Myron L ULTRAPEN PTBT2 Bedienungsanleitung

4 5
PTBT2
S
ACCURATE
S
RELIABLE
S
SIMPLE
Professional Water Analysis
myronl.com
PACKAGE CONTENTS:
PTBT2 Pocket Tester•
Pen - battery installed
Soaker Cap•
Scoop•
Pocket Clip•
Holster•
Lanyard•
Operating Instructions•
MADE IN USAMADE IN USA
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
pH and Temperature Pen
for use with your mobile device
ULTRAPEN™
• Accuracyof±0.01pH
• ProprietarySensorDesign
• 1,2and3pointcalibration
• AutomaticTemperatureCompensation
• TemperatureReadout
• Waterproof
• Poweredby1NTypebattery(included)
• MeasurementandCalibrationstoragewith
time,dateandlocation.
• Exportstoredmeasurementdataviaemail.
Enabled
®
ULTRAPEN™
PTBT3
ORP & Temperature
ULTRAPEN™ PTBT4
Free Chlorine Equivalent
(FC
E
) & Temperature
ULTRAPEN™ PTBT1
Conductivity, TDS, Salinity
& Temperature
Coming soon
Also Available
Available at
Apple App Store
Vi sit our w eb site:
myronl.com/PTBT2.htm

1
INTRODUCTION
Thank you for purchasing the Myron L® ULTRAPEN™ PTBT2 pH Pen. This
instrumentisdesignedtobeextremelyaccurate,fastandsimpletouseindiverse
waterqualityapplications.Advancedfeaturesinclude1,2and3pointcalibration;
automatic temperature compensation; highly stable microprocessor-based
circuitry;waterproofhousing;andMobileDeviceApplicationthatisuserintuitive.
ThePTBT2iseasytocalibrateandeasytouse.
Downloading the Mobile Device App
1. GototheAppleAppStore.
2. Searchforthe“ULTRAPENPTBTx”iPhoneApp.
3. OpentheULTRAPENpageandtaptheInstallbutton.
4. BesuretoreadtheEndUserLicenseAgreement(EULA)andtheDataUseLiability
Agreement(DULA).
Opening the ULTRAPEN App
1. TapULTRAPEN App icononthemobiledevice’sHOMEscreen.
2. TheApp’smainMeasurementscreenwillappear.
NOTE: AppleiOSdevices:ActivateSIRIandsay,“ULTRAPEN.”
FEATURES
PTBT2
1. Push Button -turnsPTBT2on;enterscalibrationmode,activatesPTBT2preference
menus.
2. Battery Cap -providesaccesstobatteryforreplacement.
3. Pocket Clip -holdspentoshirtpocketforsecurestorage.
4. Wireless Transceiver Window -Bluetoothantennaislocatedhere.
5. LED Indicator Light -Tellstheuserwhentodip&swirlPTBT2duringmeasurementsand
calibration.IndicatesPTBT2isinConfigurationmode.
6. pH Sensor-measureshydrogenionconcentrationofsolution.
7. Soaker Cap-ContainspHSensorStorageSolutiontomaintainsensorhydration.To
remove,twistthecapwhilepullingoff.Usecautionnottospillsolution.Toreplace,fill
thecaphalf full ONLYwithsolution.Twistthecapwhilepushingon.Becareful,excess
solutionmaysquirtout.
DoNOTpushthesoakercapbeyondtheCapStopassensordamageWILLoccur.
NOTE:TheformationofKClcrystalsaroundthesoakercapisnormalanddoesnotaffectthe
sensorlife,performance,oraccuracyprovidedtheyarerinsedoffwithwaterpriortoatest.
8. Cap Stop -ShowshowfartopushthesoakercapwhenputtingitonthePTBT2.
9. Scoop-Usedtoholdsamplesolutionwhendippingisnotpossible.Touse,pushthe
scoopontothesensorwhileshiftingside-to-side.andholdthescoopdirectlyundera
verticalstream,avoidingbubbles.Toremove,pullthescoopoffwhileshiftingside-to-side.
IfthesensorisnotfullyseatedinthePTBT2,reinstallperpHSensorReplacementsection
onpage5.
Not shown: LanyardandHolster.
ULTRAPEN App Measurement Screen
1. Mobile Device Status Bar -standardStatusBarforyourmobiledevice.
2. Measurement Value Field -displaysthemeasuredvalueofthesolution.When
theULTRAPENturnsoff,displaysthemessage,“ULTRAPEN is Ofine”andvalues
grayout.
3. Units of Measure -displayspHunits.
4. ULTRAPEN Battery Level -ashesREDwhenULTRAPENbatteryis≤25%.
5. Solution Temperature -measuredtemperatureofthesolution.
6. Memory Store Button -tapheretorecordthemeasurementintheApp’sdatabase.
7. Memory Replace Button -tapheretoreplaceapreviouslystoredmeasurement
withdatafromanewmeasurement.
8. Feature Navigation Bar -thesebuttonsactivatevariousAppfeatures.
CAUTION:

