GE GE006 Bedienungsanleitung


Preface
32
Preface
Thank you for selecting the GE006 Blood Glucose Monitoring System. This manual provides all the
information you need to operate this product for accurate test results. Please read this entire manual
before you start any testing.
For people living with diabetes, it is important to regularly monitor blood glucose levels to effectively
reduce complications from the disease. The easy-to-use GE006 Blood Glucose Monitoring System
provides accurate, reliable test results to help you better manage your diabetes.
The GE006 Blood Glucose Monitoring System is designed for in vitro diagnostic use only (for self-testing
by a single user outside the body). Testing requires only a small amount of fresh capillary whole blood
from either the fingertip, palm or forearm.
The GE006 Blood Glucose Monitoring System is manufactured and supported by Bionime Corporation. If
you have any questions or concerns, please contact your local Bionime Customer Service or call us at
886 0800 241 688 GMT+08:00, 08:30~17:30.
The GE006 Blood Glucose Monitoring System is a personal blood glucose monitoring system to be used
for self-testing only. It is not recommended for multiple users.
A healthcare professional should be contacted when Customer Service is not available.

Caution
54
Precaution
- Before using the GE006 Blood Glucose Monitoring System to test your blood glucose, please read all of
the instructions and conduct all of the tests including the quality control test (Refer to page 37).
- Please perform the quality control test regularly to make sure the test results are accurate.
- The GE006 Blood Glucose Meter can only be used with GE006 Blood Glucose Test Strips. Other brands’
test strips should not be used under any circumstances. The use of other brands’ strips may give
inaccurate results.
- The GE006 Blood Glucose Monitoring System is intended for in vitro diagnostic use only. The GE006
Blood Glucose test results using fresh capillary whole blood samples from the fingertip, palm and
forearm are calibrated to be the equivalent to that of plasma testing.
- The GE006 Blood Glucose Monitoring System is intended for self-testing. It should not be used to
screen for or diagnose diabetes mellitus.
- If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature,
please wait 30 minutes before measurement.
- Follow all environmental protection regulations when disposing of batteries.
- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the GE006 Blood Glucose Monitoring System is 0.75µL:
( )
Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading. An Er4
reading will be displayed if the sample size is too small. In this case, repeat the test with a new test strip.
Sample Size Example 0.75µL 1.0 1.5µL 2.0µLµL 3.0µL
Important Safety Notes:
- The GE006 Blood Glucose Meter and Lancing Device are for single patient use. Do not
share them with anyone including other family members.
- All parts of the kit are considered biohazardous and can potentially transmit infectious
diseases, even after following the cleaning and disinfecting procedures.
- Users should wash their hands thoroughly with soap and water before and after handling
the meter, lancing device, or test strips.
Please refer to the
section "Cleaning and Disinfecting Procedures" on page 47.

Table of Contents
76
Table of Contents
About the Blood Glucose Monitoring System
The Blood Glucose Monitoring System ........ ................................... ................
The GE006 Blood Glucose Meter ................................................. ..................................
The GE006 Blood Glucose Test Strip ......................................................... .................
Before Testing
Meter Activation and Battery Change ............................................................................... .......
Setting Up Your Meter - Setting the Date, Time and Volume ............................................... ................
Turning On/Off the Meter ........................................................................................... ....................
Turning On/Off the Bluetooth .............................................................................. .......................
Handling the GE006 Blood Glucose Test Strip ................................................ ..................................
Testing Procedure
Getting Ready for Testing ......................................................................................... .....................
Performing a Blood Glucose Test .............................................. .............................................
Setting Markers .......
Alternative Site Testing ......................................................... ..........................................................
View Window Appearance .................................................... ......................................................
Understanding Test Results and Messages ......................................................... ....................................................
GE006
GE006 ................................................. .........
.............................................................
...............................................................
................................................
.........................
................................................
................................................
...................................
................................................
...................................................
...........................................................................................................................................................................
................................................
.................................................
.............
Quality Control
About Quality Control Testing ......................................................................................................................................................
Performing a Quality Control Test .............................................................................................................................................
Understanding Control Test Results ..........................................................................................................................................
.....................................................
.....................................................
.....................................................
......................................................
.....................................................
..................................................
....................................................................................................................................................................
.........................................................................................
NCC 警語 ...............................................................................................................................................................................................
..................................................
..................................................
.................................................
.................................................
.................................................
.................................................
.................................................
..................................................
Test Memory
Recalling Test Results ......................................................... .........................................................
Recalling Average Test Results ....................................................... ........................................
Caring for Your Meter .............................................................................. ....................................
Cleaning and Disinfecting Procedures .............................................. .................................
Error Messages and Troubleshooting ................................................. ................................
Limitations ................................................................................................. ..........................................
Real Time Synchronize
Specifications ................................................................................................ ....................................
Warranty ......................................................................................................... .....................................
Customer Service .................................................................................................................. ............
Expected Blood Glucose Values Without Diabetes ........................................... ................
Component Manufacturer Information ........................................................... .......................
Log Book .......................................................................................................................... .....................
Warranty Card ...................................................................................................................... .............
Emergency Card ..................................................................................................... ..........................
FEDERAL COMMUNICATIONS COMMISION (FCC) STATEMENT
40
42
43
45
46
47
51
55
56
58
60
61
63
64
65
66
67
82
85
8
10
14
16
18
21
22
23
25
26
32
34
35
31
37

