Level Plus®Board Replacement
Operation Manual
I3 I
Fig. 3: LP Dashboard - Connect Screen
Fig. 4: LP Dashboard - Save Settings - Read from Gauge
3. Open LP Dashboard and connect to level transmitter. Select the
correct Protocol from the drop-down menu. Select the correct
COM Port. Do note that only active COM Ports will be shown so
make sure the converter is plugged into the computer before
opening the LP Dashboard. Select the correct Device Address.
Default address is 0 for HART®, 0 for SIL, 247 for Modbus, 192
for DDA, and 192 for USTD.
4. Click on the three white bars in the top left and select Save
Settings. Click Read Settings from Gauge. All of the boxes will
turn red. Wait for them to change two white signaling that the
parameters have been updated.
Fig. 5: LP Dashboard - Save Settings - File Save
Fig. 6: LP Dashboard - Save Settings - Write to Gauge
5. Save a backup file of the settings for the level transmitter by
clicking Save Settings to a File. Label the file with customer name,
site, and tank number. The board set will come with the factory
programming based off of the serial number but this will not
include any changes made by the end user after receipt.
6. Disconnect power to the tank before going to change the board
set.
7. Load backup file into new board set. This can be done in the
office before going on top of the tank. No SE or DT cable is
required to program the board set. Connect as shown in Step 2.
8. Connect to the level transmitter as shown in Step 3.
9. Click on the three white bars and select Save Settings. Click Read
Settings from File and select the previously saved file. Once the
file loads click Write Settings to Gauge.