Joerns Oxford Pro Digital Weigh-Scale Bedienerhandbuch

User Instruction Manual
Oxford®Low Profile Weigh Scale
To avoid injury, read user manual prior to use. For alternative
languages, contact your authorised service provider.
Manuel de l’utilisateur
Pèse-personne compact Oxford®
Afin d’éviter tout accident, veuillez lire attentivement la notice
avant utilisation. Pour obtenir ces informations dans une autre
langue, veuillez contacter votre prestataire de service agréé.
Manual de Instrucciones
Balanza de Bajo Perfil Oxford®
Para evitar posibles daños, lea previamente el
manual de usuario. Para otros idiomas, póngase en
contacto con su proveedor de servicios autorizado.
Benutzerhandbuch
Oxford® Flache Patientenwaage
Um Verletzungen zu vermeiden, lesen Sie bitte die
Gebrauchsanweisung vor der ersten Benutzung. Für andere
Sprachen wenden Sie sich an Ihren autorisierten Dienstanbieter.

Oxford®
Low Profile Weigh Scale
2
English
Contents
1. Introduction ........................................................................................................ 3
2. Component Parts............................................................................................... 4
3. Assembly ........................................................................................................... 5
4. Operation ........................................................................................................... 7
5. Operation Features............................................................................................ 8
6. Cleaning............................................................................................................. 8
7. Technical Specifications .................................................................................... 9
8. Battery Replacement ....................................................................................... 11
9. Maintenance Inspection and Test.................................................................... 12
10. Warranty .......................................................................................................... 12
11. Logbook for Scale............................................................................................ 13

Oxford®
Low Profile Weigh Scale
3
English
1. Introduction
About your scale:
The Oxford Digital Scale is a technologically advanced, microprocessor controlled, digital
weighing device. It features two large, easy to read, Liquid Crystal Displays (LCD) that show the
patient’s weight from either side.
The controls are simple and straightforward to use. The Oxford Digital Scale is powered by an
inexpensive, common 9 volt battery.
The Oxford Digital Scale can be used in conjunction with the Oxford Presence and Stature. The
model numbers for the scale are:
OXF-UNISCALE100 4-point cradle with scale (calibrated for UK)
OXF-UNISCALE100EU 4-point cradle with scale (calibrated for non UK)
OXF-UNISCALE101 6-point spreader bar with scale (calibrated for UK)
OXF-UNISCALE101EU 6-point spreader bar with scale (calibrated for non UK)
OXF-UNISCALE102 4-point Adaptive power cradle with scale (calibrated for UK)
OXF-UNISCALE102EU 4-point Adaptive power cradle with scale (calibrated for non UK)
The CE mark (Europe Only): The Oxford Digital Scale carries the CE mark and
complies with the following EC Directives:
• Medical Device Directive (MDR) 2017/745
• EMC Directive (2014/30/EU)
• NAWI Directive (2014/31/EU)
• RoHS Directive (2011/65/EU)
Intended Use:
The Oxford Low Prole Weigh Scale is a RoHS compliant precision scale for use in portable patient
weighing systems. The scale is designed to be connected between patient hoist mechanism and
the patient sling apparatus. The Oxford Low Prole Weigh Scale is NOT intended to be used during
patient transportation.
Serious Incident Reporting
In the event of a serious incident taking place during use of this product, affecting the patient and/
or care-giver, it must be reported to the product manufacturer or authorised distributor. Should
the incident take place within the European Union (EU), it must also be reported to the local
competent authority within the member state.

Oxford®
Low Profile Weigh Scale
4
English
2. Component Parts: Digital Scale
The following components are included with your digital weigh scale for the Presence or Stature lift.
1. 1 x Digital Scale with cradle/spreader bar
2. 1 x User Manual
3. 2 x Grey Spacer Disks
Fulcrum Bush
2 Grey Spacers
(Presence lift only)
Fig. 1
Quick Release Pin
Scale Assembly
Fulcrum Spacer

Oxford®
Low Profile Weigh Scale
5
English
3. Assembly
Your weigh scale comes fully integrated with the 6 point (loop) spreader bar, 4-point (non-
powered) cradle or a 4-point adaptive power cradle (powered) cradle.
Attaching the scale/cradle assembly to the lift is quick and easy and will require a small blade
screwdriver.
You need to rst remove your existing spreader bar or cradle from the lift.
Removal of the spreader bar/cradle:
To attach your new Oxford digital scale you may rst need to remove the spreader bar/cradle from
the lift by removing the quick release pin that retains the 4-point cradle or 6-point spreader bar
(Fig. 2). A small at blade screw driver is the only tool required.
Caution: Prior to removing the quick release pin, the cradle needs to be supported by a stationary
object such as a work bench or table. This will prevent the cradle from dropping suddenly
preventing an injury or damage to the cradle.
Fig. 2

Oxford®
Low Profile Weigh Scale
6
English
Connection of the Spreader Bar/Cradle:
To connect, raise the spreader bar/cradle up towards the boom slot. Once aligned in the boom
slot reinsert the quick release pin fully into the fulcrum bush until it locks into position. YOU
MUST CHECK THAT THE PIN IS SECURELY FITTED; a noticeable “click” will sound when the
connection has been made secure.
To do this, press your nger into the barbed end. If the pin remains stationary the lift is safe to use
(Fig. 3).
WARNING
The pin and fulcrum bush must engage through the boom slot. If the fulcrum bush
becomes misaligned this may result in serious injury/death.
Fig. 3

