Dillon Quantrol BFG 10 Bedienungsanleitung

Basic Force Gauge
User’s Manual

2Basic Force Gauge User’s Manual
Risk of electrical shock. Do not remove cover. No user serviceable
parts inside. Refer servicing to qualified service personnel.
Weigh-Tronix reserves the right to change
specifications at any time.
CAUTION
EUROPEAN COUNTRIES
WARNING
This is a Class A product. In a domestic environment this
product may cause radio interference in which the user
may be required to take adequate measures.
02/04/03 BFG_U.P65 PN 29798-0013A e1 Printed in USA

3
Basic Force Gauge User’s Manual
Table of Contents
Table of Contents ........................................................................................... 3
Specifications ................................................................................................. 4
Capacity and Divisions ........................................................................... 4
Accuracy ................................................................................................. 4
Output ..................................................................................................... 4
Introduction .................................................................................................... 5
BFG Power Supply ................................................................................. 5
Operation ....................................................................................................... 6
Fitting Accessories ................................................................................. 6
Mounting to a test stand ......................................................................... 6
Powering up ............................................................................................ 6
Display of Tension/Compression ............................................................ 7
Zeroing the Gauge ................................................................................. 8
Changing the unit of measurement ........................................................ 8
Max (Peak) Readings ............................................................................. 8
“Max” mode ..................................................................................... 8
Dual Max ......................................................................................... 9
Max Tension .................................................................................... 9
Max Compression ........................................................................... 9
“Normal” mode ................................................................................ 9
Data Output ................................................................................... 10
Optional Settings .................................................................................. 10
Backlit Display ............................................................................... 10
Auto-off .......................................................................................... 10
Invert Display Text ......................................................................... 11
Load cell Diagnostic test ............................................................... 11
Appendix A ................................................................................................... 13

4Basic Force Gauge User’s Manual
Specifications
Model No: mN N kN g-f kg-f oz-f* lb-f*
BFG 10 10000 x 2 10 x 0.002 - 1000 x 0.2 1 x 0.0002 35 x 0.01 2.2 x 0.0005
BFG 50 50000 x 10 50 x 0.01 - 5000 x 1 5 x 0.001 180 x 0.05 11 x 0.002
BFG 200 - 200 x 0.05 - 20000 x 5 20 x 0.005 720 x 0.2 44 x 0.01
BFG 500 - 500 x 0.1 - 50000 x 10 50 x 0.01 1800 x 0.5 110 x 0.02
BFG 1000 - 1000 x 0.2 1 x 0.0002 - 100 x 0.02 3500 x 1 220 x 0.05
BFG 2500 - 2500 x 0.5 2.5 x 0.0005 - 250 x 0.05 - 550 x 0.1
Capacity and Divisions
Accuracy ±0.25% of full-scale, ±least significant digit
Calibration temperature: 20°C ±2°C
Operating temperature: 10°C - 35°C
Temperature shift at zero load: ±0.09% of full-
scale/°C
RS-232: 8 data bits, 1 Start bit, 1 Stop bit, no parity
Digimatic (Mitutoyo) format
Analog: 2.5V at zero load, ±1.5V approximately for
full-scale tension/compression
Voltage obtained from analog conditioning
of the load cell. Actual voltages will vary by
capacity and also by individual instruments.
Actual voltage can be captured by applying
reference load and measuring or capturing
voltage.
Output
* Due to internal rounding, occassional jumps of twice the increment size may occur.

5
Basic Force Gauge User’s Manual
Introduction
BFG Power
Supply
When alkaline
batteries are fitted, the
AC adaptor/charger
must NEVER be
connected to the BFG
due to risk of acid
leakage which could
damage the instru-
ment.
The QuantrolTM Basic Force Gauge (BFG) can
measure tensile and compressive forces accurately
while still being easy to operate.
Please check the unit for any physical damage. If any
damage is found, please notify your authorized
Quantrol dealer.
The most commonly used features of the BFG are
accessed with a single key press on the instrument:
• displaying force
• peak hold
• zero
• changing units of measure
Other features can be accessed easily by pressing
and holding different 2 key combinations.
The BFG comes with five Nickel Metal Hydride AAA
rechargeable batteries. For safety reasons the
batteries are shipped discharged. To obtain maximum
battery life we recommend that you charge them with
the charger/adaptor supplied for at least 14 - 16 hours
when you first receive the BFG.
To insert the batteries you must first remove the
battery cover on the rear of the gauge by unscrewing
the two retaining screws. Fit the batteries in the
battery holder. Be sure to match the marked polarity.
Refit the battery cover and tighten the screws.
Connect the AC adaptor/charger to the BFG charger
socket and charge the batteries for 14 - 16 hours.
Only use the adaptor/charger supplied.
The BFG can also be powered directly from the AC
adaptor. You can do this with or without the recharge-
able batteries installed. Connect the AC adaptor/
charger to the BFG and a properly grounded outlet of
the correct voltage. Only use the adaptor/charger
supplied.
If the rechargeable batteries are installed, a trickle
charge will continue during AC use of the unit.
The BFG can also be powered by AAA 1.5V alkaline
batteries (not supplied). Follow the same instructions
as above, except for use of the charger. See warning
at left.

