ERTE B3 Bedienungsanleitung

www.erte.com.tr
B3 INDICATOR
Version P.31
DE-19-NAWID-PTB006 R76/2006-A-DE1-2019.06

ERTE® Endustriyel Elektronik Sanayi e Ticaret Limited Sirketi
Address IZMIR / TURKEY
Telephone +90 232 2 9 7400
Fax +90 232 2 9 3960
E-mail [email protected]
Web site www.erte.com.tr

Contents
1. MATTERS TO CONSIDER........................................................................................................................... 2
2. INTRODUCTION.......................................................................................................................................... 3
Interface....................................................................................................................................................... 3
Keys and Functions................................................................................................................................ 3
Signs and Meanings............................................................................................................................... 4
Connection Diagram....................................................................................................................................
Notes...................................................................................................................................................... 6
Dimensions.................................................................................................................................................. 7
3. WEIGHING PARAMETERS AND ADJUSTMENT.........................................................................................8
Weighing Parameters.................................................................................................................................. 9
Adjustment................................................................................................................................................. 10
4. SER MENU (SER2, SER3)......................................................................................................................... 11
. ALIBI MENU................................................................................................................................................ 12
6. LANGUAGE AND INFORMATION MENUS................................................................................................13
Language Menu......................................................................................................................................... 13
Information Menu....................................................................................................................................... 13
7. ERROR AND WARNING MESSAGES....................................................................................................... 14
1

1. MATTERS TO CONSIDER
●The device must be operated from an earthed socket.
●It should not be connected to the same power line with powerful electrical
machines and should not be operated near
●Make sure that the electrical connection is cut before the cover of the
device is disassembled.
●Recommended cable thickness for load cell connection is 1 mm². If it is
necessary to use thinner cable, it should be thickened by using double (or
more) ends for ±supply. This is especially important for long cable
connections.
●If the load cell/collection box connection does not have sensing leads (4-
way connection) device will give Error 3. In such cases, connect the
sensing leads to the supply terminals at the closest point to the load cell.
(See wiring diagram.)
●Shielded and twisted-pair cable for load cell and communication
connections should be used.
●For the most accurate weighing results, the device should be left in
operation for at least half an hour before calibration.
2

2. INTRODUCTION
Interface
* Display digit height is 2 mm
Keys and Functions
KEY EXPLANATION APPLICATION
FUNCTION
MENU
FUNCTION
NUMERIC INPUT
FUNCTION
SELECT FROM
LIST FUNCTION
Zero Key Zero weight Exit Menu Reset value Select 0 if it is in
the list
Tare Key
(Scroll) Tare in/out Show Previous
Menu
Go to the next
digit Show previous
Print Key
(Increase)
Print,
save to alibi
memory
Show Next Menu Increase value Show next
Star Key
(Enter) Enter Menu Enter Menu Approve value Select
Function
Key
(Exit)
Differs by
application Exit Menu Cancel Cancel
3

Signs and Meanings
R1, R2 Range indicator: If configured as a multi range instrument, indicates the current weighing
range.
Center-of-zero indicator: Indicates the deviation from zero is not more than ±0.2 e.
NET Net indicator: Indicates that a tare device is in operation and the displayed weight value is a
net weight.
Stability (No-motion) indicator: Indicates that stability of equilibrium has been reached.
kg Unit of weight: Indicates the unit of the weight, if a weight value is displayed.
Battery : Indicates the battery level in devices with battery option. When it starts to flash, the
device should be charged as soon as possible.
4

Connection Diagram
On/Off Switch
Power 8 – 264 VAC, 47 - 63 Hz, 0.2 A
Note: Make sure to connect the device to a grounded socket
Load Cell
Pin Explanation
1 - Signal
2 + Signal
3 - Sense (Note: Connect to - Excitaion if load cell sense not available)
4 + Sense (Note: Connect to + Excitaion if load cell sense not available)
Chassis
6 + Excitation
7 - Excitation
Remote Display
Pin Value ERTE Remote Display Connection
1, 2, 3 Not connected Not connected
4, ,
6, 7,
8, 9
4 4
6 6
7 7
8 8
9 9
RS-232
Pin Value PC Connection
2 Tx 2
3 Rx 3
7Gnd
USB USB PC Connection
Note: USB driver must be downloaded and installed from www.erte.com.tr before the
cable is connected.
5

Notes
•In this position the calibration is OFF / LOCKED
•In this position the calibration is ON / UNLOCKED.
•3. Optional RS-232 output must be connected from Op1
•Relay option must be connected from Op2
6

Dimensions
7

3. WEIGHING PARAMETERS AND ADJUSTMENT
In order to setup weighing properties, follow these steps:
• Switch off the device and disconnect the power cable.
• Dismount the front panel and put the adjustment switch in UNLOCKED position.
• Mount the front panel and switch on the device.
• Perform weighing setup.
• When the weighing setup is done, switch off the device and disconnect the power cable.
• Dismount the front panel and put the adjustment switch in LOCKED position.
• Mount the front panel and switch on the device.
Note: Device is shipped with the switch in UNLOCKED position.
While the device is powered with UNLOCKED adjustment switch, functions of the keys are as following:
Press this key to enter menu. When pressed, first menu item is displayed. Press
key to display next item, press or to exit menu.
pAr Weighing parameters
AdJuSt Calibration
When this key is pressed, the weight value is displayed in increased resolution
(increment/10). In this position, the weight unit sign (kg) goes out. When the
calibration switch is in the off position, it automatically returns to the normal display
after seconds.