Druck DPI 705E Bedienungsanleitung


Copyright 2019 Baker Hughes Company.
English–137M7122 DPI 705E / DPI 705E IS User Manual | i
Introduction
This user manual covers the DPI 705E / DPI 705E-IS measurement instrument and its
associated remote probes, PM 700E / PM 700E-IS for pressure measurement and
RTD-INTERFACE / RTD-INTERFACE-IS for temperature measurement.
Unless specifically stated in a section all references to DPI 705E cover both DPI 705E (non-
Intrinsically Safe) and DPI 705E-IS (Intrinsically Safe). This also applies to PM 700E, PM 700E-
IS and RTD-INTERFACE, RTD-INTERFACE-IS.
DPI 705E Series Digital Pressure Indicator
The Druck DPI 705E pressure indicator uses a silicon transducer to produce a pressure reading
in various pressure measurement units. These user instructions include the operations for all
DPI 705E Pressure Indicators, safety instructions and the requirements for intrinsically safe
instruments.
The PM 700E provide a remote pressure sensor function for the DPI 705E to allow more
convenient connection when direct connection of the handheld instrument would be
inconvenient. The PM 700E is a “plug and play” device holding its own calibration data, so it can
be used to have multiple pressure range sensors available for immediate use with a single
DPI 705E instrument.
The RTD-INTERFACE is a remote interface for use with the DPI 705E or the DPI 620G to allow
connection of a PT 100 RTD probe to the instrument for the purpose of measuring temperature.
DPI 705E DPI 705E-IS
RTD-INTERFACE
PM 700E (Differential)
PM 700E (Absolute, Gauge)

Copyright 2019 Baker Hughes Company.
ii | 137M7122 DPI 705E / DPI 705E IS User Manual–English
For all accuracy specifications etc refer to the relevant product datasheet, available for download
from Druck.com/essential.
Commercial and Intrinsically Safe Product versions
The DPI 705E is available as either non-Intrinsically Safe (blue in color) or Intrinsically safe
(yellow in color) versions, with case coloring used as well as labeling to indicate the version.
The remote pressure sensor PM 700E and the remote RTD interface RTD-INTERFACE are
available as either non-Intrinsically Safe or Intrinsically safe versions. The outer metal body of the
product is marked with its part number and relevant approvals.
Safety
For all safety information on both commercial and Intrinsically Safe versions refer to the Quick
Start and Safety Manual provided with the unit and also available for download on
Druck.com/essential.
Pressure Connectors and Adaptors
The DPI 705E and PM 700E are fitted with female G1/8 pressure connectors for the main
pressure port P1 for ranges 25 mbar - 200 bar(0.36 psi - 2900 psi). For pressures 350 bar-
1400bar(5000 psi - 20000 psi) Autoclave male pressure connectors are used. The main
pressure port P1 is shown on the left in the picture below.
For differential units/sensors fitted with a reference port a Legris adaptor is fitted giving a 4mm
tubing fitting. This is shown as the Ref port on the right in the image below.
Various adaptors are available (either as accessories or configured when ordering) for the G1/8
connection to convert to other standards. Please see datasheet for full details.
Part Code Description
IO-ADAPT-G1/4 G1/4 female adaptor
IO-ADAPT-1/4NPT 1/4 NPT adaptor
IO-ADAPT-1/8NPT 1/8 NPT adaptor
IO-ADAPT-QF Quick-fit hose adaptor

Copyright 2019 Baker Hughes Company.
English–137M7122 DPI 705E / DPI 705E IS User Manual | iii
See below for cautions on adaptor and connector fitting:
Batteries
For compatible battery types refer to Quick Start and Safety Manual provided with the unit and
also available for download on Druck.com/essential
The instrument requires 4 xAA/LR6 batteries, fitted as shown below. Use a small Posidrive
screwdriver to loosen the screws in the battery compartment cover and fit as per the battery
orientation shown on the back of the instrument as illustrated below:
1 DPI 705E or PM 700E G1/8 port
2 Bonded seal 110N3018
3 ISO 228/1 G1/8
4 NPT thread pressure connector IO-ADAPT-1/4NPT or IO-ADAPT-1/8NPT
5 NPT female to G1/8 male adapter part number IO-ADAPT-1/4NPT
τ
1P
LQOE
4
4
3

Copyright 2019 Baker Hughes Company.
iv | 137M7122 DPI 705E / DPI 705E IS User Manual–English
External Sensors
The DPI 705E supports two types of external sensor, both of which connect via a cable into the
right-hand side of the unit.
The sensor must be connected via supplied connection cable (it will not engage correctly if
directly plugged in). The connectors push in (with polarity protection features aligned) then the
outer body rotates clockwise to lock the connector in position. Multiple cables may be used in
series (i.e. as extenders). Up to a maximum of 10 in series if very long cabled connections are
needed.
When an external sensor is connected it is automatically detected and has precedence over the
internal sensors. The display shows “ ” (External Sensor) to show that the reading
is from the external sensor. To show a reading from the internal pressure sensor the external
sensor must be unplugged.
The external sensor may be either:
• A pressure sensor PM 700E ranging from 25 mbar to 1,400 bar (0.36 psi to 20,000 psi)
full-scale.
• A temperature sensor adaptor RTD-INTERFACE which supports a 4-wire PT100 probe.
If a PM 700E pressure sensor is plugged in then the pressure Full-scale will be displayed (only
at first plug-in), followed by the calibration due date in days. Normal pressure measurement
display will then start.
If an RTD-INTERFACE temperature sensor is plugged in then temperature/resistance full-scale
will be displayed (only at first plug-in), followed by the calibration due date in days. Normal
resistance or temperature measurement display will then start.
RTD Wiring
The DPI 705E M12 connector is to be connected to an external RTD probe as shown below.
The RTD-INTERFACE is optionally supplied with a field-rewirable M12 connector for customers
to connect their own wire-ended RTD's into. This is accessory part number IO-RTD-M12CON.
External sensor connector

