hadewe Azuras CHROME Bedienungsanleitung

art. 0982
FB 04-303 Rev. 1
Date: 30.01.2019
Manual
Azuras CHROME

2
Directory
Foreword 3
Safety Instructions 4
Cleaning & Disinfection 10
Discover Device 11
Before Use 12
Area of Application 12
Operation 12
Usage 14
Operating Steps 14
Work Positioning 15
Infilling the Spray Speed 17
Spray 17
Insert Bur 17
Adjusting the Speed 17
Memory Keys 18
Foot Switch 18
Select Menu 19
Help Menu 19
Settings Menu 20
Confirm Washer Change 20
Turn Light ON/OFF 20
Pictures & Videos 21
Messages 24
Maintenance & Care 25
Washer 25
Cleaning the Chuck 25
Water Filter 26
Cleaning Function 26
Maintenance Overview 27
Upload Firmware 27
Handling Instructions 28
Troubleshooting 29
Technical Details 30

3
Foreword
The Azuras CHROME foot care unit was produced according to strict quality criteria
and uses the latest technology in its design. The images in this manual may vary
slightly from the product you have bought as we are constantly developing our
products.
This user manual includes a detailed description and explanation of how to use the
Azuras CHROME unit. If you have any questions or suggestions, you can contact us by
telephone or email.
This user manual belongs with the Azuras CHROME unit. Keep it somewhere safe. If
you give this product to someone else, please include this document as it includes
important information on how to correctly use the product.
Short description of functions
The Azuras CHROME is a device for the treatment of callus and nails. The handpiece
is compatible with rotating tools such as diamond polishers, steel cutters and ceramic
cutters. These can be worked with up to a speed of 40,000 rpm thus providing an
efficient way of removing tissue. The direction of rotation can be changed by pressing a
button. The integrated spray function emerges reliably bundles emerging dusts. The
rotation speed can be adjusted using the keyboard. The amount of spray liquid is
adjusted via knob. All other functions can be accessed with the touch display. The
device display will notify you when the washer needs to be changed and when the
device is due for inspection. Functions and maintenance steps will be described with
pictures and videos shown in the display. You can further use an USB stick featuring
your own pictures. For example, you may want to show your customers the separate
successful steps from treatment to treatment.

4
Safety Instructions
The installation of the working area has to
correspond to the relevant regulations. Set up
the unit in such a way, that the air slits are not
blocked and the extracted air can escape well.
This device shall only be used by persons who
are familiar with feature and function of the unit.
This appliance can be used by children aged
from 8 years and above and persons with
reduced physical, sensory or mental capabilities
or lack of experience and knowledge if they
have been given supervision or instruction
concerning use of the appliance in a safe way
and understand the hazards involved. Children
should be supervised not to play with the
appliance. Cleaning and user maintenance shall
not be made by children without supervision.
To avoid the risk of electrocution only connect
the device to the main power supply. If you pull
the power plug, it will stop the connection to the
main power supply. Set up the unit so that the
power plug is always accessible. The power
switch should also always be easily accessible.
Only switch on the unit if the handpiece rests
secured in the handpiece holder, or if you hold it
safely in your hand.

5
Safety Instructions
Prevent any liquids from entering into the control
unit, since this could damage the electronic
elements. Clean the outside of the box only with
a damped cloth. Make sure, that you have
unplugged the unit before.
Only when pulling the mains cable, you can
ensure that the device will be completely
disconnected from mains. Check the mains
cable regularly for damages and in case of
such, have it changed or repaired by an
authorized service center.
Prevent your hair from wrapping itself around
moving parts. If applicable, wear a hair net.
When working with materials, which might
create dust or moisture, use a drilling unit with
suction or spray technique. Wear a nose and
mouth protection. Always follow the safety
instructions, provided by the manufacturer of the
materials being formed and used.
For safety reasons, you must always wear
disposable gloves and protective glasses when
operating this unit. In addition, a face mask
conforming to EN 14683 type IIR or EN 149 FFP
2 must also be worn.

