sky vision Linx 300 Bedienungsanleitung

Pre-Installation Manual
SkyVision Linx 300
Integration Control System
TEC-R-0049 REV2
08/05/2015

Distributed by:
US - SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512 (616) 656-2900
www.skytron.us
The base language for this document is ENGLISH. Any translations must be from the base language document.
Printed copies are not controlled documents.
Although current at the time of publication, SKYTRON'S policy of continuous development makes this manual subject to
change without notice. If current manuals are required, contact your local SKYTRON representative or contact SKYTRON
directly at the distribution addresses listed above.

Page 1
SkyVision Linx 300 Pre-Install• REV2
TABLE OF CONTENTS
TITLE PAGE
SECTION 1. INTRODUCTION ....................................................................................................... 3
SECTION 2. REQUIREMENTS...................................................................................................... 4
2-1. Environmental.......................................................................................................................... 4
2-2. Power ................................................................................................................................... 4
2-3. Grounding ............................................................................................................................... 4
2-4. Rack Mounting......................................................................................................................... 5
2-5. Display Power Management................................................................................................... 10
2-6. Optional Equipment ................................................................................................................ 10
2-7. Remote Rack Locations.......................................................................................................... 11
2-8. Connections............................................................................................................................ 12
2-9. Network .................................................................................................................................. 16
SECTION 3. Installation RESPONSIBILITIES ............................................................................. 18
3-1. SKYTRON Responsibilities .................................................................................................... 18
3-2. Customer Responsibilities ...................................................................................................... 18
SECTION 4. REVISION HISTORY................................................................................................ 23

Page 2
SkyVision Linx 300 Pre-Install • REV2

Page 3
SkyVision Linx 300 Pre-Install• REV2
SECTION 1.INTRODUCTION
The SKYTRON SkyVision Linx 300 Integration Control System installation can vary according to the
specic equipment needs of the OR facility. What is common in all installations is that video and data
input sources are routed to various outputs such as displays, video recorders, or networks providing
communication and video conferences.
This SkyVision Pre-Installation Manual provides an overview of typical installation considerations and
specic installation requirements associated with a SkyVision system.
Only SKYTRON certied technicians are authorized to install a system. The SkyVision installation team
may consist of two (2) or more technicians who coordinate with the hospital staff and local subcontractors
to complete an installation. Technicians follow specic installation details mandated by the customer and
specied in the SkyVision OPI (Order / Production / Installation) document. SKYTRON supports SkyVision
installations with thorough documentation, classroom training, and technical support.

Page 4
SkyVision Linx 300 Pre-Install • REV2
SECTION 2.REQUIREMENTS
2-1. Environmental
The SkyVision system must be stored and operated in a clean, dust free environment. Ambient room
temperatures should be maintained between 60˚ and 75˚ Fahrenheit with non-condensing humidity
between 20% to 50%.
The SkyVision Linx 300 system components will generate up to 2900 BTU's/hr. when fully equipped and
under maximum load during peak operating times. Mechanical and HVAC designs of the SKYTRON
Nurse Documentation Center (NDC), SORIC system, and Third Party provided versions must address
heating, ventilation and exhaust requirements in a manner consistent with the operational requirements
of the SkyVision system. When installing the system within custom cabinetry or a NDC, a minimum of
50 cfm through a 12" square opening is required for constant return HVAC. This will allow the systems
to operate within the equipment temperature specications and ensure product performance, reliability,
and longevity.
2-2. Power
The SKYTRON Linx 300 System requires two (2)120VAC/20A, GFCI protected outlets. One circuit is
dedicated to the component rack and the second circuit is dedicated to the monitor power supplies.
All receptacle outlet strips, adapters, receptacle covers, and faceplates shall be marked with the following
words or equivalent:
WARNING - TECHNICAL POWER
DO NOT connect to lighting equipment
For electronic equipment use only
60/120VAC
GFCI protected
The Uninterruptable Power Supply (UPS) that is provided for power conditioning on the Linx 300 system
is not designed for use with Isolated Power Supply systems.
2-3. Grounding
SKYTRON requires a minimum 10 AWG copper ground wire to be provided to the 12" x 12" junction box
for connection to the equipment rack.
Refer to Article 647.6(B) (Grounding Conducters Required for Sensitive Electronic Equipment) of the NEC.
Permanently wired utilization equipment and receptacles shall be grounded by means of an equipment
grounding conductor run with circuit conductors to an equipment grounding bus prominently marked
"Technical Equipment Ground" in the originally branch circuit panel board.