9
10 *
53 *2
6 *
1
8
4 *
7 *
8 *
9 *
The GE006 Blood Glucose Monitoring System
Contents
Your GE006 Blood Glucose Monitoring System consists of several items. Please identify each item, learn
its name and how it is used.
Below are the items included in your GE006 Blood Glucose Monitoring System:
1. Getting Started Guide
2. User's Manual (includes Log Book, Warranty Card, Emergency Card )
3. GE006 Blood Glucose Test Strip Package Insert *
4. Rightest® Control Solution GC550 Package Insert *
5. GE006 Blood Glucose Meter ( with 2 AAA batteries installed )
6. GE006 Blood Glucose Test strips ( 10/25 pcs )*
7. *
8. Lancing Device *
9. Clear Cap *
10. Disposable Sterile Lancets ( 10 pcs ) *
11. Carrying Case ( not shown )*
12. Instruction for the lancing device ( not shown )*
(* Different packages have different bundled items. Some of packages might not include * items.)
*
*
Rightest® Control Solution GC550
Rightest® GD500
The GE006 Blood Glucose Monitoring System
If you want to purchase optional items, please contact the Customer Support Center at
866-0800-241-688.

1110
The Blood Glucose MeterGE006 The Blood Glucose MeterGE006
Left Button ( )
Main Button ( )
Display Window
Right Button ( )
Test Strip Port
Meter Battery
Product name
Battery Cover
Meter Serial Number

1312
Indicates a test result stored in memory
Indicates a test result not included into the Average
function of this meter
Indicates the average result
Indicates a control solution test result
Unit of test result
Test result
Warns when the battery is low or must be replaced
Indicates when to apply the blood sample
Indicates the time in 12H format
The GE006 Blood Glucose Meter The GE006 Blood Glucose Meter
Appears when an error occurs
Appears after you insert test strip into meter
Current date under time mode or testing date under
memory mode
Warns when the operational temperature limit is
exceeded during testing
Displays current time under time mode or testing time
under memory mode
Manufacturing use only
Indicates a meal marker with test result
Indicates that the Bluetooth is turned on
Indicates that the Bluetooth is turned off
Indicates the time period of test results

1514
The GE006 Blood Glucose Test Strip The GE006 Blood Glucose Test Strip
The GE006 Blood Glucose Meter can only be used with GE006 Blood Glucose Test Strips and Rightest™
Control Solution GC550. The use of other test strips or control solutions can lead to incorrect results.
Sample Port
Apply blood/control solution here.
Noble Metal Electrodes
Electrochemical sensor
View WIndow
PRECAUTION
- Close the GE006 Blood Glucose Test Strip vial immediately after removing a test strip.
- Do not reuse GE006 Blood Glucose Test Strips.
- Do not use expired GE006 Blood Glucose Test Strips.
- When you open a new vial of GE006 Blood Glucose Test Strips, record the date on the vial.
Discard the vial of test strips after 3 months from opening.
- Store the GE006 Blood Glucose Test Strip between 4-30°C (or 39-86°F) and in a location (<
90% relative humidity). Do not expose to direct sunlight or heat.
- If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial change in
temperature, please wait 45 minutes before measurement.
- For detailed information, please refer to the GE006 Blood Glucose Test Strip Package
Insert.
Electrode Contacts
Hold here to insert
test strip into
meter.