1
OPERATING INSTRUCTIONS
The following are basic instructions for operating the PTBT2 ULTRAPEN &App. To
accessafullOperationsManual:
1. InsidetheApptapthePreferencesbuttonontheFeature
NavigationBar.
2. SelectHelpfromtheSettingsMenuscreen.
3. ThebrowserofyourmobiledeviceswillopenandlinktotheMyronL®Company
websitepagewherethePTBTxOperationsManualcanbeaccessed.
I. Connecting the ULTRAPEN to the Mobile Device
TopairandconnectthePTBT2toyourmobiledevice:
1. InsidetheApptaptheConnectbuttonintheFeatureNavigationBar.
2. PushandreleasethebuttononthePTBT2toturnitON.
3. WhenthePTBT2’snameappearsontheConnectScreen,selectthatline.
4. AcheckmarkwillappearnexttoPTBT2’sname.
NOTE: The default name for all PTBT2’s is, “MLC-PTBTx”.To edit the name, Tap the
Preferences button and go to ULTRAPEN Settings section. See the PTBTx
OperationsManualforfullinstructiions.
TounpairfromaPTBT2:
1. TaptheredUnpairbuttonnexttothePTBT2’sname.
2. AnUnpairInstrumentdialogueboxwillappear.
3. TaptheDisconnectbuttoninthedialoguebox.
II. Temperature Unit Selection
ThePTBT2allowsyoutoselecteitherCentigrade(°C)orFarenheit(°F)whendisplaying
temperatureontheiOSscreens.
ToselecttheTemperatureUnits:
1. InsidetheApptapthe PreferencesbuttonontheFeatureNavigationBar.
2. SelectApplication Preferencesfromthelistofsettingstopics.
3. SlidetheTemperatureUnitsswitchleftorrighttodisplay
temperaturevaluesaseither°Cor°F.
III. pH Mode Selection
ThePT2allowsyoutoselectbetweentwopHmeasurementmodes:
Hold(Default):Oncestabilized,thereadingiscapturedthendisplayed.
LIVE:Real-timereadingsaredisplayedcontinuouslyduringmeasurementuntilthe
ULTRAPENtimesout.
ToselectthepHMode:
1. InsidetheApptap the PreferencesbuttonontheFeature
NavigationBar.
2. SelectULTRAPEN Settingsfromthelistofsettingstopics
3. PushandreleasethebuttononthePTBT2toturnitON.
4 When the PTBT2 LED begins to ash, TheApp will display, “Press and hold the
button on the ULTRAPEN to enter Preferences”
5. PressandholdthePTBT2’spushbutton.
6. WhentheLEDgoestosolidONtheULTRAPENSettingsscreenwillappear.
7. SelectPen ModefromtheULTRAPENSettingsscreen.ThePenModescreenwill
appear.
8. TapthePen Modefieldonthescreen.
9. Choosethedesiredmode(HoldorLive)fromthelistthatappears.
10. Tapthe Applybutton.
IV. pH Measurement
ThefollowingtableexplainstheLEDIndicatorsignalsandtypicaldurations.
LED Indicator Signal Meaning Duration
RapidFlashing Dippeninsolution 6sec
SlowFlashing SwirlPen.Measurementinprocess
InLIVEmode,thereadingsare
dispayed.
10-30secinHoldmode
5minutesinLivemode
SolidONLight Measurementiscomplete
Finalreadingisdisplayed.
≤6sec
To measure solution at the extremes of the specified temperature range,
allowthepen to equilibrateby submerging thecellin the samplesolution of thesame
temperaturefor1minutepriortotakingameasurement.
NOTE:Ifyoucannotdipthepeninthesamplesolution,pourthesampleintoaclean
container.Ifyoudon’thaveasamplecontainerorneedtotestaverticalstreamof
solution,usethescoop.
2
CAUTION:

