NuFlo Barton 202E Bedienungsanleitung

BarBar
BarBar
Bartonton
tonton
ton®®
®®
®
MODEL 202E
DIFFERENTIAL PRESSURE RECORDERS
User Manual
Manual No. 10060, Rev. B
05/2004
CONTENTS
Section 1 - Introduction ................................................................. Page 1
Pressure Elements/Chart Drive Tables ..................................... Page 5/6
Section - Installation/Calibration Check ................................... Page 7
Section 3 - Maintenance & Calibration ........................................ Page 13
Section 4 - Troubleshooting ........................................................... Page 6
Section 5 - Installation Drawings .................................................. Page 8
Section 6 - Parts Drawing/List ....................................................... Page 34
Specification (DPU/Instrument)................................................... Page 39
SECTION 1 - INTRODUCTION
General
The Barton® Model 202E single DP Flow Recorders measure and record system
operation parameters including gas and liquid flow rate liquid level in a vessel
and other system variables that can be measured by differential pressure and
static pressure methods.
Main Components
Differential Pressure Unit
The Barton® Model 199 Differential Pressure Unit (DPU) actuates the recorders.
The DPU is a dual bellows assembly enclosed within pressure housings. The
bellows are liquid filled and withstand overranges equal to the working pressure
of the housing without calibration change. Movement of the bellows is transmit-
ted through a torque tube from the DPU to the recording mechanism. The
torque tube is hermetically sealed to eliminate friction and leakage; it does not
require lubrication.
Before installing this instrument become familiar with the installation instructions
in Section 2. WARNING notes that appear on the following pages of this manual
should be reviewed before proceeding: 8 20 21 22 and 25.
SAFETY !
WARNING notes indicate the presence of a hazard which can cause severe personal injury, death, or substantial
property damage if warning is ignored
CAUTION notes indicate the presence of a hazard which can cause moderate personal injury or property
damage if warning is ignored

4HE$05ISATTACHEDTOTHEBACKOFTHERECORDERCASE4HETORQUETUBEIS
INSERTEDTHROUGHAHOLEINTHECASEANDCONNECTEDTOTHE$0DRIVEARMOFTHE
RECORDERMECHANISM-OVEMENTOFTHEBELLOWSISTRANSMITTEDBYTHE$05
DRIVEARMTOTHETORQUETUBEASAROTARYMOTIONWHICHINTURNROTATES
THERECORDERMECHANISM4HEPRESSUREHOUSINGSOFTHE$05ARECONNECTEDBY
PIPEORTUBINGTOTHEHIGHANDLOWPRESSURESIDESOFTHEPRIMARYDEVICELOCATED
INTHEPROCESSSYSTEM
4HERANGEOFTHE$05ISDETERMINEDBYTHEFORCEREQUIREDTOMOVETHEBELLOWS
THROUGHTHEIRNORMALRANGEOFTRAVEL4HE2ANGE3PRINGSWHICHAREAVAILABLE
INVARIOUSRANGESBALANCETHE$0APPLIEDTOTHEUNIT4HENUMBEROFSPRINGS
ANDTHEIRRATINGSDEPENDSONINDIVIDUAL$0RANGEREQUIREMENTS3EE"5!PHOTOS
ONPAGE
4HEBUILTIN0ULSATION$AMPENERCONTROLSTHEmOWOFlLLLIQUIDBETWEENTHE
HIGHANDLOWPRESSUREBELLOWSWITHANEXTERNALLYADJUSTABLENEEDLEVALVE)N
APPLICATIONSWHEREPULSATIONISNOTAPROBLEMTHENEEDLEISSETTOTHEFULLOPEN
POSITION

