Centron Technologies CM735 Bedienungsanleitung

USER’S MANUAL
BLOOD COLLECTION MIXER
Model CM735
No. CAT.CM73522Ce
Centron Technologies Corporation
319-25 Sadang-4-dong, Dongjak-ku
Seoul, Korea 156-823
Tel. +82-2.522.7807 Fax +82-2.522.7806

1
CM73522UMC.0112
Table of Contents
Page
Important Notes 2
1. Scope 3
1-1. Features 3
1-2. Accessories 5
2. Specifications 6
3. Views and Locations 7
3-1. Top View 7
3-2. Rear View 8
3-3. Control Panel 9
3-4. LCD Display Panel 10
3-5. Tube Clamp Module 11
3-6. Battery Charger and Adaptor 12
4. Functional Descriptions 13
5. How to Use Equipment 16
5-1. Preparation before Use 16
5-2. Operating Procedures 17
5-3. Audible Alarm Signals 19
5-4. Recharging the Battery 21
5-5. Moving the Clamp Module 23
6. Maintenance 24
6-1. Troubleshooting 24
6-2. Calibration 25
6-3. Leveling the Tray 26
6-4. Service Call 27

2
CM73522UMC.0112
Important Notes
Power Source
The Blood Collection Mixer model CM735 is operated by either a battery pack or an
AC power adaptor.
Caution
Do not put excessive weight on or push the tray downward in a forceful manner.
(This avoids damage to the weight-sensitive device under the tray.)
Do not place weights over 2 Kg on the tray.
Battery Capacity
Turn power switch off to save the battery when the machine is not to be used for
hours. If a machine, with a fully charged battery pack is left in but unused, the
battery will be exhausted in about 15-20 hours. A machine with a fully charged
battery can operate continuously for approximately 8 hours.
Battery Recharging
A low-voltage symbol flashes on the display panel and an alarm will sound when the
battery voltage gets low. To recharge the battery, just plug the AC power adaptor in
the rear panel jack, or take out battery and put it on the battery charger unit.
Cleaning Instructions
Use a soft cloth and mild detergent to clean the equipment. DO NOT use paint
thinner, benzene, solvent or strong detergent.
Keeping
Avoid using or leaving the machine in a location where the temperature is higher
than 122 °F (50 °C). Keep it away from any heating source or direct sunlight.

3
CM73522UMC.0112
1. Scope
1-1. Features
•The Blood Collection Mixer CM735 provides efficient mixing of the blood with the
anticoagulant in the blood bags.
•It is a lightweight mobile unit with maximum portability.
•The panel display indicates the net volume of blood collected in the blood bag. It
automatically closes the tube clamp when the collected volume reaches the preset
selected figure. The display shows the total weight on the tray at the end of the
donation and elapsed collection time.
•Various audible and visual alarms are provided to ensure accurate and safe collection
and mixing.
The mixing motion is controlled by a microcomputer. When the blood collection is
started, the tray rocks for 3 cycles and pauses for 2 seconds, then resumes rocking. As
the collection approaches the preset donation amount, the tray stops rocking to ensure an
accurate donation. When the preset amount is reached, the clamp closes and tray resumes
rocking.
The Blood Collection Mixer monitors blood flow rate during the collection and donation
draw time. When the draw is less than 30 ml/minutes, the machine will alarm. The alarm
is both audible and visual. The alarm repeats every 60 seconds if blood flow remains low.
The LCD panel displays pertinent information.
The following information can be viewed on the LCD:
- total weight [gram] of blood and blood bags
- collected volume [ml] of blood
- target volume to collect [ml]
- elapsed time [min : sec]
- low battery voltage condition
- unit of displayed figure ([ml] or [gram])
- alarming condition
The unit can be powered by either a battery pack or an AC power adaptor.
A Nickel-Cadmium battery pack of 12 Volts/2AH capacity is used. It is replaceable and
rechargeable. An audible alarm sounds when battery voltage drops to its lower limit.
The battery can be recharged while in the mixer or with a separate battery charger unit.
The battery can be recharged while the mixer is in use if the AC power adaptor is
attached at rear panel jack of main unit. The battery can also be recharged by removing it
from the mixer and connecting it to a separate battery charger unit.