1
1. Rinsethepen.Ifmeasuringfromacontainer,submergethepenandswirlitaround
inFRESHsamplesolution3times.Alternatively,swirlfor30secondsunder
astreamorinabodyofwater.
2. IftheMeasurementscreenisnotdisplayed,taptheMeasurement
buttonintheFeatureNavigationBar.Thescreenwillsay,
“Paired ULTRAPEN is Ofine“.
3. Graspthepennearthepencaptoavoidsamplecontamination.
4. Removepenfromtherinsesolution.
5. PressandreleasethePTBT2’spushbuttontoturnitON.
6. WhentheLEDashesrapidly,IMMEDIATELYdipthepeninthesamplesolutionso
thatthecelliscompletelysubmerged.Ifyoudonotsubmergethecellinsolution
beforetheashingslows,allowthepentopoweroffandretakethereading.
7. TheAppscreenwilldisplay“Measuring ...”
8. WhentheLEDashesslowly,swirlthepentoremovebubbles,keepingthecell
submergedandpreferablyatleast1inch(2½cm)awayfromsides/bottomof
container.
a.InHoldmodewhentheLEDturnsonsolidthedisplaywillshowthefinalpHand
temperaturereadings.
b.InLIVEmodeallowthepentoremaininsolutionwhiletheLEDashesslowly.
ThedisplaywillshowlivepHandtemperaturereadings.Thepenwilltimeout
after5minute.Pushandreleasethepen’sbuttontopoweritOFFatany
timeduringLIVEmeasurement
9. TheULTRAPENwillturnOFFoncethemeasurementiscompleted.
V. Recording and Recalling Measurements
1. Onceameasurementisfinished,taptheMS button(MemoryStore).
2. TheAppscreenwilldisplaythemeasurementasanunsavedrecord.
3. TaptheSavebuttonintheupperrightcorneroftherecorddisplay.
4. TherecordwillbesavedandtheAppwillreturntothe
Measurementscreen.
5. TaptheMR button(MemoryRecall)intheFeatureNavigationBarto
displayalistofstoredrecords.
6. Swipeupanddowntorevealthefulllist.Tapanyrecordonthelist
toviewit’sdetails.
VI. Calibration
The Myron L® Company recommends calibrating twice a month depending on usage.
However, you should check the calibration whenever measurements are not as
expected.3-pointWetCalibrationismostaccurateandisrecommended.
NOTE:IfthemeasurementisNOTwithincalibrationlimits,“Error. Clean sensor. Check
Solution.”willdisplay.ChecktomakesureyouareusingaproperpHbuffersolution.If
thesolutioniscorrect,cleantheglassbulbofthesensorwithacottonswabsoakedin
isopropylalcohol.Restartcalibration.
NOTE:Smallbubblestrappedinthesensormaygiveafalsecalibration.
NOTE: If at any point during calibration, you do not submerge the sensor in solution
beforetheashingslows,allowthepentopoweroffandstartover.
A. Calibration Preparation
1. Formaximumaccuracy,fill2cleancontainerswitheachpHbufferyouwillbeusing.
Arrangetheminsuchawaythatyoucanclearlyrememberwhichistherinsesolution
and which is the calibration buffer. Always rinse the pH sensor between buffer
solutions.
2. EnsurethepHsensoriscleanandfreeofdebris.
NOTE:Ifyoudonothaveenoughbuffer,youcanuse1containerofeachbufferfor
calibrationand1containerofcleanwaterforallrinsing.
B. Multi-Point Calibration (2 and 3 point Calibration).
Toperformamulti-PointCalibrationyoumustalwaysstartwith7.0pHbuffer.
1. TaptheCALbuttonintheFeatureNavigationBar.
TheAppscreenwillappearandsay,“Paired ULTRAPEN is Ofine”.
2. Graspthepennearthebatterytoavoidsamplecontamination.
3. Removepenfromtherinsesolution.
4. Pressandreleasethepen’spushbutton.TheCalibrationscreenwilldisplay,“Press
and hold the button on the ULTRAPEN to enter CAL mode.”
5. PressandHoldthepen’sbutton.TheLEDwillgotosolidONandtheApp’sCalibration
screenwillappear
6. TapthegreenCALbuttonontheCalibrationscreen.
7. Whenthepen’sLEDashesrapidly,IMMEDIATELYdip
thepenincalibration7.0pHbuffersothatthecellis
completelysubmerged.TheCalibrationscreenwilldisplay.“Calibrating pH ...”
3
Continued on page 4 ...