!DJUSTING0ULSATION$AMPENER
2EMOVEDUSTCOVERANDINSERTHEXKEYINTO
NEEDLEVALVE
4URNVALVECLOCKWISETOCLOSEDPOSITIONAPPROX
TURNS
"ACKOUTVALVETURNORASREQUIREDTO
REDUCEPULSATIONSORSHOCKPRESSURES
2ECORDING-ECHANISM
4HERECORDINGMECHANISMISALINKAGEANDPEN
SYSTEMTHATPERMANENTLYRECORDSDATA)TCONVERTS
MECHANICALINPUTSFROMTHE$05ANDSTATICPRESSUREELEMENTTOTRANSCRIBELINES
ONAREVOLVINGCHART!LLOPERATIVEPARTSOFTHERECORDERMECHANISMAREMADEOF
STAINLESSSTEELFORALONGERlELDLIFE4HEPENMOUNTISEXCEPTIONALLYRUGGED!LL
LINESAREADJUSTABLE3CREWADJUSTMENTSFORZERORANGEANDLINEARITYASSUREFAST
ANDACCURATECALIBRATION4HE$00ENMECHANISMISSHOWNBELOW

3TATIC0RESSURE%LEMENT
4HISELEMENTMEASURESSTATICPRESSUREINAPIPINGSYSTEMFROMTOINCHESOF
(GMERCURYVACUUMORFROMTOPSI4HESENSORELEMENTISABOURDON
TUBECONSISTINGOFASLIGHTLYmATTENEDCROSSSECTIONOFTUBINGCOILEDINTOAHELIX
ORmATSPIRAL4HEOUTERENDOFTHETUBINGISSEALEDANDATTACHEDTOADRIVEARM
WHICHISATTACHEDTOTHESTATICPRESSURECONNECTIONTUBING
4HE3TATIC0ENMECHANISMISSHOWNBELOW
4HESTATICPRESSURECONNECTIONISLOCATEDONTHEBACKOFTHERECORDERCASE3TATIC
PRESSUREINTRODUCEDTHROUGHTHETUBINGCAUSESTHESPIRALORHELIXTOUNWIND!
REDUCTIONOFPRESSUREWITHINTHETUBINGCAUSESTHEELEMENTTOWINDMORETIGHTLY
3ENSORELEMENTSAREAVAILABLEINVARIOUSALLOYSFORAVARIETYOFPRESSURERANGES
3EE3TATIC0RESSURE%LEMENT4ABLEONNEXTPAGE

3TATIC0RESSURE%LEMENT4ABLE

#HART$RIVE
3PIRALWOUNDCHARTDRIVESELECTRICALORMECHANICALCANBESELECTEDTOlTAWIDE
VARIETYOFCHARTSPEEDSORTIMEINTERVALSTOREDUCEMAINTENANCETIME!LLCHART
DRIVESAREINTERCHANGEABLEANDlTTEDWITHAmOWRECORDERHUBCLIPWHICHLOCKS
THECHARTINPLACE%XPLOSIONPROOFELECTRICALDRIVESAREALSOAVAILABLE
"ELLOWS5NIT!SSEMBLY"5! $RIVE!RM
(OLE0LUG
$O.OT,OOSEN
2ANGE3PRINGS
2ANGE3PRINGS
,OCK.UT
0USH2OD
2ETAINER3CREWS
4ORQUE4UBE
'LAND.UT
$O.OT,OOSEN
4ORQUE4UBE
'LAND.UT
$O.OT,OOSEN
4ORQUE4UBE
3HAFT
4ORQUE4UBE
3HAFT
2ANGE3PRING
!SSEMBLY
,OCK.UT
0USH2OD
3PRING
%ND#AP
2ANGE3PRING
!SSEMBLY
2ANGE3PRING
0OST
"5!
"5!

