EC910 Portable Multiparameter Water Quality Meter
Instruction Manual
AMTAST
AMTAST USA INC

EC910 Portable Multiparameter Water Quality Meter 1
Introduction
Thank you for selecting the AMTAST EC910 portable multiparameter water quality meter. This manual provides a -by-step guide to help you operate the
meter, please carefully read the following instructions before use.
UNPACKING THE METER:
Before unpacking, ensure that the current work environment meet following conditions.
●Relative humidity is less than 80%.
●Ambient temperature is greater than 0oC and less than 60oC.
●No potential electromagnetic interference.
The following list describes the standard accessories of the meter. After the unpacking, please check all accessories are complete. If any are damaged or
missing, please contact nearest distributor.
ACCESSORIES:
●pH Electrode
●Conductivity Electrode
●Dissolved Oxygen Probe
●Temperature Probe
●pH Buffer Pouches (pH4.01/7.00/10.01)
●Conductivity Calibration Solutions (84µS/cm, 1413µS/cm, 12.88mS/cm)
●Electrolyte Solution
●Membrane Cap
pH Electrode Conductivity Electrode Dissolved Oxygen Probe
Temperature Probe pH Buffer Pouches Conductivity Calibration Solutions
Membrane Cap Electrolyte Solution
DO
Electrolyte
Solution
Very Important:
If meter shows battery low, please stop use and change the battery ASAP, if you still use it, and meter
power off automatically, it will lost the system and need RETURN to us for install the system again.
The LCD will be blank and display nothing!!!

2
Display
EC910 portable multiparameter water quality meter is equipped with a clear and bright LCD display that used to show measured values, mode indicators
and help message. The following table describes the meaning of each indicator.
INDEX:
Measurement mode indicator:
Indicates meter is in the measurement mode
Low Battery Alarm:
When battery is depleted, the indicator will disappear
Calibration mode indicator:
Indicates meter is in the calibration mode
Stable indicator:
Indicates the measured value has stabilized
Setup mode indicator:
Indicates meter is in SETUP mode
Hold indicator:
Indicates the displayed value has been frozen
Memory mode indicator:
Indicates data is stored into memory
Calibration Due Reminder:
Prompts user to calibrate the meter regularly
Electrode slope indicator:
Indicates the average slope of the pH electrode
Automatic Temperature Compensation:
Indicates meter is in the temperature compensation mode
EC910 Portable Multiparameter Water Quality Meter
AMTAST

3
Keypad
The meter has a succinct membrane keypad, names and symbols describe the each function key controls.
INDEX:
KEY DESCRIPTION
MEAS|
●Power the meter ON/OFF.
●Freezes the measured value on the display, press the key again to resume measuring.
●In the calibration or setting mode, exits current mode and returns to measurement.
MODE|oC ●Toggles between available measurement modes (pH, mV, ORP, Ion, Conductivity, TDS, Salinity, Resistivity, DO).
●Press and hold the key to enter temperature setting mode.
CAL| ●Press the key to enter the calibration mode.
●Press and hold the key to enter the setup menu.
MI|
●Press the key to store current measured value.
●Press key in setup mode to scroll up through menu.
●Press key in temperature setting mode to increase the setting value.
MR|
●Press the key to view calibration report or stored data.
●Press key in setup mode to scroll down through menu.
●Press key in temperature setting mode to decrease the setting value.
ENTER| ●Confirms the calibration, setting value or displayed option.
●Press and hold the key to turn On/Off the backlight.
EC910 Portable Multiparameter Water Quality Meter

4
Connectors
EC910 portable meter provides 3 connectors for connecting the various types of sensors. Listed in the below table are the details of these connectors.
INDEX:
NO. CONNECTOR FUNCTION
1 BNC Connector For connecting the pH, ORP or Ion Selective Electrodes
2 Phone Jack For connecting the temperature probe
3 6-pin Connector For connecting the conductivity electrode or dissolved oxygen probe
Inserting the Batteries
Before using the meter, insert three 1.5V “AA” alkaline batteries into battery compartment, follow the procedure outlined below.
1. Remove the battery cover from meter’s backside.
2. Insert the batteries into the battery compartment, note polarity.
3. Replace the battery cover into its original position. Installation is completed.
When batteries are depleted, the meter allows you to use the USB cable connected to computer as a temporarily power supply.
2
1
3
EC910 Portable Multiparameter Water Quality Meter

5
Connecting the Sensor
Take out the sensor from the carrying case.
FOR THE pH, ORP AND ION SELECTIVE ELECTRODES:
●Insert the BNC connector into corresponding connector socket. Rotate and push the connector clockwise until it locks.
FOR THE CONDUCTIVITY AND DISSOLVED OXYGEN ELECTRODES:
●Insert the 6-pin connector into corresponding connector socket, ensure the connector is fully seated.
●After connection is completed, DO NOT pull on the sensor cord. Always make sure that the connector is clean and dry.
EC910 Portable Multiparameter Water Quality Meter