4
CM73522UMC.0112
(A large battery charger, model BC162, which charges 6 battery packs same time, is
available for purchase from your distributor. With this charger, one pack among 6 is
recharged fast in 3 hours, and other 5 packs in 10 hours.)
It takes about 10 hours to recharge a pack from exhausted level to fully charged level.
A fully charged battery will operate the mixer continuously for approximately 8 hours.
The blood collection mixer unit can also be powered directly by using the AC power
adaptor. When both a battery pack and a power adaptor are connected at the same time,
the power adaptor supplies operating power and the battery is under recharging.
The tube clamp module can be positioned on either the left side or the right side of the
tray. It can easily be remounted on either side of the tray by the user.
The equipment is stored in a soft carrying bag, and weighs less than 4 kilograms.
When the power switch is turned on, the machine requires 30 seconds to stabilize
weight-sensing circuits. During this time the display shows a 30-second count down.
The Blood Collection Mixer is a unit designed with solid-state electronics and advanced
technology.

5
CM73522UMC.0112
1-2. Accessories included with each unit
The Blood Collection Mixer CM735 includes the items shown below.
(Quantity - 1 each)
CollectionMixerBase Collection/WeighingTray
BatteryPack BatteryChargerUnit ACPowerAdaptor
And,
Carrying/Transportation Case
User's Manual

6
CM73522UMC.0112
2. Specifications
a. Collection volume display ; 0 ~ 999 ml, with 1 ml increments
b. Total weight display ; 0 ~ 999 grams, with 1 gram increments
c. Elapsed time display ; 0 ~ 19 min 59 sec
d. Target volume set
(1) Range ; 0 ~ 995 ml, with 5 ml steps
(2) Set method ; up/down and preset buttons
(3) Preset values
(*1) ; 350, 400, 450, 500 ml
(4) Default value
(*1) ; 450 ml
e. Alarms
(*2)
(1) Low battery voltage ; audible, visual alarms at low battery voltage
(2) Low flow rate ; audible, visual alarms at 30 ml/min or lower
(3) Tray removal ; audible, visual alarms at tray removal
(4) Long collection time ; beeps at 10 and 20 minutes after start
f. Tray moving
(1) Speed ; 60 RPM speed nominal, with +/- 12.5 degrees
(2) Motion mode ; 3 cycles moving - 2 seconds pause
g. Accuracy
(1) Weight calibration tolerance ; +/- 1 gram
(2) Decimal truncation error ; 1 gram/ml/sec maximum
(3) Clamp closing time delay error ; 1 gram maximum
(4) Collection volume accuracy ; - 0 ml to + 3 ml
(5) Total weight accuracy ; - 0 gram to + 3 grams
(6) Time display accuracy ; 1 second maximum
h. Controls ; POWER,START,END,TIME,CLAMP,PRESET,UP,DOWN
i. Indication panel ; custom-designed composite LCD display
j. Operating power source ; 12 volts DC battery pack or AC power adaptor
k. Battery pack ; Nickel-Cadmium rechargeable, 12 Volts DC/2AH
l. Power adaptor ; 220-240 VAC (*3), 50-60 Hz In, 18 VDC/1 Ampere Out
m. Battery charging ; by built-in charging circuit, or by separate charger unit
n. Power consumption ; 2 watts maximum
o. Dimensions ;
Equipment 11.5 L x8.0 W x5.5 H inches
29.2 L x20.3 W x14 H cm
p. Weight ;
Equipment with battery 3.2 Kg (7.0 lbs)
q. Temperature characteristics ;
Operating 0 ~ 40
oC (32 ~ 104
oF)
Storage - 20 ~ 70
oC (- 4 ~ 158 oF)
Note : (*1) These figures may vary upon user’s request.
(*2) Certain version doesn’t have 10 & 20 minutes and/or tray removal alarms.
(*3) A power adaptor of 110-120 VAC input is available as necessary.

7
CM73522UMC.0112
3. Views and Locations
3-1. Top View
(1) Control Panel
(2) LCD Display Panel
(3) Power Switch
(4) Tube Clamp Module (now mounted at left side)
(5) Tray
(6) Handle
(7) Weight Calibration Hole
(8) Tray Leveling Screw
(9) Clamp Module Connection Hole
(10) LED lamp

8
CM73522UMC.0112
3-2. Rear View
(1) Battery Pack
(2) Battery Compartment Door
(3) Power Adaptor Inlet
(4) Tray
(5) Tray Mount
(6) Handle
(7) Tube Clamp Module

9
CM73522UMC.0112
3-3. Control Panel
(1) [START] Switch
(2) [END] Switch
(3) [TIME] Switch
(4) [CLAMP] Switch
(5) LCD Display Panel
(6) [UP] Switch
(7) [DOWN] Switch
(8) [PRESET] Switch