Density 3.0 (D3) - Calibration:
Place the D3 filter in die recess instead of the normal aperture and press
the arm downwards. A measurement will be triggered. Now continue to
hold the arm down and at the same time press the S2 button to start the
D3 calibration. Keep the arm down until the density value is displayed,
which means the calibration routine is finished. This can take up to 4
seconds. Now return the arm back to the horizontal position.
Settings-Menu
You can access the densitometer settings menu by pressing the S2 button
for three seconds, while the arm is in the upper, horizontal position (see
Fig.: 5 main menu). Press the S1 button to scroll through the menu and
press the S2 button to selection a menu setting. Choose "Exit" to navigate
to the main menu and to leave the settings menu altogether (see Fig.: 6 -
exit and Fig.: 7 - Exit 2).
(Bild 5 –Main menu)
(Bild 6 –Exit) (Bild 7 –Exit 2)