SeQual Eclipse Betriebsanleitung

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
1
Eclipse & Eclipse 2
Eclipse & Eclipse 2
Troubleshooting
Troubleshooting
Alarm Conditions
Alarm Conditions

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
2
Alarm Types
I. Flow Chart (pg 4 & 5)
II. Quick Reference Guide (pg 6 & 7)
1. Low Power Cartridge or warm Power Cartridge (pg 9)
ECLIPSE is: Beeping has a YELLOW light flashing and the unit IS running.
2. Low Oxygen Purity (Between 85% & 70%) (pg 9)
ECLIPSE has: a YELLOW light flashing and the unit IS running. (NO Alarms).
3. Low Oxygen Purity Below 70%) (pg 10)
ECLIPSE is: Beeping with a RED light flashing and the unit IS running.
4. Flow Rate Error / Blocked Flow (pg 11)
ECLIPSE is: BEEPING has a YELLOW BLINKING LED and the unit IS running.
5. No Inspiration Detected (In Pulse Mode Only) (pg 12)
ECLIPSE has: GREEN & YELLOW STEADY LED and the unit IS running.
6. Loss of External Power or Hot Power Cartridge. (pg 13)
Eclipse has: a CONTINUOUS ALARM, a RED FLASHING LIGHT and the
Eclipse is NOT Running.
7. System Malfunction (pg 14)
Eclipse has: a RED LIGHT FLASHING and ALARMING CONTINUOUSLY, and
the unit is NOT Running.
Troubleshooting Guide
Troubleshooting Guide

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
3
Non Alarming Problems
8. Display Panel Malfunction (Buttons).
• The Buttons on Eclipse are not responding to inputs.
• The Eclipse will not Power Down (Turn OFF).
• The Eclipse Powers ON automatically
9. My Power Cartridge will not charge more than 5 bars (Eclipse 1
only).
10. Power Cartridge Re-Calibration.
11. Power Cartridge will not charge (Cascade) when it is in the Eclipse,
and is running off an AC / DC Power Source.
12. Eclipse beeps once every 1-3 min’s (No LED’s) when running off the
12 volt DC System (In the Car, RV, Boat).
Troubleshooting Guide
Troubleshooting Guide

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
4
Alarming
Alarms
Beeping/
Continuous
Beeping
Continuous
LED
Yellow Red
Flashing Steady
Flashing Steady
LED
Yellow Red
Flashing Steady
Flashing Steady
Warm
Battery
Low
Battery
Flow
Block
Loss of
External
Power
Hot
Battery
Syst
Malf
(See
System
Codes)
N/A Low O2
< 70 % N/A
N/A
N/A
NO YES
Go to
Non-
Alarming
Codes
Page
Alarm Code Flow
Chart
Troubleshooting Guide
Troubleshooting Guide

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
5
LED
Yellow Red
Flash Steady
Flash Steady
Not
Applicable Not
Applicable
Can not
Charge
Power
Cartridge
or
No Breath
Detected
(Pulse Mode
Only)
Low O2
<85% -
>70%
No Alarming Codes
Troubleshooting Guide
Troubleshooting Guide

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
6
Quick Reference Alarm Guide
Alarm Priority Green Yellow Red Alarm Code
All OK, warm-up done 13/None ON OFF OFF OFF 00
Warm-up: %O2 < 85 12/Low ON ON OFF OFF 08
Warm-up: %O2 < 70 11/Low ON ON ON OFF 04
Cannot Charge Battery 10/Low N/A ON OFF OFF 01
No breath for 45 seconds 9/Medium ON ON OFF OFF 20
70% < %O2 < 85% 8/Medium OFF (1s/1s) OFF OFF 08
Flow rate error 7/Medium OFF (1s/1s) OFF (0.5s/119.5s) 20
Warm battery 6/Medium ON (1s/1s) OFF (0.5s/119.5s) 02
Low battery 6/Medium ON (1s/1s) OFF 2 Beeps 0.5s/0.5s every 30s 10
DC cable detected 4/Medium N/A (1s/1s) OFF (0.5s/1.5s) N/A
%O2 < 70 3/High OFF OFF (0.25s/0.25s) 3 Beeps (0.5s/0.5 every 120s) 04
Loss of power 2/High OFF OFF (0.25s/0.25s) ON 40
Malfunction 1/High OFF OFF ON 10s then silence N/A
Troubleshooting Guide
Troubleshooting Guide