3%#4)/.).34!,,!4)/.#!,)"2!4)/.#(%#+
'ENERAL
4HEINSTRUMENTSHOULDBEINSPECTEDATTIMEOFUNPACKINGTODETECTANYDAMAGE
THATMAYHAVEOCCURREDDURINGSHIPMENT
3TANDARD0RACTICE#ONSIDERATIONS
4HEFOLLOWINGPRACTICESSHOULDBEOBSERVEDUPONINSTALLATION
4HEDISTANCESBETWEENTHEPRIMARYDEVICEANDTHE$05SHOULDBEMINI
MIZED$ISTANCESEXCEEDINGFEETARENOTRECOMMENDED&ORDISTANCES
UPTOFEETUSEINCHORINCHPIPEORTUBING&ORRUNSOFTO
FEETUSEINCHPIPEORTUBING4HERECOMMENDEDLIMITATIONDOES
NOTAPPLYWHENANAIRPURGEORBLOWBACKSYSTEMISUSED
3LOPEALLPIPINGATLEASTINCHLINEARFOOTTOAVOIDLIQUIDORGASENTRAPMENT
0ROVIDETWOFEETOFUNINSULATEDPIPINGBETWEENTHE$05ANDTHEPRIMARY
DEVICEFOREACH&#INEXCESSOF&#
!SSURETHATTHETEMPERATUREOFTHE$05NEVEREXCEEDS&#
7HENSTEAMTRACINGISNECESSARYTHESTEAMPRESSURESHOULDNOTEXCEEDlVE
POUNDSPERSQUAREINCHANDINSULATIONSHOULDNOTBEUSED)FPRESSUREMUST
EXCEEDlVEPOUNDSPERSQUAREINCHLIMITTHELENGTHOFTUBINGAROUNDTHE
$05TOTWOTURNSANDDONOTINSULATE
)NSTALLASUITABLEPULSATIONDAMPENINGDEVICEUPSTREAMOF$053EVERE
PULSATIONWILLAFFECTTHEACCURACYOFmOWMEASUREMENT!VOIDSEVERE
PULSATINGmOWSSUCHASRECIPROCATINGPUMPSORCOMPRESSORS
-OUNTTHE$05ONASOLIDSUPPORTTOMINIMIZEVIBRATION4IGHTENALLPOINTS
USINGASUITABLECOMPOUND,EAKSINPIPINGCANCAUSEMEASUREMENTERRORS
2OTATETHEHOUSINGASNECESSARYTOPLACETHECONNECTIONINTHEPROPERPOSI
TION4HE$05HASCONNECTIONSINTHEPRESSUREHOUSINGSTOACCOMMODATE
VARIOUSPIPESIZES3EE3ECTION
)NSTALLAVALVEMANIFOLDCONNECTINGTHE$05ANDTHESOURCEOFDIFFERENTIAL
PRESSURETOFACILITATEOPERATIONANDCHECKINGOFTHE$05
,OCATEALLSHUTOFFANDBYPASSVALVESSOTHEYAREACCESSIBLEFROMTHEFRONTOF
THEINSTRUMENT,OCATEBLOCKVALVESATTHESOURCEOFDIFFERENTIALPRESSURE
!PPLICATION#ONSIDERATIONS
&LOW-AKESURETHE$05HIGHPRESSUREHOUSINGISCONNECTEDTOTHE
UPSTREAMTAPOFTHEPRIMARYDEVICE.OTE4OPREVENTOVERHEATINGTHE$05
DURINGBLOWDOWNTHEOPERATORSHOULDMONITORTHETEMPERATUREBYPLACINGA
HANDONTHEPIPEBETWEENTHE$05ANDTHEMANIFOLDPIPECONTAININGTHEVENT
VALVES
,IQUID,EVEL4HEPROCESSMEDIACANBEUSEDASAREFERENCELEGSEALmUID
WHENITISOFATYPETHATWILLCONDENSEINTHEREFERENCELEGUNDERALLCONDITIONS
)FPROCESSORPROCESSMEDIACHARACTERISTICSPREVENTMEETINGTHEABOVEREQUIRE
MENTSASPECIALREFERENCELEGSEALmUIDWILLBEREQUIRED4HESPECIALSEALmUID
MUSTNOTBEVOLATILEANDMUSTNOTBEMISCIBLEWITHTHEPROCESSMEDIA!LSOTHE
DIFFERENCEINTHEDENSITIESOFTHESPECIALSEALmUIDANDTHEPROCESSMEDIAWILL
REQUIRECOMPENSATIONINCALCULATINGTHE$0RANGEOFTHE$05

