2.3 Exhaust temperature transmitter set-up
Configuration changes made using the PDT should be performed with the pump not running. It is recommended
that the CXS dry pump is then power-cycled following configuration changes.
The Pump Display Terminal (PDT) must be used to configure the exhaust temperature transmitter. The PDT must be
connected and used in accordance with the instructions in the CXS instruction manual M528-00-880.
Figure 6 - Pump Display Terminal (PDT)
The PDT should be plugged in and the pump switched on. When the PDT is powered up, select and set the transmitter
accessory options as follows:
1. Press 'setup' - scroll using arrow keys to 'Service'.
2. Scroll to 'fit accessory'.
3. If loom is connected to ' Bulkhead Connector 8' then scroll to transmitter (PT100 1) and select 'fitted' using arrow
keys and press 'enter'.
4. If loom is connected to 'Bulkhead Connector 2' then scroll to transmitter PT100 2 and select 'fitted' using arrow
keys and press 'enter'.
5. To exit the menu press the 'cancel' key three times where the indicator lights will revert to the initial menu.
The PDT is a non-Ex device, suitable for use in a hazardous area. It must be connected using a
suitable cable according to EN60079-14 and routed into the safe area.
The test/diagnostics port on the Ex d control box door is intended for connection to a PDT for
commissioning, service and diagnostic purposes, only when:
- the CXS dry pump is installed in a safe area, or if safety within the hazardous area can be
guaranteed, for example by the use of a suitable gas detector to ensure the absence of any
explosive atmosphere.
After the PDT socket has been used, the explosion-proof stopping-plug must be refitted. Check the
condition of the threads and O-ring, tighten hand-tight then a further half turn with a wrench.
Incorrect use of the PDT socket and PDT will invalidate the explosion-proof rating of the CXS dry
pump and may result in an explosion in the hazardous area.