Model 61059
SELECTIS ALTERNATING PRESSURE PUMP AND MATTRESS
USER MANUAL

Important Safeguards
1

Pump
The funcons of the Serenty Elite pump are described below. Please refer to the
below figures.
Power Switch(1)
The switch is at the right side of the pump.
Turn ON/OFF the power and press Power buton on the panel,the pump will
start/stop operaton.
Press Cycle Time(2)
The cycle me is selectable from the panel.
The cycle me value can be either 10, 15, 20, or 25 minutes.
Press Pressure Range(3)
20mmHg~55mmHg is offered from the panel, providing op ns of pressure
value by easy press.
Alarm Mute (4)
The audible/visible alarm turns on aer 1 minute when low pressure,power
failure or alternate failure.
To mute the audible alarm, press the Alarm Mute buton. The visible alarm
indicator will flash l problem solved.
Re-press the Alarm Mute buton to reac ate the alarm.
Alternate mode(5)
Press to set the air matress in alternate mode.
Statc mode(6)
Press to set the air matress in statc mode.
Max Firm(7)
Press to set the air matress in quick inflaaon mode, which facilitates nursing
and caring.
Seat Inflate (8)
Press to set the air matress in seat inflate+ alternate or seat inflate+statc
mode, pressure value will increase by 5mmHg, which facilitates sea g.
Under max firm mode, because the pressure is already set at the largest
value, no more 5mmHg will be increased.
Lock buton (9)
The panel will automacaly lock in 5 minutes without opera n.
Press the lock buton and hold for 3 seconds,to lock or unlock the panel.
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
5 6 7 8
9
Matress
Elite
Intended Use
The Selecs Serenity Elite pump and maress are intended to reduce the incidence of
pressure ulcers while opmizing paent comfort. It also provides the following:
Individual home care seng and long-term care
Pain management as prescribed by a physician
Product Funcons

4 5
Place pump on horizontal surface or hang it by means of the hook.
Open maess on the bed. Secure it with end flaps or straps.
Connect air feed tubes to air outlet on the maess and pump, ensuring that the
tubes are not“kinked”or twisted. Allow space for bed rails to drop freely.
Turn on the power. Aer a beep sound, the pump will begin the operaton.
The pump will enter statc mode initally when the power is turned on. The cycle
me is set at the 1st LED (10 min.), and the pressure level is set at the 3rd LED
(30±2mmHg).First it will inflate the maess to 10-15±
2 mmHg and then enter
alternate mode.
During normal opera n, the pump will monitor pressure changes. If the pressure
is below a user-defined pressure level, the pump will automacaly start to inflate
the maess.The pump will stop when the user-defined pressure level is reached.
The first infllaon of new maess: the alarm func n (visible and audible alarm)
will actvate if matress is not inflated completely within 40 minutes.
During pump is running, low pressure alarm will actvate within 1 minute aer air
tubes are disconnected or air cells are broken.
If there should be power failure,for example caused by power cord unplugged or
power offwhen pump is working,the audible alarm will be actvated and power
failure indicator will flash.When power is back,pump will automatcally start
working.
If the matress stops alternatng or alternate abnormally,the audible alarm will be
actvated and alternate failure indicator will flash.
For nursing and caring convenience,press the max firm buon to automacaly
inflate the maesses to the maximum level (55mmHg) for about 20 minutes. The
pressure will return to alternate mode aer 20 minutes.
Installaon & Operang Instrucons Environment Requirements
·Operaton Temperature: 50℉~95℉(10℃~35℃)
·Storage Temperature: 5℉~122℉(-15℃~50℃)
·Operaton Humidity: 20%~80% non-condensing
·Storage Humidity: 10%~90% non-condensing
Cleaning Guidelines
·Can be cleaned by disinfectant soluon
·Do not immerse the pump in water
·Do not heat or steam in autoclave
·The matress and cover are aerated unl they are fully dry
Maintenance
General
Check the power cord and plug to see if there are abrasions or excessive wear.
Check the maress cover for signs of wear or damage. Ensure the maress cover and tubes
are connected together correctly.
Plug in the pump unit and check the airflow from the hose connecon port. The airflow should
alternate between ports every half-cycle me.
Check the air hoses to see if there are any kinks or breaks. For replacement, please contact
your local agent or dealer.
Make sure the maress tube is well connected.
Check the pump unit and make sure both power indicators are off when the switch is turned
off.
Low pressure
Examine if there is any air leakage between the pump and the maress connecons or from the air
maress tubes:
Check connectors between the air maress and pump. If there is any disconnecon, please
reconnect it.
Check the CPR Valves. Ensure their outlets are sealed.
Check the air-connecng tubes. Ensure each single cell is not broken.
Set the pressure at Max firm.Keep the tubes fully inflated and inspect for air leakage.
Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage occurs,
please contact your local agent or dealer.
·
·
·
·
·
·
·
·
·
·
·

Matress
Item: Seren Elite- 61059
Size:80"(L) x 36" (W) x8"(H)
Top cover: Quilted bi-elastc cover with
zipper
Air cells:Nylon PVC
(8 cell on cell constructon with 5 upper
layer alternatng and 3 lower layer statc)
Base: Nylon /PVC 840D
20 air cells
CPR valve for emergency procedures
Transport functon
Quick connectors
Pillow on
Pump
Item: Serenity Elite
Power supply: 120V 60Hz 1A
Air output: 8 litermin.
Pressure range: 20 mmHg 55 mmHg
Cycle 10/15/20/25 min.
Max firm mode
Alarm:Low pressure/Power failure/
Alternate failure
Alternate/Statc /Seat inflate mode
Size:31cm(L) x 12.5cm(W) x 21cm(H)
Weight: 2.5 kg
Technical Specific
6 7
Tel:718-942-3372
Email: info@emeraldsupply.net
www.emeraldsupply.net
""
"
-
/
/,
Inhaltsverzeichnis
Beliebte Wasserpumpe Handbücher anderer Marken

Sykes AmeriPumps
Sykes AmeriPumps GP100M Fehlerbehebungshandbuch

DUROMAX
DUROMAX XP WX Series Bedienungsanleitung

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 Bedienungsanleitung

Franklin Electric
Franklin Electric IPS Bedienungsanleitung

Xylem
Xylem e-1532 Series Bedienungsanleitung

Milton Roy
Milton Roy PRIMEROYAL Bedienungsanleitung