4 5
8. WhentheLEDashesslowly,swirlthepenaroundtoremovebubbles,keepingthe
cellsubmerged.TheCalibrationscreenwilldisplay.“Calibrating 7.0 pH ...”
9. WhentheLEDlightgoessolidON,theCalibrationscreenwilldisplay
“7.0 ph CAL SAVED”alongwiththebeforeandaftercalibrationvalues.
10. DONEandCONTINUEButtonswillappearatthebottomoftheCalibrationScreen.
NOTE:TaptheDONEbuttontoendthecalibrationatthispoint,or;
11. Rinse the pH sensor with the next pH buffer (it can be either 4.0 pH or 10.0 pH
buffer).
12. TaptheCONTINUEButtontocalibratethenextpoint.
13. Repeatsteps7trough10.Theinstrumentwill
automaticallyrecognizethecalibrationpointbasedonthebufferused.
14. Afterthe3rdcalibrationpointissuccessfullysavedonlythe
DONEbuttonwillappear.Tappingthisbuttonendsthe
calibration.
C. Single-Point Calibration.
1. Usingeither4,7,or10pHbuffer,performthecalibrationasdescribedinsteps1
through9ofSectionB.
2. Afterthecalibrationpointissuccessfullysaved,taptheDONEbutton
D. Factory Calibration
WhenpHbuffersarenotavailable,thePTBT2canbereturnedtofactorydefault
calibrationusingtheFACCALfunction.Thiswillerasestoredwetcalibrationconstants.
NOTE:defaultfactorycalibrationresetstheelectronicsonlyanddoesNOTtakethe
conditionofthesensorintoconsideration.
1. InsidetheApptaptheCALbuttonintheFeatureNavigationBar.TheAppscreen
willsay,“Paired ULTRAPEN is Ofine.”
2. Pressandreleasethepushbutton.Thescreenwilldisplay,“Press and hold the
button on the ULTRAPEN to enter CAL mode”.
3. PressandHoldtheULTRAPEN’sbuttontheLEDwillgotosolidONandtheApp’s
Calibrationscreenwillappear.
4. TaptheredFAC CALbutton.Awarningdialoguebox
willappear.
5. TaptheOKbuttoninthedialoguebox.TheApp’sCalibratingscreenwilldisplay,
“FAC CAL Reset”.
MAINTENANCE
I. Battery Replacement
ThePTBT2Appdisplayhas an indicatorthatdepictsthepenbattery’s charge level.
Whenthechargelevelfallsbelow25%,theindicatorwillashred.Imediatelyreplace
thebatterywithannewNtypebattery.
1. InaCLEAN DRYplaceunscrewthebatterycapinacounter-clockwisemotion.
2. Slidethebatteryandbatteryhousingoutofthepen.
3. Removethedepletedbatteryfromitshousing.
4. Insertanewbatteryintothebatteryhousingorientedwiththenegativeendtouching
thespring.
5. Alignthegroovealongthebatteryhousingwiththeguidebumpinsidethepencase
andslidethebatteryhousingbackin.
6. Screwthebatterycapbackoninaclockwisedirection.
... Continued from page 3
II. Routine Maintenance
1. ALWAYSrinsethepHsensorwithcleanwateraftereachuse.
2. ALWAYSreplacethesoakercaphalffilledwithpHSensorStorageSolutionto
preventthesensorfromdryingoutaftereachuse.
3. Ifthesensorbecomesdirty,cleanthesensorsurfacewithanisopropylsoaked
cottonswab.Thenrinsethoroughlywithcleanwater.
4. Donotdrop,throworotherwisestrikethepen.Thisvoidsthewarranty.
5. Donotstorethepeninalocationwheretheambienttemperaturesexceedits
specifiedOperating/StorageTemperaturelimits.
Do not over tighten.