Oxford®
Low Profile Weigh Scale
7
English
4. Operation
The Oxford scale has been designed and certied to weigh up to 400kg in 0.2 increments. It has
been certied to weigh in temperatures between 5˚C/35˚C.
To weigh a patient follow the instructions below:
1. To power up the scale press and release the ON/ZERO button. The display will show either 0.0
or a random measurement.
2. Attach the sling to the spreader bar/cradle that will be used to weigh the patient (this is
necessary so the weight of the sling is not included with the patient’s weight).
3. Press and hold the ON/ZERO button for approximately 2 seconds until “ZERO” is displayed.
The scale will zero and is now ready to weigh a patient.
NOTE: The upper left corner of the display will have the “>0<” symbol. This indicates that the
scale has been properly zeroed. No more than 40kg can be tared to zero.
4. Remove the sling from the spreader bar/cradle and place the sling around/under the patient for
proper lifting.
5. Reattach the sling to the spreader bar/cradle; raise the patient so that they are suspended
from the lift.
NOTE: The patient must not be touching anything such as a chair or bed since this will cause
an error in weighing.
6. Once the digital display has stabilized, record the weight and then lower the patient onto the
appropriate surface (i.e. bed, chair).
NOTE: The upper left corner of the display will have the “>p<” symbol. This indicates that the
weight being measured is stable.
7. The scale has been programmed to turn o after 1 minute to conserve battery life. The zero
(TARE) will be maintained. To restart the scale prior to lowering the patient, push the ON/
ZERO pad momentarily (do not hold for >2 seconds or the scale will be zero under the weight
of the patient). The display will show the start up screen and then show the patient’s weight
NOTE: To restart the scale, be sure to push the ON/ZERO. If the scale is “turned on” and the
pad is pushed again or pressed for more than 2 seconds, the scale may go to zero (0.0) and
the patient’s weight will be lost.
8. The RECALL feature is designed to retain in the memory the last stable weight for a patient
NOTE: Pressing the RECALL button will display the last viewed stable weight, indicated by
the MEM symbol in the upper left corner of the display.

Oxford®
Low Profile Weigh Scale
8
English
5. Operation Features
• All functions are accessible to the user without opening the case.
• Starting or zeroing the scale is done through a single touchpad located on the front of the case.
• Stable readings are achieved through digital averaging. Once the digital display has stabilised,
the stable weight will be held. To make the scale active again, push the ON/ZERO pad
momentarily. The held weight is stored in a non-volatile memory for later recall if required.
• Display turns o after 60 seconds of inactivity to conserve battery life.
• Recall of the last held weight is available by pressing the RECALL pad. This allows a reading of
the last patient’s weight, after the patient has been removed from the scale.
• Battery change is made easy with external access to the battery holder.
6. Cleaning
Clean with ordinary soap and water and/or any hard surface disinfectant. Harsh chemical
cleaners or abrasives should be avoided as these may damage the surface finish of the product.
Avoid wetting any of the electrical parts. After cleaning, the unit should be thoroughly dried.
CAUTION
Do not scratch the display window.
Do not immerse the Weigh Scale in liquid or spray liquid directly on the unit.
Do not use pressurised water or steam on the weigh scale.
End of Life Disposal
Battery: All batteries in the product and any accompanying accessories must be recycled
separately. Batteries should be disposed of in accordance with local or national regulations.
Slings: Slings and associated material accessories should be sorted as combustible waste in
accordance with local or national regulations.

Oxford®
Low Profile Weigh Scale
9
English
7. Technical Specifications
• Electronics..................Microprocessor controlled.
• Power .........................9 volt alkaline battery PP3.
• Battery Life .................Minimum 3000 cycles (60 seconds default)
• Display........................LCD (liquid crystal display) 70 x 30mm screen with 25mm high characters
• Weighing Units ...........Kilograms (kgs).
• Load Cell ....................Rated at 400 kgs.
• Case ...........................ABS
Weights:
• Oxford Low Prole Weigh Scale........................................................................0.9 kgs
• Oxford Low Prole Weigh Scale tted to 6 Point Spreader Bar ........................2.0 kgs
• Oxford Low Prole Weigh Scale tted to 4 Point Cradle ...................................6.8 kgs
• Oxford Low Prole Weigh Scale tted to the APC.............................................9.3 kgs
WARNING
Do not exceed the capacity of the lift. Use above this capacity is unsafe and may result in
serious injury or death.
CAUTION: The digital weigh scale is to be used in a clean, dry, temperature controlled indoor
environment.

Oxford®
Low Profile Weigh Scale
10
English
KEY SYMBOLS:
The following symbols are used on the charger, control unit and battery:
This product is a Medical Device in accordance with EU Medical Device
Regulation 2017/745.
Type B equipment, as per EN 60601-1.
Class 2 equipment.
For indoor use.
ATTENTION, consult accompanying documents.
Refer to the user instructions before use.
WEEE Symbols:
The disposal of batteries should not be mixed with general household waste.
Batteries should be disposed of in accordance with local or national regulations.
The disposal of electronics should not be mixed with general household waste
The crossed-out wheeled bin means that within the European Union the product must be taken to
separate collection at the product end-of-life. This applies to your device but also to any enhancements
marked with this symbol. Do not dispose of these products as unsorted municipal waste.
WEEE Producers registration number WEE/GG0464R2
Inhaltsverzeichnis