HPG Manual.docx 4 12/01/2018
HPG Series Diode Pumped Solid State Laser
1. Basic Operation
The HPG Series Lasers are microprocessor controlled, and designed for simple,
user friendly operation from most PC type computers through the serial (RS232)
port.
Ensure that the Laser Head is connected correctly to the Power Supply by means of
the umbilical cable supplied.
NB: Care should be taken to correctly align the connector before pushing home and
turning the locking ring
Connect the mains supply of the appropriate voltage to the Power Supply by the
connector on the rear panel.
For HPG 532nm lasers connect the two pin plastic PZT connector at the power
supply rear panel and laser head.
Connect the 9-way serial cable supplied to the RS232 D-connector on the rear of the
Power Supply, and the USB serial converter (supplied).
1.1 Operation through RS232
Before laser operation it is necessary to follow the set up
procedure for your PC as shown in appendix 1
The connection detail for the RS232 cable is given later in this manual.
Ensure that the external interlock circuit is closed, either by use of the supplied
shorted connector (temporarily for test only), or via a user supplied door or
enclosure protection circuit.
Boot the computer, if not already running, and run a dumb terminal program such as
HYPERTERMINAL (supplied with Win 98/ME/2000/XP) and accessible from the
Windows Start Menu and following:
Programs > Accessories > Communication > HyperTerminal