Page 5
SkyVision Linx 300 Pre-Install• REV2
2-4. Rack Mounting
The SKYTRON SkyVision System is offered in a single 18U wall mounted rack, that does NOT stand on
the oor, or as part of a NDC in two (2) 12U racks.
NOTICE
For Siesmic mounting specications, refer to the Middle Atlantic Products
letter for "Seismic Certication of the DWR Series Wall Mounted Electronics
Rack Enclosures" available on SkyNet.
a. Rack - 18U Wall Mount
Overall Width Overall
Height Overall Depth Weight
Capacity
Minimum
Structural
Requirements
Mounting Hole
Spacing
23-3/8”
[594mm]
38-1/2”
[978mm]
26-5/16”
[668mm] 250 lbs 500 lbs 16”
[406mm]
Minimum Clearance - 6” from non-hinged side, 20” from hinged side, 4” on top and bottom.
NOTICE
It is the responsibility of the contractor to provide sufcient backing/studs in
the wall to support the rack (the rack does NOT stand on the oor).
An alternative to mounting directly to the studs is to mount plywood to the entire wall and use toggle bolts
to mount the rack to the plywood.
Whatever method is used to mount the rack, the contractor is responsible for making sure it will hold the
minimum structural required load of 500 pounds.

Page 6
SkyVision Linx 300 Pre-Install • REV2
6" MIN.
CLEARANCE
WALL
WALL
WALL
26 5/16”
6-1/2"
20-1/2"
20" MIN.
CLEARANCE
90°
90°
(4) MOUNTING
HOLE LOCATIONS
A
B x B
CD
D
E
E
F
F
G
G
J
H
H
B
K
L
I
Letter Dimension
A38-1/2"
B12-1/2"
C22-3/4"
D 1-3/8"
E16"
F 5-3/8"
G 2-11/16"
H13"
I 4"
J33-1/4"
K 5-7/8"
L 4-3/8"
B
CONDUIT
BOXES
RACK
Figure 1. Rack Mounting and Junction Box Location (Left-Hand Swing Cabinet)

Page 7
SkyVision Linx 300 Pre-Install• REV2
WALL
6" MIN.
CLEARANCE
20" MIN.
CLEARANCE
6-1/2"
WALL
WALL
20-1/2"
90°
90°
26-5/16"
(4) MOUNTING
HOLE LOCATIONS
A
B x B
C
DE
G
G
H
H
B
I
J
K
L
DE
F
Letter Dimension
A38-1/2"
B12-1/2"
C22-3/4"
D 1-3/8"
E16"
F 5-3/8"
G 2-11/16"
H13"
I 4"
J33-1/4"
K 5-7/8"
L 4-3/8"
RACK
CONDUIT
BOXES
B
F
Figure 2. Rack Mounting and Junction Box Location (Righ-Hand Swing Cabinet)

Page 8
SkyVision Linx 300 Pre-Install • REV2
12" x 12" x 4"
JUNCTION BOX
(PROVIDED BY
CUSTOMER) TO BE
ROUGHED IN BY THE
CONTRACTOR.
MOUNT THE POWER
OUTLETS IN BOX AS
SHOWN.
10 AWG GREEN
STRANDED WIRE
Figure 3. Left-Hand Swing Installation with Junction Box Termination