Copyright 2019 Baker Hughes Company.
English–137M7122 DPI 705E / DPI 705E IS User Manual | v
The pin numbering is printed on the rear of the connector body and is reproduced below for
clarity.
Controls and Functions
Function Control Description
Power Turn the instrument on or off.
Filter Use to get a stable pressure reading over a noisy
pressure source.
It has a secondary function as a Back key.
Leak Use to measure pressure leaks over a configurable
time interval
It has a secondary function as an OK key.
Tare Set a temporary zero for relative measurements by
using the Tare feature.
It has a secondary function as an Increment key.
Units Change the measurement unit of the instrument
It has a secondary function as a Next key.
Backlight Turn the instrument backlight on/off.
Alarm Set a high and low alarm based on a high/low
configurable threshold.
(Press Simultaneously)
Zero Set the Zero point for your gauge or differential sensor
before performing measurement.
(Press Simultaneously)
Settings Perform Advanced operations such as sensor
calibration, configure calibration interval, units, user
PIN, display lock behavior and factory reset. Also
check sensor full-scale values, system date/time,
software version and other General settings.
(Press Simultaneously)
Customer M12 connector pinout
Connector front Connector rear (wiring side)

Copyright 2019 Baker Hughes Company.
vi | 137M7122 DPI 705E / DPI 705E IS User Manual–English
Display Segments
The DPI 705E display segment is explained below:
1. Leak Test Indicator 2. Battery Level Indicator
3. Filter On Indicator 4. External Sensor Functional Indicator
5. Calibration Due Indicator 6. Remote State Indicator
7. Internal Sensor Functional Indicator 8. Lock Indicator
9. Alarm State Indicator 10. Temperature Degree Indicator
11. Resistance (Ohms) Indicator 12. Custom Units Indicator
13. Pressure Units Indicator 14. Tare Indicator
15. Sensor Type Absolute/Gauge Indicator 16. Minimum Symbol Indicator
17. Negative Value Indicator 18. Maximum Symbol Indicator
1
18
17
16
15
14
13 12
11
2 3 4 567 8 9
10

Copyright 2019 Baker Hughes Company.
English–137M7122 DPI 705E / DPI 705E IS User Manual | vii
Contents
1. Basic Functionality 1
1.1 Turn on/off 1
1.2 Pressure Measurement 2
1.3 Temperature Measurement 2
1.4 ZERO 3
1.5 TARE 4
1.6 FILTER 4
1.7 UNITS 4
1.8 Backlight 5
2. Advanced Functionality 5
2.1 Maximum Reading 5
2.2 Minimum Reading 6
2.3 Leak Test 7
2.4 Alarm 8
2.4.1 Set High Alarm 9
2.4.2 Set Low Alarm 10
3. Settings 12
3.1 Advanced Settings 13
3.1.1 Adjust Sensor Calibration 13
3.1.2 Calibrate Sensor 15
3.1.3 Set Calibration Date 17
3.1.4 Set Calibration Due Interval 17
3.1.5 Change User PIN 18
3.1.6 Units Lock 20
3.1.7 Lock Display 21
3.1.8 Factory Reset 22
3.2 General Settings 23
3.2.1 Full-scale Value 24
3.2.2 Calibration Due Date 24
3.2.3 Audio 25
3.2.4 Auto shutdown 26
3.2.5 Battery Type 27
3.2.6 Date 28
3.2.7 Time 30
3.2.8 Custom Unit 31
3.2.9 Software Version 33
4. Accessories 35
5. Error Codes & Diagnostics 36
5.1 Error Codes 36
5.2 Diagnostics 37
6. Approved Service Centers 37
6.1 Return Goods/Material Procedure 37

Copyright 2019 Baker Hughes Company.
viii | 137M7122 DPI 705E / DPI 705E IS User Manual–English
Andere Handbücher für DPI 705E
2
Dieses Handbuch passt für folgende Modelle
1
Inhaltsverzeichnis
Andere Druck Messgerät Handbücher

Druck
Druck DPI 705E Bedienungsanleitung

Druck
Druck DPI 800 Bedienungsanleitung

Druck
Druck DPI 705E Bedienungsanleitung

Druck
Druck PACE1000 Bedienungsanleitung

Druck
Druck LR-Cal TLDMM 2.0 Series Bedienungsanleitung

Druck
Druck DPI 705 Gebrauchs- und Pflegehandbuch

Druck
Druck IDOS UPM Bedienungsanleitung

Druck
Druck DPI 705 Gebrauchs- und Pflegehandbuch