6Basic Force Gauge User’s Manual
Operation
Fitting
Accessories
Mounting to a
test stand
Powering up
Affix either the short extension rod (30mm long) or the
long extension rod (130mm long) to the load cell
probe in the hole at the bottom of the gauge by
tightening it gently with the fingers.
Your chosen grip or accessory may now be connected
to the extension rod.
When fitting a grip ensure that it is screwed finger-
tight only. Excessive torque can damage the load
cell.
On the rear of the gauge there are two M5 threaded
holes, which can be used for mounting the gauge to a
Quantrol test stand.
Each Quantrol test stand is supplied with a dedicated
‘dovetailed mounting bracket’ and screws for this
purpose.
If you wish to mount the gauge to another type of
stand, ensure that the screws used are threaded into
the gauge to a maximum depth of .21 inch (5.5mm).
The final thread is peaned but if screws are fitted
beyond this depth, damage to the load cell will occur.
As shown in Figure 1 the control panel has 6 keys.
Figure 1
BFG keys

7
Basic Force Gauge User’s Manual
A BFG measuring very
low forces may not
show zero if it is
moved during the self
test routine. Once it is
properly mounted and
zeroed the reading will
be stable.
To power up the gauge press the ON/OFF key. A
short self test runs during which the display will
show the model and capacity in Newtons.
After the self test, providing no load has been
applied to the instrument, the display will show all
zeroes. This is because the gauge re-zeroes itself
during the self test routine. See note at left.
If a force is applied perpendicularly via the sensor
probe (hole at bottom of BFG), the reading on the
display will register the applied force.
Forces greater than 120% of full-scale will produce an
audible beep until load is released and an OL symbol
will appear on the display for 30 seconds.
Forces greater than 150% of full-scale will produce an
audible beep until load is released and an OL symbol
will appear permanently on the display. Consult your
Quantrol distributor to arrange repair.
To power down the BFG press the ON/OFF key.
When tensile forces are applied, the tension symbol
shown in Figure 2 is displayed.
Figure 2
Tension symbol
When compression forces are applied, the compres-
sion symbol shown below is displayed.
A load indicator bar alerts the operator to how much
load has been applied to the transducer. As the load
approaches the maximum rating of the transducer, the
indicator bar changes appearance when above
approximately 80% of the rated capacity. This warns
you that steps should be taken to prevent overloading
Do not overload the
transducer, as this will
cause irreparable
damage.
Display of
Tension/
Compression
If the BFG has
suffered a serious
overload condition, the
load indicator bar will
be partially displayed
even when no load is
present. This is a
warning that the load
cell is damaged and
you should immedi-
ately contact your
Quantrol distributor to
arrange repair.
The load indicator bar
is independent of the
zero function and thus
provides good visual
reference of potential
overload conditions.

8Basic Force Gauge User’s Manual
Zeroing the
Gauge
Any load applied prior
to zeroing must be
considered as part of
the full-scale capacity.
Changing the unit
of measurement
Max (Peak)
Readings
“Max” mode
the transducer.
For tensile forces the indicator bar is solid then
dotted. For compressive forces the indicator bar is
dotted then solid. See Figures 3b & 3c
During the operation of the gauge it is often necessary
to zero the display – e.g. when you wish to tare out
the weight of a grip, so it does not become part of the
measured reading. Press and release the ZERO key.
The display will blink momentarily as the zero opera-
tion is carried out.
You can choose from the following units of measure-
ment depending on the capacity of your gauge:
milliNewtons, Newtons, gram-force, kilogram-force,
ounce-force or pound-force.
To change the display units press and release the
UNITS key. Each successive key press will select the
next available unit of measure until the gauge returns
to its original setting. The BFG automatically converts
readings as new units of measure are selected.
The gauge detects and stores maximum (peak) force
in both compressive and tensile directions.
Press the MAX key. The display will show the word
MAX together with the highest tensile force and the
highest compressive force detected during the test.
The current load being applied to the transducer is
also displayed.

9
Basic Force Gauge User’s Manual
Dual Max
Max Tension
“Normal” mode
Max Compression
Figure 3a
MAX dual readings
Press the MAX key again and the display will show
the maximum tensile force identified by the symbol.
Figure 3b
MAX tension
Press the MAX key again and the display will show
the maximum compressive force identified by the
symbol.
Figure 3c
MAX compression
Press the MAX key again and the word MAX disap-
pears from the display. The display indicates forces
applied in both directions as they are applied to the
transducer and maintains a running display.
Dieses Handbuch passt für folgende Modelle
5
Inhaltsverzeichnis
Andere Dillon Messgerät Handbücher

Dillon
Dillon AP Dynamometer Bedienungsanleitung

Dillon
Dillon ED-2000plus Bedienungsanleitung

Dillon
Dillon FI-80 LCD Bedienungsanleitung

Dillon
Dillon FI-127 Bedienungsanleitung

Dillon
Dillon AFG Series Bedienungsanleitung

Dillon
Dillon FI-90 Bedienungsanleitung

Dillon
Dillon EDXtreme Bedienungsanleitung

Dillon
Dillon GTX-Plus Series Bedienungsanleitung

Dillon
Dillon EDXtreme EDX-1T Bedienungsanleitung

Dillon
Dillon AP Bedienungsanleitung