7!2.).'
%XPLOSION(AZARD.O/RGANICCOMPOUNDSOILGREASEDIRTOR
SCALEOFANYKINDCANBETOLERATEDINANOXYGENINSTALLATION
#ALIBRATION#HECK
.OTE4HERESHOULDBEENOUGHPENPRESSURETOPRODUCEACONTINUOUSLINE
4HEPENSHOULDSPRINGFREEOFCHARTWHENCHARTISPUSHEDINWARDABOUT
INCH)FNECESSARYADJUSTPENPRESSUREBYBENDINGPENARMINTHE
REQUIREDDIRECTION
! $IFFERENTIAL0RESSURE0EN
4OADJUSTTHEINSTRUMENTTOFACTORYSETCALIBRATIONLEVELS
6ERIFYTHEINSTRUMENTISPOSITIONEDAPPROXIMATELYLEVELANDISSECURELY
MOUNTED
#ONNECTTHEHIGHPRESSUREHOUSINGINTOTHECALIBRATIONSETUP

6ENTTHE$05LOWPRESSURESIDETOATMOSPHEREBYREMOVINGONEVENT
PLUGFROMTHELOWPRESSUREHOUSING
2EMOVETHEDUSTCAPOFTHEPULSATIONDAMPENERVALVEUSINGAREGULAR
SCREWDRIVER/PENTHEDAMPENERVALVEBYTURNINGITFULLYCOUNTERCLOCK
WISEUNTILITSTOPSTURNINGUSINGAINCH!LLENWRENCH
0LACEARECORDINGCHARTONTOTHECHARTPLATE,OCKTHECHARTINTO
POSITIONWITHTHEHUBCLIP
3ETTHEPENPRECISELYATZEROUSINGTHEPENZEROADJUSTMENTSCREW
!PPLYPRESSURE6ERIFYTHATTHEPENMOVESACROSSSCALETOTHE
PRESSUREINDICATION
!PPLYPRESSURE6ERIFYTHATTHEPENINDICATESPRESSUREON
THECHART
)FTHEPENDOESNOTACCURATELYINDICATETHEPRESSUREBEINGAPPLIED
FURTHERADJUSTMENTSARENECESSARYPERFORMTHECALIBRATIONPROCEDURE
OUTLINEDIN$00EN#ALIBRATIONONPAGE
" 3TATIC0RESSURE0EN
)FTHEINSTRUMENTINCLUDESASTATICPRESSUREELEMENTUSETHEFOLLOWINGTO
CHECKTHESTATICCALIBRATION
#ONNECTTHESTATICPRESSUREELEMENTINTOTHECALIBRATIONSETUP
2ELEASEPRESSURE3ETTHEPENTOZEROUSINGTHEZEROADJUSTMENTSCREW
!PPLYPRESSURE6ERIFYTHATTHE30PENMOVESACROSSSCALETOTHE
PRESSUREINDICATION
!PPLYPRESSURE6ERIFYTHATTHE30PENINDICATESPRESSURE
ONTHECHART
)FTHEPENDOESNOTACCURATELYINDICATETHEPRESSUREBEINGAPPLIED
FURTHERADJUSTMENTSARENECESSARYPERFORMTHECALIBRATIONPROCEDURE
OUTLINEDIN300EN#ALIBRATIONONPAGE