6
Safety Instructions
Ensure that the patient’s foot and the handpiece
are held firmly in place during the operation. To
prevent injury, work with extreme caution on the
foot of the patient and use only light pressure.
If water leaks from the unit, then a disturbance
is present. The unit has to be separated from
the socket and cannot be used any longer.
Only use distilled water plus a sterilizing
additive, e.g. hadewe AquaHy (art. no. 8555). If
you use other brands, pay attention to their
manuals if you can use them for this purpose.
Never use tap water or alcoholic fluid, or other
burnable liquids. In case, you are working with
an alcoholic fluid, or other burnable liquids,
there is always the danger of fire or explosion
hazards. Never refill the tank directly over the
unit to avoid in case you accidently spill water,
this will flow into the unit.
Change the spray liquid every day. Always add
disinfection additive to the distilled water.
Rinse the tank daily and fill it with fresh water
every day.

7
Safety Instructions
Empty the unit if it is not in use for a longer
period of time, e.g. before vacations. Clean out
the tank. Set the spray function on maximum.
Now start the spray function, and the water will
spray for approx. 3 min. The fluid will be blown
out of the complete spray system and
handpiece.
To ensure that the bur/tool sits firmly in the
chuck, do not work with tools which
•have an oily shaft.
•are worn.
•are bent.
Before using the equipment, assure that
•the tool can be pushed far enough into the
chuck (test this with a dipstick and clean if
necessary),
•the inserted tool sits properly in the chuck
and will not cant/jam while clamping it,
•the tool is approved for using it at the max
speed (see manufacturer’s instructions), or
•only use the speed required by the
instrument,
•the tool is absolutely dry.
For safety reasons, do not change the tool while
the motor is running.

8
Safety Instructions
Switch off the unit immediately if
•the bur comes off during the treatment itself,
•the motor blocks during the treatment.
The light on the handpiece is for lighting the
area being treated. Do not look directly into the
light.
Always follow the operating instructions. Do not
carry out any repairs, modifications or
maintenance work yourself. This is only to be
accomplished by an authorized specialist. The
mains cable may also only be changed by the
manufacturer, or an authorized specialist.
Maintenance records may be supplied to repair
workshops and trained technicians upon
request. Never remove the housing, especially
not while the equipment is plugged into the
socket, since some components are under
electric voltage. Any unauthorized or improper
handling (e.g. trying to repair the unit) of the
equipment will immediately nullify the
guarantee/warranty and any other liabilities of
the manufacturer.
Using the memory buttons: Only use burs/tools
which are permitted for speed level saved under
(15) to (17)!

9
Safety Instructions
When the device is brought out of standby
mode, it will run with the rotation speed used
last. Take care to not start the mill at an overly
high rotation speed. Select the required rotation
speed before changing the milling tool.
In case the touch display shows a damage or
does not work properly the device may not be
used any longer. Please send it to a qualified
service workshop for repair.

10
Cleaning & Disinfection
The device is designed to be compatible with rotating
instruments which do not penetrate the skin or mucosal or
inadvertently may still cause injuries. For this application
please adhere to the cleaning and disinfection instructions
provided below. After every treatment please carry out the
following steps:
•Remove the rotating tool (e.g. bur, capper etc.) first. Then
handle the tool according to the instructions of its
manufacturer.
•Remove tissue remains by wiping off the unit with a mildly
moistened paper cloth. Afterwards clean all surfaces with
disinfection cloth (e.g. Mikrozid AF Tücher of Schülke and
Mayr). There is no limit to the number of times this device can
be cleaned and disinfected during its recommended lifespan.
Inhaltsverzeichnis
Andere hadewe Pediküre Spa Handbücher

hadewe
hadewe Auroria Bedienungsanleitung

hadewe
hadewe Argos 2 Bedienungsanleitung

hadewe
hadewe Xantos-BT Bedienungsanleitung

hadewe
hadewe SB2 Series Bedienungsanleitung

hadewe
hadewe Azuras PLUS Bedienungsanleitung

hadewe
hadewe HELIUS-40 Bedienungsanleitung

hadewe
hadewe Helius 22 Bedienungsanleitung

hadewe
hadewe Veloria MD Bedienungsanleitung

hadewe
hadewe Veloria Chrome Bedienungsanleitung

hadewe
hadewe Medeas Bedienungsanleitung