Peak Corona Bedienungsanleitung

Corona Air Compressor
User Manual
Copyright © Peak Scientific Instruments Ltd 2018 — Corona Air Compressor User Manual Rev 4 PN UM-0017 RSID 152 EN —23/03/2018

Contents
Change History 3
How to use this Manual 3
Introduction 4
Warranties and Liabilities 5
Safety Notices 6
Symbols 6
Safety Notice to Users 6
Declaration of Conformity 7
Environmental Declaration 8
Technical Specification 9
Fittings Kit Contents 11
Unpacking Instructions 11
Installation 12
Compressor Environment 12
Compressor Overview 13
General Dimensions 13
Rear Connections 14
Electrical Connection 15
Connecting to the application 16
Tubing Lengths 16
Normal Operation 17
On Demand Gas 17
Unusual Operation 17
Service Requirements 18
Service Schedule 18
Peak Protected 19
Cleaning 20
Troubleshooting 21
Page 2

Page 3
Change History
Rev Comment Name Date
4 Service Kit Ordering Update Liam Couttie 23/03/2018
How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your air compressor.

Page 4
Introduction
The Corona Compressed Air Generator is an external air compressor solution designed
to supply compressed air to the Peak Corona Nitrogen generator for CAD applications.
Features include:
• Compact, stackable design
• Also stackable with Corona Veo CAD
• Product developed in conjunction with Thermo
• Status indicative lighting feature
To ensure this air compressor model meets our high expectations with regards
to reliability and performance, we have tested this new model extensively at our
manufacturing plant and with end users around the world to ensure reliability and
longevity of the system.

Page 5
Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods
shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free
from defects in material and workmanship for the lesser of a period of twelve
months from the date of delivery or thirteen months from the date of dispatch
from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person
dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common
law are hereby expressly excluded save to the extent they may not be lawfully
excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of
this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of
clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed,
commissioned, stored, serviced and used and without prejudice to the generality
of the foregoing that any defect is not the direct or indirect result of lack of repair
and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or
incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any relevant
packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in sub-
clause 2 hereof within a reasonable time of it becoming apparent and subject always
to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or
repair the defective Goods or refund a proportionate part of the Price. The Company
shall have no further liability to the Customer (save as mentioned in sub-clause 6
hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim
for death or personal injury to any person in so far as such is attributable to the
negligence or breach of duty of the Company or any failure by the Company to
comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect
of any claim by the customer for costs, damages, loss or expenses (whether direct,
indirect, consequential or otherwise) or indemnity in any respect howsoever arising
including, but not by way of limitation, liability arising in negligence (other than
pursuant to clause 6 above) that may be suffered by the Customer or any third
party.

WARNING
WARNING
WARNING
WARNING
CAUTION
Page 6
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this manual refer to the most
likely potential hazards, but by definition cannot be all inclusive. If the user employs
an operating procedure, item of equipment or a method of working which is not
specifically recommended by Peak Scientific, the user must ensure that the equipment
will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe
and proper use of the air compressor.
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the air compressor or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the air compressor has been
removed before proceeding.
Safety Notice to Users
These instructions must be read thoroughly and understood before
installation and operation of your Peak Corona Air Compressor. Use of
the air compressor in a manner not specified by Peak Scientific MAY
impair the SAFETY provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.

Page 7
Declaration of Conformity
We Peak Scientific Instruments Ltd.
Of Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Declare that:
Equipment: Air Compressor
Models: Corona Air Compressor 110V & 230V
To which this declaration relates, is in conformity with the applicable EC directives,
harmonized standards, and other normative requirements.
• Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Electrical Equipment for measurement, control and laboratory use.
• Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for measurement, control and laboratory use.
• FCC47 CFR Part 15 Class B
Unintentional radiators; Conductec and Radiated emissions limits.
All evaluation, testing and certification issued by:
York EMC Services Ltd.
Earls Road, Grangemouth Technology Park
Grangemouth
KY11 9HZ
Signed: Name: Chris Pugh
Date: 02/09/2016 Position: Engineering & Operations Director

Page 8
Environmental Declaration
We Peak Scientific Instruments Ltd.
Of Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Declare that:
Equipment: Air Compressor
Models: Corona Air Compressor
Is fully compliant with the following Directives
• 2012/19/EU WEEE (Waste of Electrical and Electronic Equipment)
• 2011/65/EU RoHS2 (Restriction of Hazardous Substance)
Peak Scientific Instruments Ltd fully complies with its obligations towards the European
WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These
obligations are being met within the B2B compliance group.
Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to
ensure that our products comply with the principles and requirements of the European
recast RoHS2 (Restriction of Hazardous Substances) Directive 2011/65/EU. Similar
directives in the United States and China, for example, have also been captured within
this program.
Where a specific certificate of compliance is required, this can be requested, on
a product serial number basis, directly from Peak Scientific Instruments Ltd, by
contacting us through our website on www.peakscientific.com
Signed: Name: Chris Pugh
Date: 02/09/2016 Position: Engineering & Operations Director

Page 9
Technical Specification
Corona Air Compressor
Environment
Minimum Operating Ambient Temperature 5°C (41°F)
Maximum Operating Ambient Temperature 35°C (95°F)
Maximum Altitude 2000 m
Maximum Relative Humidity 80% Non-Condensing
Minimum Storage Temperature -20°C (-4°F)
Maximum Storage Temperature 60°C (140°F)
Compressor Outlets
Maximum Flow Rate 30 l/min
Maximum Pressure 7.58 bar (110 psi)
Particles <0.01µm
Phthalates None
Suspended Liquids None
Gas Outlets 1 x ¼” BSPP
Drain Outlets 1 x ¼” BSPP
Pressure Gauges 1
Electrical Requirements
Voltage 110VAC 230VAC
Frequency 60 Hz 50/60 Hz
Current 5.3A 2.8A
Input Connection C14 Plug
Power Cord (Supplied) C13 Socket to Local Connection
Pollution Degree 2
Insulation Category II
General
Dimensions cm (inches) WxDxH 41.6 (16.3) x 54.0 (21.2) x 40.6 (15.8)
Compressor Weight Kg (lbs) 41.5 (91.5)
Shipping Weight Kg (lbs) 64 (141)

Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking.
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require
two people to remove the unit from the shipping crate and to manoeuvre the air
compressor onto the bench.
Please save the product packaging for storage or future shipment of the air compressor.
Note: Included with the air compressor is a “Fittings Kit” containing mains power leads
for UK, EU & US and also all the required fittings and warranty registration card. Be
careful not to discard these with the packaging.
Page 10Page 10
Inhaltsverzeichnis