5
TOTAL (Accumulator)
The TOTAL annunciator is lit when the accumulation function is active. The accumulator
will track the number of weights and total amount of weight on the scale. Do not change
units during an accumulation process.
Accumulate
1. Place item on the scale
2. Press TOTAL to accumulate current weight
3. Add--- is displayed while saving to memory
4. Remove item from scale
5. Weight must return to zero before the next weight can be added to the
accumulated total.
Display Accumulator
1. Press and hold TOTAL until total is displayed
2. Cn is displayed showing the current count
3. Press GROSS to display the total accumulated weight
4. Press PRINT to exit to weighing
Clear Accumulator
1. Press and hold TOTAL until total is displayed
2. Cn is displayed showing the current count
3. Press GROSS to display the total accumulated weight
4. Press ZERO to clear the accumulator
5. Press PRINT to exit to weighing
COUNT (Piece Counting Function)
Press COUNT to switch between weight and count display. When displaying piece
count, a nis displayed in the leftmost digit. Parameter F2.4 controls the method of
piece count entry. F2.4 = 0 will use platform weight to calculate piece weight. F2.4 = 1
will use manual piece weight entry.
Set piece weight with platform weight (F2.4 = 0)
1. Place the container on the scale
2. Press ZERO
3. Place the piece sample into the container
4. Press and hold COUNT until SANPLE is displayed
5. Press PRINT
6. Enter the number of pieces on the scale
7. Press PRINT
8. Piece count is displayed