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
7
Troubleshooting Guide
Troubleshooting Guide
Quick Reference Alarm Guide-System Codes.
Malfunction Code Resolution
Invalid RESET 80
Remove ALL Power Sources
(External Power & Power
Cartridge) for 20 sec's then
restart Eclipse.
If problem persist replace
and/or repair unit.
I/O port failure 81
RAM failure 82
FLASH failure 83
EEPROM failure 84
IPC watchdog timeout 90
Compressor motor stalled 94
Battery communication failure 95
IPC failure A0
Product pressure sensor failure A1
Breath pressure sensor failure A2
Ultrasonic failure A3
Loss of product tank pressure failure A5
Purity calibration data failure A6
Flow calibration data failure A7
Breath sensitivity data failure A8
Hour of operation data failure A9
Invalid assembly option AA
Ambient pressure sensor failure AB

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
8
Malfunction Code Resolution
Compressor motor too hot 91 Remove ALL Power for 20
min's then restart Eclipse.
If problem persist replace
and/or repair unit.
PCB too hot 92
Product temperature sensor failure A4
Troubleshooting Guide
Troubleshooting Guide
Quick Reference Alarm Guide-System Codes

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
9
1. Low Power Cartridge or warm Power Cartridge
Eclipse is: Beeping, has a YELLOW LED flashing and the unit IS running.
A. Is Eclipse plugged into an External Power Source or running off of the Power
Cartridge?
i. If POWER CARTRIDGE, then go to Step B
ii. If External POWER SOURCE (AC or DC), then go to Step C
B. How many bars are on the Power Cartridge Icon?
i. 1 bars or Less
i. The Power Cartridge power is low in the Eclipse. Replace the Power Cartridge
with a fully charged one, or switch to an external power source (Use AC or DC
Adapter).
ii. 2 Bars or More
i. The Power Cartridge is possibly too warm. Replace the Power Cartridge with a
spare one (that has cooled), or switch to an external power source (Use AC or
DC Adapter), and remove the Power Cartridge for approximately 15 to 20
minutes to allow to cool properly, before replacing.
** NOTE: If Eclipse is plugged into an external power source and running from the Power Cartridge
and will not switch over to an external power source (AC or DC) then there may be a system
fault. Exchange out the unit for evaluation.
Troubleshooting Guide
Troubleshooting Guide
Continued on Next Page Æ

11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
10
1. Low Power Cartridge or warm Power Cartridge
Eclipse is: Beeping, has a YELLOW LED flashing and the unit IS running.
ÅContinued from Previous Page
C. Is Eclipse running from the AC or DC Adapter?
i. AC Adapter – There may have been an electrical fluctuation (an electrical dip or
spike), or a power outage at the electrical outlet, Eclipse has switched to the Power
Cartridge as a backup, and now the Power Cartridge is either low or warm (see
Steps B i or B ii)
ii. DC Adapter – The voltage in the vehicle has possibly fallen out of range and
Eclipse has switched to the Power Cartridge as a backup, and now the Power
Cartridge is either low or warm (see Steps B i or B ii)
D. Press the No Smoking Icon once, and observe what number appears after the “AC=” in
the LCD window.
(The Diagnosis Code will disappear after 5 sec’s if no button is pushed).
** NOTE: If Eclipse is plugged into an external power source and running from the Power Cartridge
and will not switch over to an external power source (AC or DC) then there may be a system
fault. Exchange out the unit for evaluation.
Troubleshooting Guide
Troubleshooting Guide
Andere Handbücher für Eclipse
1
Dieses Handbuch passt für folgende Modelle
1
Inhaltsverzeichnis
Beliebte Sicherheitssystem Handbücher anderer Marken

EDM
EDM Solution 6+6 Wireless-AE Bedienungsanleitung

Highway Safety Group
Highway Safety Group EA401 Bedienungsanleitung

Siren
Siren LED GSM Bedienungsanleitung

Detection Systems
Detection Systems 7090i Montageanleitung

Se-Kure Controls
Se-Kure Controls MicroMini SK-4841 Bedienungsanleitung

Siemens
Siemens FDM273 Bedienungsanleitung