance (such as chemical indicator or biological
indicator) into the equipment, run the corresponding program to monitor the sterilization effect,
and judge according to the result. If it fails, find the cause or contact the manufacturer.
A three-hole socket (220 ±22VAC / 16A / 50Hz) must be used to ensure that the socket is reliably
grounded. Do not place the device where it is difficult to disconnect the power.
Do not use other power supplies with different voltages or frequencies.
Do not touch the plug and socket with wet hands.
Do not damage, modify, pull, excessively bend or twist the power cord, and do not place heavy
objects on the power cord.
Please do not place the sterilizer on an unstable workbench, such as a shaking table, a sloping
surface or a place where it will vibrate.
Do not block or cover the sterilizer door, vents or heat dissipation windows.
Please do not place anything on the sterilizer.
If the sterilizer is not used for a long time due to various reasons, the power cord plug should be
disconnected from the power supply and placed in a dry and cool place.

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
7
1. Brief introduction
LF series high-pressure steam sterilizer is specially designed and manufactured by for clinics, hospital
departments, laboratories and other occasions requiring frequent sterilization.Used by a physician or
professional.The sterilizer uses microprocessor intelligent control, man-machine interface, easy to
operate, safe and reliable. The parameters and states in the process of operation are displayed
dynamically by the LCD screen. Automatic fault diagnosis, over temperature, over pressure automatic
protection, ensure the effectiveness of sterilization and disinfection.
1.1 Product categories:
The product is classified as class I type B equipment according to electrical safety;
The product is classified as type B equipment according to YY/T 0646 "small steam sterilizer -
automatic control" standard.
This product is classified as class A equipment according to the electromagnetic compatibility GB/T
18268 standard.
1.2 Product structure:
This product mainly consists of sterilization chamber, sterilization chamber door and seal ring, steam
generator, condenser and fan, vacuum pump, water pump, solenoid valve, sensor, heating ring,
bacteria filter, piping system, power supply, control system, accessories (tray, tray rack, tray handle).
1.3 Range of application
It is used for sterilizing medical instruments which are resistant to heat and humidity.
Don't sterilize the liquid! And ensure that the sterilized equipment can withstand heat and
humidity.
1.4 Sterilization principle, main sterilization factors and strength
1.4.1 Sterilization principle
This equipment is suitable for the vacuum pump to exclude the cold air in the sterilizing room,
take saturated steam as the sterilizing factor, and use the characteristics of high latent heat and high
penetrability of saturated steam to sterilize the equipment.
1.4.2 Sterilization factor and strength
1)sterilization factor and strength of the sterilization procedure at 134℃: the temperature of
saturated water vapor is between 134℃and 137℃, and the temperature difference between each

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
8
point in the sterilization chamber at the same time is no more than 2℃, and the maintenance time is
4min;
2)sterilization factor and strength of the sterilization process at 121℃: the temperature of
saturated water vapor is between 121℃and 124℃, and the temperature difference between each
point in the sterilization chamber at the same time is no more than 2℃, and the maintenance time is
20min.
1.5 Visual section of the device
1
Touch Screen
7
Socket
13
Switch sensor
2
Handle
8
Ventage
14
Door locker
3
Printer
9
Outlet of waste water tank
15
Outlet of waste water tank
4
Clean water tank
10
Outlet of clean water tank
16
Door seal
5
Power switch
11
Safety valve
17
Button battery housing
6
Fuse
12
SD card port

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
9
2. Specifications
2.1 Chamber specification:
Available space size:
Model No.
Dimensions(DN * Length mm)
A50B
L=175, H=55, D=280
2.2 Parameter of Chamber:Design
pressure: -0.1/0.27MPa;
Design temperature:140℃
2.3 Parameter of autoclave:
Max working temperature:137℃
Max working pressure:0.24MPa
Safety valve set-pressure:0.24MPa;Safety valve open-pressure:0.24MPa~0.26MPa
Capacity of clean water tank:1L
Capacity of waste water tank:0.8L
Power input:220V±22V, 50Hz,2900VA
Life span:5 年
2.4 Environmental requirements for use: Environment
temperature: 5℃~40℃;
Relative humidity: ≤85%;
Ambient atmospheric pressure: 80kPa~106kPa;
Water usage requirements as shown in the table below:
Evaporation
residue
Silica
Iron
Cadmium
Lead
Other heavy
metals
Space available

No.:GC-JS-21 Version: 01/00 Effective date:2020-03-30
10
≤10mg/L
≤1mg/L
≤0.2mg/L
≤0.005mg/L
≤0.05mg/L
≤0.1mg/L
Chloride
Phosphate
Conductivity
PH value
Conductivity
hardness
≤2mg/L
≤0.5mg/L
≤15uS/cm
5~7.5
Colorless clean
without
precipitation
≤0.02mmol/L
2.5 Transportation and storage conditions:
Ambient temperature: 5 ~ 40 ℃
Relative humidity: ≤80%
No corrosive gas
Inhaltsverzeichnis
Beliebte Sterilisator Handbücher anderer Marken

Dust Free
Dust Free LightStick Bedienungsanleitung

Euronda
Euronda E9 Next 18L Bedienungsanleitung

UVPLUS
UVPLUS ALS UVP201R Bedienungsanleitung

SterileLight
SterileLight Illumicide SL-UVC-IDC-01-B Bedienungsanleitung

Norka
Norka BUENOS AIRES UV-C 997 100 02 Series Bedienungsanleitung

CertoClav
CertoClav Vacuum Pro 45 Bedienungsanleitung

Chicco
Chicco SterilNatural 2in1 Bedienungsanleitung

Market Forge Industries
Market Forge Industries Sterilmatic Bedienungsanleitung

Philips AVENT
Philips AVENT SCF271/20 Bedienungsanleitung

Easypix
Easypix SteriBox SB1 Bedienungsanleitung

ASP
ASP sterrad NX Bedienungsanleitung

Vital Baby
Vital Baby NURTURE advanced pro UV mini Bedienungsanleitung