4
one flash and program two gives two flashes (see below for program
characteristics).
Very weak or no signal: T-SIGN is dark
Weak signal: Flashes green in a soft manner
Normal signal: Steady green light
Strong signal: Steady red light
Installation (Refer to the enclosed Quick Guide)
1. Preparations before the setup: A functioning hearing loop setup
in accordance with the SIS 60118-4 standard and a device providing
sound source to the hearing loop, e.g. a microphone, TV or a
computer.
2. Choose an appropriate place for T-SIGN. It is not recommended to
mount T-SIGN close to electronic devices due to signal interference.
3. Position T-SIGN at or nearby the chosen mounting place and
connect power to it. Make sure the hearing loop is not in use. Verify
that there is no signal interference that aects T-SIGN when the
hearing loop is not in use. This is shown by that T-SIGN remains
dark when it has been turned on. If T-SIGN is flashing or is indicating
steady green with the hearing loop turned o then consider a
dierent mounting location with less signal interference from the
surroundings. Depending on the magnetic field strength from the
hearing loop, at the chosen mounting location, sensitivity needs to
be adjusted on T-SIGN. A high sensitivity setting in combination with
background signal interference might cause the T-SIGN to show false
indications from the hearing loop.
4. Calibrate T-SIGN in accordance with the instructions in paragraph
”Calibration” below. After successful calibration continue with step 5.
5. Drill two screws onto the wall according to the enclosed drilling
template (page 9) and mount T-SIGN in place.
ENGLISH