Safety Precautions and Warnings - 5
Safety Precautions and Warnings
Warning: Read and fully understand this user manual and all other
safety instructions before operating the PGHH pump. The pressure
inside the pump built-up during use can be extremely high.
Warning: Only personnel with good experience and knowledge of
pressure media, pressure instruments and connections are allowed to
work with the PGHH pump. Incorrect use may result in damage to the
pump, the instrument connected to the pump and/or personal injury.
Warning: Wear protective glasses.
Warning: Use only the Pressure Hose delivered by Beamex, marked
with "Max. 630 bar". Other hoses may not withstand the pressure
generated by the PGHH.
Warning: Never exceed the maximum pressure of the connected
devices. The maximum pressure of Beamex EXT External Pressure
Module is stated on the module's sticker.
Caution: Only use the original Beamex pressure fittings. Impurities
from wrong materials may clog the pump.
Note: The pump is tested using WACKER® AK 10 Silicone Fluid.
There may still be small residues of fluid inside the pump when you
receive it. Before connecting any instruments to the PGHH pump,
flush the pump with the pressure media to be used.
Warning: Always depressurize the system before opening or
connecting any pressure fittings or connectors. Use proper valves for
venting the system. Ensure that all connections are made correctly
and that the hose, the connectors and the pressure fittings are intact.
Caution: If possible, flush the DUT (Device Under Test) before
connecting PGHH to it. Impurities and other process residues
accumulated in the instrument can get into the PGHH and damage
the seals, O-rings and the Reservoir. As soon as first cloudiness,
particles or dirt become visible, change the fluid.