6
Preparing the pH Buffer Solutions
EC910 portable multiparameter water quality meter is packaged with three pH buffer packets required for calibration.
●Open the pH7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until
reagent is completely dissolved.
●Preparation of pH4.01 and 10.01 standard buffer solutions are the same as above.
●Prepared standard buffer solutions should be stored in hermetically sealed glass containers.
Refilling Electrolyte Solution for DO Probe
1. Unscrew the membrane cap from the bottom of the dissolved oxygen probe.
2. Fill the membrane cap halfway with electrolyte solution.
3. Screw the membrane cap onto the probe, excess electrolyte will drain out.
4. Be sure the cathode of probe makes contact with membrane cap, the electrolyte solution in membrane cap should be without an air bubble.
…………Cathode
EC910 Portable Multiparameter Water Quality Meter

7
Prior to Use
Remove the protective cap from the bottom of the sensor.
●FOR THE pH ELECTRODE:
If the glass sensitive membrane on the electrode dries out, soak the electrode in 3M KCL solution or tap water for atleast 15 minutes. DO NOT use
distilled or deionized water, it will shorten the life of sensor.
●FOR THE ORP ELECTRODE:
Soak the electrode in3M KCL solution or tap water for 5 minutes.
●FOR THE ION SELECTIVE ELECTRODES:
Soak the electrode in the diluted standard solution (e.g., 100ppm) for at least 15 minutes.
●FOR THE CONDUCTIVITY ELECTRODE:
Soak the electrode for a few minutes in tap water to remove dirt and oil stains on the electrode.
●FOR THE DISSOLVED OXYGEN PROBE:
Connecting the probe to meter, turn on the meter 10 to 15 minutes and wait for the probe to polarize.
Power On/Off
●Press MEAS key to turn on the meter, the display shows measured values, mode indicators and help messages.
●Press and hold the MEAS key for 3 seconds, the meter will turn off.
●If you do not press any key within the specified time period, the meter will turn off automatically.
To disable the auto-off function, please read the SETUP MENU section.
……..…. Glass Sensitive Membrane
EC910 Portable Multiparameter Water Quality Meter

8
Setup Menu
EC910 portable multiparameter water quality meter contains an integrated setup menu that allows you to customize each displayed option to meet
measurement requirements. In the different modes, the meter will show the corresponding options. For the general options, once it is changed that will
be applied to all modes.
pH MODE:
PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT
pH Buffers:
Select your desired pH buffer group for calibration
and auto-recognition.
USA (1.68, 4.01, 7.00, 10.01, 12.45) ●
NIST (1.68, 4.01, 6.96, 9.18, 12.45)
DIN (1.09, 4.65, 6.79, 9.23, 12.75)
User-Defined Buffer
Calibration Points:
Select the number of calibration points you intend
to calibrate the meter.
1 point
2 points
3 points ●
4 points
5 points
Resolution:
Sets the resolution for pH measurement.
0.001pH ●
0.01pH
0.1pH
Measurement Unit:
Sets the default temperature unit.
Degrees Celsius ●
Degrees Fahrenheit
ORP MODE:
PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT
Resolution:
Sets the resolution for ORP measurement.
0.1mV ●
1mV
ION MODE:
PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT
Measurement Unit:
Sets the default ion concentration and temperature
units.
ppm Parts per million ●
mg/L Milligrams per liter
mol/L Moles per liter
Degrees Celsius ●
Degrees Fahrenheit
EC910 Portable Multiparameter Water Quality Meter

9
Calibration Points:
Select the number of calibration points.
2 points ●
3 points
4 points
5 points
Ion Valence:
Select the ion valence of sensor.
Monovalent ●
Divalent
CONDUCTIVITY/TDS/SALINITY/RESISTIVITY MODES:
PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT
Cell Constant:
Select the cell constant of conductivity electrode.
K=0.1
K=1 ●
K=10
User-Defined
Temperature Coefficient:
Select the temperature coefficient of sample. Setting Range: 0.0 to 10.0%/oC 2.10
Calibration Points:
Select the number of calibration points.
1 point ●
2 points
3 points
4 points
5 points
Pure Water Coefficient:
Pure water coefficient will be calculated and applied
automatically for ultra pure water measurement if
enabled.
Enable
Disable ●
Normalization Temperature:
Sets the normalization temperature for conductivity
measurement and calibration.
25oC ●
20oC
TDS Factor:
Sets the default TDS conversion factor. Setting Range: 0.40 to 1.00 0.5
Measurement Unit:
Sets the default temperature unit.
Degrees Celsius ●
Degrees Fahrenheit
EC910 Portable Multiparameter Water Quality Meter
Inhaltsverzeichnis
Andere AMTAST Messgerät Handbücher

AMTAST
AMTAST MB65 Bedienungsanleitung

AMTAST
AMTAST AMT605 Bedienungsanleitung

AMTAST
AMTAST RHS-4ATC Technisches Handbuch

AMTAST
AMTAST UV340B Bedienungsanleitung

AMTAST
AMTAST AMT-316 Bedienungsanleitung

AMTAST
AMTAST DO901 Bedienungsanleitung

AMTAST
AMTAST GM006 Bedienungsanleitung

AMTAST
AMTAST AMT27 Bedienungsanleitung