V20160909
Focal plane change for
every 1mm of secondary
spacing change
Allowable primary-to-secondary
spacing tolerance to keep focal
plane within 5mm of optimal
Figure 9 shows a Ronchi simulation with the Ronchi screen set at the focal plane and also
with the focal plane 10mm away from the Ronchi screen. The image on the left is what you
will see if the primary-to-secondary spacing is perfect. The image on the right is what you
will see if the primary-to-secondary spacing is off by about 2mm. Remember, you want
better than 1mm spacing accuracy for the CDK17.
Adjusting the Secondary Spacing
The CDK secondary mirror is not spring loaded, so you must always keep tension in the
collimating screws and the central bolt. To move the secondary toward the primary mirror:
1. Loosen three screws (four screws for the CDK12.5) . Loosen them in equal amounts
to maintain your collimation.
2. Hold the secondary housing with one hand. This is to keep the housing from rotating.
(The collimation screws sit in shallow recesses and you would like them to go back in
the recesses when the procedure is complete).
3. Take a flat-head screwdriver and place the tip in the central bolt of the secondary
assembly. Rotate the central bolt in a clockwise direction. For reference, rotating
¼ turn moves the secondary .4mm, which moves the focal plane 1.76mm in the
case of the CDK17.
4. Turn the collimation knobs in equal amounts until the assembly is tight again.
5. Check the Ronchi grating and repeat this process as necessary until you get a null.
To move the secondary away from the primary:
1. Loosen the collimation screws. Loosen them in equal amounts to maintain your
collimation.
2. Hold the secondary housing with one hand. This is to keep the housing from rotating.
(The collimation screws sit in shallow recesses and you would like them to go back in
the recesses when the procedure is complete).
3. Take a flat-head screwdriver and place the tip in the central bolt of the secondary
assembly. Rotate the central bolt in a counter-clockwise direction. For reference,
rotating ¼ turn moves the secondary .4mm, which moves the focal plane 1.76mm
in the case of the CDK17.
4. Rotate the four collimation knobs in equal amounts until the assembly is tight again.
5. Check the Ronchi screen and repeat this process as necessary until you get a null.
Re-checking Collimation
After you are satisfied that the secondary spacing is correct, replace the Ronchi ocular with
the low power eyepiece and repeat step 1 to verify that the collimation is still close. If re-
collimation adjustments are necessary, make them and repeat step 3.