4 5 PTBT2TFM 08-14
SPECIFICATIONS
MYRON L®COMPANY
2450ImpalaDrive•Carlsbad,Ca92010-7226•Phone:+1-760-438-2021
E-Mail:CustomerService–[email protected]•TechnicalSupport–[email protected]
Website:www.myronl.com
MADEINUSA ©2014 MYRON L® COMPANY
Do not over tighten.
AppRequirements: AppleiOS6oriOS7.OptimizedforiOS7.
MeasurementRange: 0.00–14.00pH;Temperature:0°-71°C/32-160°F
Accuracy(AfterWetCalibration): ±0.01pH;Temperature±0.1ºC/±0.1ºF
Repeatability: ±0.01pH;Temperature±0.1ºC/±0.1ºF
Resolution: 0.01pH;Temperature±0.1ºC/±0.1ºF
TimetoReadingStabilization: 10-30seconds
ActiveModePowerConsumption: ActiveMode137mA,SleepMode2µA
TemperatureCompensation: Automaticto25ºC
PhysicalDimensions: 17.15cmLx1.59cmDor6.75in.Lx.625in.D
Weight: 54gor1.94oz
CaseMaterial: AnodizedAircraftAluminumwithProtectiveCoating
BatteryType: Ntype,Alkaline1.5V
CalibrationStandardSolution: 4.0pH,7.0pH,10.0pH
Operating/StorageTemperature: 0-55ºC/32-131ºF
WaterResistance: IP67andNEMA6
EN61236-1:2006-AnnexA:
2008;1999/5/ECAnnexIII
Electrostaticdischargetocaseofinstrumentmay
causePTBT2tospontaneouslypoweron.Inthiscase,
thePTBT2willpoweroffafterseveralseconds
FCCID:T7VPAN17 TheBluetoothtransceiverdevicemeetstherequirementsfor
modulartransmitterapprovalasdetailedinFCCpublicNotice
DA00-1407.
Canada(IC),license:
IC:216Q-PAN17
TheBluetoothtransceiverdevicemeetstherequirementsfor
modulartransmitterapprovalasdetailedinRSS-GEN.
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
QUICK REFERENCE INSTRUCTIONS
1. TurnonULTRAPENApp:TapULTRAPENicononmobiledevice’shomescreen.
2. TurnonPTBT2:Pressandreleasepushbuttononpen.
3. LEDashesrapidly:Dippeninsamplesolutionsocellistotallysubmerged.
4. LEDashesslowly:Swirlpentoremovebubbles,keepcellsubmerged.
5. LEDturnssolidONthenturnsOFF:Measurementiscompleteandreadingisdisplayed.
Feature Navigation Bar
ThisbuttonalwaysreturnstheApptotheMeasurementScreen.
Thisbuttondisplaysalistofsavedmeasurements.
ThisbuttonplacestheAppinCalibrationmode.
ThisbuttonopenstheApp’sBluetoothConnect/Disconnectscreen.
ThisbuttonopenstheApp’spreferencesandsettingsmode.
AninternetlinktofullOperationsManual(Helpbutton)isfoundhere.
ACCESSORIES
STANDARDSOLUTIONSFORCALIBRATION:TheULTRAPEN™PT2requirespH4,pH7,andpH10
buffersolutionsforwetcalibrationandpHSensorStorageSolutionforproperstorage.
— Order MODEL#s: PH4, PH7, PH10, and PHSS.
ReplacementpHSensor(withinstructions)— Order Model: RPT2.
III. pH Sensor Replacement (Follow the instructions that come with your Replacement Sensor.)
OnlyRemove/ReplacethepHsensorinaCLEANandDRYenvironment!
ToremovethepHsensor:
1. Removethesoakercap;makesurethePTBT2(includingthepHsensor)isclean
anddry.Loosenthebatterytraytoallowpressureequalization.
2. ThenfirmlygraspthepHsensorbodyandslowlypullthepHsensorout.
ToinstallanewpHsensor:
1. LineupthealignmenttabsonthepHsensorwiththealignmentslotsonthePT2
unit.
2. GentlypushthepHsensorintoposition,closebatterycap.
CAUTION:
MYRON L®WARRANTY
Exceptingsensorandbattery,theMyronL®CompanyPTBT2pHULTRAPENiswar-
rantedtobefreefrommanufacturedefectfor1year.ThepHsencerhasasix(6)month
limittedwarrenty.
Andere Handbücher für ULTRAPEN PTBT2
2
Dieses Handbuch passt für folgende Modelle
3
Inhaltsverzeichnis
Beliebte Temperaturregler Handbücher anderer Marken

Swegon
Swegon LUNAd MB Bedienungsanleitung

Lucky Reptile
Lucky Reptile Thermo Control II Bedienungsanleitung

SENSORMETRIX
SENSORMETRIX Argon 100 Bedienungsanleitung

Fuji Electric
Fuji Electric PXE4 Bedienungsanleitung

Yuyao gongyi
Yuyao gongyi XMT-808 series Bedienungsanleitung

Selec
Selec TC544C Bedienungsanleitung