Preeflow eco-PEN450 Leitfaden

Dosing system
Commissioning and
maintenance manual
eco-PEN450


eco-PEN450
Commissioning and maintenance manual
Copyright © Version 1.8 3 / 15
Contents
1This manual 5
2The dispenser 5
3Safety 5
4Starting up for the first time 6
4.1 Inserting the stator 6
4.1.1 Connecting the dosing unit to the drive unit 7
4.1.2 Supply and bleed the medium. 7
5Connection options 9
6Cleaning 10
7Spare parts 11
8Maintenance 12
8.1 Changing the stator 12
8.2 Troubleshooting 12
9Disposal 12
10 Technical specifications 13

eco-PEN450
Commissioning and maintenance manual
4 / 15 Version 1.8 Copyright ©

eco-PEN450
Commissioning and maintenance manual
Copyright © Version 1.8 5 / 15
1This manual
Dear customer,
This manual forms part of the operating and maintenance manual for the dosing system
eco-CONTROL EC200. The operating and maintenance manual must have been read
and understood in full before commissioning or carrying out maintenance work on the
eco-PEN450.
2The dispenser
The dispenser is supplied with these components and accessories:
3Safety
All instructions in the Safety section of the maintenance and operating manual for the
dosing system must be complied with and implemented.
fig. 1
A Dosing unit
B Stator
C Drive unit
D Assembly aid
E Installation tool
(2 Allen wrenches)
F Spraying device (without fig.)

eco-PEN450
Commissioning and maintenance manual
6 / 15 Version 1.8 Copyright ©
4Starting up for the first time
4.1 Inserting the stator
Preparation: Unplug the power supply to the drive unit.
1. Undo the union ring (2).
2. Pull off the end piece (3).
3. Couple the assembly aid (6) to the dispenser housing (10)
(the star-shaped coupling (5) must be attached to the dispenser housing).
4. Wet the rotor (4) with the medium or a suitable lubricant.
5. Turn the stator (7) (on the bearing side, see Fig. A) on the rotor (4) until the dowel pin
begins to dip into the keyway (9).
6. Lightly press the stator in the direction of the dispenser housing (10) and turn the
assembly aid in the direction of the arrow until the stator has been guided into the
dispenser housing.
7. Uncouple the assembly aid, install the end piece and the union ring, and put in place
the required Luer-Lock needle.
Dispenser Do not switch on until medium has been delivered to it. Otherwise there is a
risk of damage to the equipment. Even a brief period of dry running can lead to the
stator being destroyed.
Fig. A
Medium
fig. 2

eco-PEN450
Commissioning and maintenance manual
Copyright © Version 1.8 7 / 15
4.1.1 Connecting the dosing unit to the drive unit
1. Turn the set screws (14) in the thread such that they do not protrude into the coupling
area. Danger of damage to the fit.
2. Attach the star-shaped coupling (5) onto the coupling of the drive unit( 11).
3. Couple the drive unit (11) to the dosing unit (20) until there is a gap <1 mm between
the anti-rotation lock (8) and the dosing unit (20).
4. Set the anti-rotation lock (8) correctly in place by rotating the dosing unit (20).
5. Bring the drive unit (11) and the dosing (20) together completely.
6. Lightly turn the set screws (14), the drive unit (11) has now been centred properly.
4.1.2 Supply and bleed the medium.
• Connect the tank (supply line, cartridge) of the medium to the dispenser medium input
(12). (See section 10, Technical specifications,, page 13 for details of the thread)
The Dispenser can be bled in one of two ways after the medium has been supplied.
Follow the safety stipulations and instructions of the manufacturer of the medium
to be used to fill the unit. If applicable, use the required protective equipment. If
the medium is being conveyed for the first time, there is a danger from medium
spurting out. Air bubbles can cause uncontrollable spurting out. Ensure that you stand
back a safe distance.
fig. 3

eco-PEN450
Commissioning and maintenance manual
8 / 15 Version 1.8 Copyright ©
Version B
(e.g. open cartridge, Dispenser fixed in place and supply of the medium under pressure)
• Turn the bleed screw (23) 180 degrees so that the bleed hole is in the position shown
(direction of the drive unit).
• wait until the medium emerges from the bleed hole free from air bubbles.
• Release (remove) the primary pressure on the medium and wait until no more medium
comes out of the bleed opening.
• Remove the ejected medium and close the bleed screw (23) again.
Connect the Dispenser to the power supply and convey the medium until there are no
more air bubbles in it and the outlet nozzle (with the Luer-Lock needle) comes out.
Alternatively, the air can also be extracted from the pump area with the syringe (9)
included in the scope of supply.
Version A
(e.g. supply via a hose, closed cartridge)
• Dispenser bring it into a position in which the end piece
(3) points upwards.
• Dispenser connect the unit to the power supply and
convey the medium until there are no more air bubbles
in it and the outlet nozzle (with the Luer-Lock needle)
comes out.
Tip: Connecting a hose can protect the Dispenser from
being wetted with the medium.
3
fig. 4
23
8
9
fig. 5

eco-PEN450
Commissioning and maintenance manual
Copyright © Version 1.8 9 / 15
5Connection options
Self-levelling liquid, low viscosity medium
Supply of the medium from a cartridge (A) or a
bottle (B).
D Power supply drive unit
Note: Select the size of the hose to be such that
the liquid comes out by gravity and without any
air bubbles at the Dispenser.
Self-suction is possible.
Not a self-levelling liquid, medium/high viscosity medium
1. Cartridge (A) supplied with compressed air (C1), connected to the Dispenser.
2. Medium supply from the pressure tank (C).
fig. 6
C1 Compressed air, 0-6 bar,
regulated
C2 Compressed air, 7 bar, dry,
cleaned, oil-free
D Power supply drive unit fig. 7

eco-PEN450
Commissioning and maintenance manual
10 / 15 Version 1.8 Copyright ©
6Cleaning
Preparation:
Remove the dosing needle, danger of injury!
Disconnect the power supply of the drive unit and uncouple it from the dosing unit (in the
reverse order to that described in 4.1.1, Connecting the dosing unit to the drive unit,
page 7).
1. Undo the screws (18) and pull out the front dispenser housing (19) with the O-ring (22)
from the module (15) while turning it slightly.
2. Undo the union ring (2).
3. Pull off the end piece (3).
4. Pull the stator (7) out of the front dispenser housing (19).
5. Undo the bleed screw (23) and the washer (24).
6. Clean all the parts.
Assembly:
1. Attach the dispenser housing (19) with the O-ring (22) to the module (15) with the
screws (18) (do not exceed a tightening torque of 0.35 Nm).
2. Install the stator in accordance with work steps 4. to 7. described in section 4.1,
Inserting the stator, page 6.
3. Connect the dosing unit to the drive unit (see 4.1.1, page 7).
18
22
15
19
Note: Do not rinse out the rotor
housing (15), because that could
damage the bearings! Clean it
with a cloth and brush. fig. 8
23
24
3
2
19
7
fig. 9