• accuracy and confidence of connections
• condition of cable insulation
• use only mains socket with fully efficient grounding pin
• to attach load (ultrasonic transducer or model resistance), use only
cables with appropriate insulation class
If the device is in operation:
• do not touch /do not short load (ultrasound transducer) connectors
• do not leave working generator without supervision
• signal output must be disabled before changing the load (ultrasound
transducer); it must be done in all operating modes (spectrum analyzer,
power analyzer, continuous work); switch button F1 in STOP position (the
color will change to red) and check if LED indicator (“Output”) on front panel is
deactivated
• do not change the fuse
• do not disconnect mains cable
• do not disconnect RS232 cable
Before replacing the fuse the generator must be turned off – exit the program
(F10), turn off the PC, turn off the device and disconnect mains cable.
Replace the fuse only with identical one (see the description below the fuse
socket) (glass body x20mm 6.3A 2 0V (quick-acting)).
WARNING!
FAILURE TO FOLLOW THESE RULES MAY CAUSE SERIOUS INJURY OR
DEATH.
The user is responsible for any damages or injuries caused by the
improper use of the device.
3.3 Installation and first use
1. Before turning on the PC – connect RS232 cable and screw it down to avoid
disconnection during work.
2. Turn on the PC.
3. Install software from included CD/DVD: click on setup.exe to start Wizard
installation program and follow the steps.
4. After installation (before first start of UG-S/A) reset your PC
. Check if the load (ultrasound transducer) is connected properly.
6. Turn on the generator (4th socket on back panel)
7. Start UG-S/A software (UG-S/A.exe)
If the software is properly installed the following files are created as in the example:
<root> UG_SA