2ECORDER-OUNTING
.OTES
"EFOREINSTALLATIONPERFORMACALIBRATIONCHECKPAGE$ONOTPLACE
UNITNEARVENTSORBLEEDHOLESTHATDISCHARGECORROSIVEVAPORSORGASES
$ONOTAPPLYWRENCHBARPRESSURETORECORDERCASE!LLlTTINGSMUSTBE
LEAKTIGHT!LLTUBINGMUSTBESECURETOMINIMIZEVIBRATION-OUNTTHE
INSTRUMENTASLEVELASPOSSIBLE
&LUSH-OUNTING
A #UTOPENINGINPANELTOTHEDIMENSIONSSHOWNIN3ECTION
B $RILLOUTPILOTHOLESLOCATEDONTOPANDBOTTOMOFCASE5SEA
.ODRILL,IMITDRILLPENETRATIONANDREMOVECHIPS
C !TTACHMOUNTINGBRACKETSTOCASEBOTTOMWITHSUPPLIEDSELFTAPPINGSCREWS
D 3LIDECASEFROMFRONTOFPANELINTOTHEPANELCUTOUT
E !TTACHREMAININGMOUNTINGBRACKETANDINSTALLPANELMOUNTINGSCREWS
0IPE-OUNTING
A 0LACEASUITABLELENGTHOFINCHPIPEINTOAWELLSECUREDmOORORWALL
mANGEORATTACHTHEINCHPIPETOEXISTINGPIPEWITHASADDLETHATIS
lTTEDWITHAINCH.04OUTLET
B 4HREADTHE$05ONTOTHEPIPEORIENTTHEINSTRUMENTWITHWRENCHOR
BARAPPLIEDTOTHE$05BRACKETONLYANDTIGHTENRETAININGSCREWS
0IPING
#ONNECTTHEPRESSUREELEMENTTOSYSTEMPRESSUREUSINGTUBINGORPIPE0RESSURE
CONNECTIONONOUTSIDEOFRECORDERCASEISINCH.04FEMALE
%LECTRICAL#ONNECTIONS
4WOKNOCKOUTSAREONTHEBOTTOMOFTHECASE/NEKNOCKOUTISONTHEBACKOF
THECASETOACCOMMODATECONDUITCLAMPINGCONNECTIONSTHROUGHWHICHWIRING
ISFEDTOTHETERMINALSTRIP4HETERMINALSTRIPISBELOWTHECHARTDRIVEMOTOR
4HESTANDARDELECTRICALPOWERSUPPLYREQUIREMENTIS6!#(Z(Z
WIRINGAVAILABLEUPONREQUEST
#HART)NSTALLATION
0ERFORMTHEFOLLOWINGSTEPSTOINSTALLTHECHART
/PENTHERECORDERDOORANDRELEASETHECHARTLOCKONCHARTDRIVEHUB
2AISETHEPENLIFTERARMANDSLIDETHECHARTBETWEENTHEPENSANDTHECHART
PLATE)NSERTTHECHARTINTHECHARTGUIDESINTHECHARTPLATEANDLOCATETHE
HOLEINTHECHARTONTOTHECHARTHUB0LACETHEPRESSURECHARTONTOTHECHART
HUB4HENLOWERTHEPENLIFTERARMANDPOSITIONTHECHARTTOPLACETHE
PENSONTHEDESIREDCHARTTIMELINE
3ECURETHECHARTINPLACEWITHTHECHARTHUBLOCK
3TARTUP
.OTE "ECAREFULNOTTOSUBJECTTHE$05TOUNNECESSARYSHOCKOROVERRANGE
PRESSUREDURINGOPERATIONS
!LWAYSSTARTWITHTHEBLOCKVALVESCLOSED.OTE$ONOTPERFORMAZERO
CHECKWITHONLYONEBLOCKVALVECLOSED)FSEALPOTSAREUSEDSOMEOFTHE
SEALINGmUIDWILLBELOSTORDISPLACED)NGASSERVICEASTANDARDWAVEEFFECT
MAYOCCURTHATWILLDISPLACETHEZEROREADING
Inhaltsverzeichnis