17
16
Meter Activation and Battery Change
Your Blood Glucose Meter comes with two AAA, batteries installed. Two new batteries will
provide power to perform approximately 800 tests under normal use with one data transfer. Press the
main button or insert a strip to activate your meter.
GE006
4. The GE006 Blood Glucose Meter
5. Press any button to exit the self-test and enter Setting Mode.
will enter Self-Testing Mode automatically when replacing the battery
(all symbols will appear on the screen).
6. You must set the time and date when replacing the battery. See Chapter “Setting Up Your Meter -
Setting the Date, Time, and Volume" on page 18. Test results are still stored in the memory.
1. Turn the
over. Press and slide the
battery cover to open.
GE006 Blood Glucose
Meter
2. Install the batteries. Be sure to
put the batteries in the correct
direction.
3. Slide the battery cover
back until it snaps into
place.
Meter Activation and Battery Change
CAUTION
- Please follow the local regulation to properly discard a used battery.
- Risk of explosion if battery is replaced by an incorrect type.
You can enter the Setting Mode two ways.
1. Replace the Battery
2. With Battery Inserted
After removing the battery, press the main button several times until there is no signal on screen,
then follow the battery installation steps to replace battery. The GE006 Blood Glucose Meter will
perform a self-test. Press the main button to exit the self-test and enter the Setting Mode.
Press the main button to turn on the GE006 Blood Glucose Meter. Hold down the main button for 7
seconds. During this time the screen will go blank until you hear a beep. After the beep, the meter will
turn on into the Setting Mode. The display screen will show setting data.
NOTE
- Use the Left and Right buttons to select your setting data and press the main button to
confirm each selection. After confirming all of the settings, you will return to the Time Mode.

1918
1)
2)
3)
4a) 4b)
5) 6)
1. Year setting
, press the Left or Right button until you see the
current year. Press the Main button to confirm.
2. Month setting
With the month blinking, press the Left or Right button until you see the
current month. Press the Main button to confirm.
With the year format blinking
Once the year is confirmed, the
Month Setting will appear.
Once the month is
confirmed, the Day Setting will appear.
Once the day is confirmed, the Time
Setting will appear.
3. Day setting
With the day blinking, press the Left or Right button until you see the current
day. Press the Main button to confirm.
Setting Up Your Meter - Setting the Date, Time and Volume Setting Up Your Meter - Setting the Date, Time and Volume
4. Time format 12/24H selection
With the time format blinking, press the Left or Right
button to switch between 12H and 24H. Press the Main
button to confirm.
5. Hour setting
With the hour blinking, press the Left or Right button until
you see the current hour. Press the Main button to
confirm.
6. Minute setting
With the minutes blinking, press the Left or Right button
until you see the current minute. Press the Main button to
confirm.
Once confirmed, the Hour Setting will
appear.
Once the hour is confirmed, the Minute Setting
will appear.
Once the minute is confirmed, the Volume
Setting will appear.
Inhaltsverzeichnis
Andere GE Blutzuckermessgerät Handbücher
Beliebte Blutzuckermessgerät Handbücher anderer Marken

ReliOn
ReliOn Prime Bedienerhandbuch

AgaMatrix
AgaMatrix Presto Bedienungsanleitung

Polymer Technology Systems
Polymer Technology Systems A1C Now Self Check Bedienungsanleitung

Medisana
Medisana GlucoDock Bedienungsanleitung

Arkray
Arkray ASSURE PLATINUM Bedienungsanleitung

Dexcom
Dexcom G4 Professional Bedienungsanleitung

LabStyle
LabStyle Dario Bedienungsanleitung

ACONLABS
ACONLABS ON CALL EXPRESS Bedienungsanleitung

Higi Technology
Higi Technology iGlucose Bedienungsanleitung

Bayer HealthCare
Bayer HealthCare CONTOUR NEXT Bedienungsanleitung

Accu-Chek
Accu-Chek ADVANTAGE Bedienungsanleitung

Sanofi-aventis
Sanofi-aventis BG STAR Technische Daten