GRAPHTEC midi LOGGER GL800 Bedienungsanleitung

GL800 Quick Start Guide
midi LOGGER
GL800-UM-851
Checking the Outer Casing
Checking the Accessories
Setting and Checking the AC Line Frequency
Quick Start Guide : 1 CD-ROM : 1
AC cable/AC adapter : 1
Set the AC line frequency in the "OTHR" menu.
This setting (50 or 60 Hz) affects the device's noise elimination
capability. elimination capability.
After unpacking, check the GL800's outer casing before use to make sure that there
are no surface scratches or other flaws such as stains or dirt.
Don't forget to
check the setting


GL800 Contents
Part Names............................................................................. 2
Connection Procedures....................................................... ... 3
Precautions to Observe When Performing Measurement...... 4
Descriptions of the Control Panel Keys............................... ... 5
Descriptions of the Menu Screens......................................... 8
Measurement Procedure........................................................ 9
1.
Preparations : How to Make the Preparations Required for Data Capture
.......... 9
2.
Setup : How to Make the Settings for Temperature Measurement
............. 10
3.Data Capture : How to Measure the Temperature.......................... 13
4.Data Replay : How to Replay Captured Data.............................. 14
Convenient Functions........................................................... 15
Trigger Functions to Control Data Capture Start/Stop Operations....... 15
Span, Position and Trace Functions to Adjust the Waveform Display
.............. 17
Specifications..................................................................... 18
Standard Specifications.......................................................... 18
External Input/Output Functions................................................. 18
Input Unit Specifications.......................................................... 19
Installation Guide............................................................... 20
1

GL800 Part Names
Top Panel
Bottom Panel
Monitor
Battery cover
2
Control panel keys
Analog signal input terminals
LAN interface terminal
USB interface terminal
USB memory terminal
Power jack for the humidity sensor
Operation status LED
・POWER
・START
・CHARGE
Model imprint and others
GND terminal
AC adapter jack External input/output terminals
・LOGIC/PULSE
・EXT TRIG
・ALARM
Power switch

GL800 Connection Procedures
Connecting the AC Adapter Connecting the Grounding Cable
Making Connections to the Analog Input Terminals
Making Connections to the External Input/Output Terminals(Using B-513)
3
Connect the output side of the AC adapter to the
connector indicated as “DC LINE” on the GL800.
Use a flathead screwdriver to push the button
above the ground terminal while connecting the
grounding cable to the GL800. Connect the other
end of the cable to ground.
CH1 CH20
+
-
b
+
+
-
+
-
+
-
-
b
Direct voltage
Direct voltage input
Thermocouple input
Direct current input
Resistance temperature detector input
A
B
b
Compensation copper wire
Direct current
Orange with red dotted line : 1
Orange with black dotted line
: 2
Grey with red dotted line : 3
Grey with black dotted line : 4
White with red dotted line : 1
White with black dotted line : 2
Yellow with red dotted line : 3
Yellow with black dotted line : 4
Pink with red dotted line : Trigger input
Pink with black dotted line
Shielded
GND
Logic/pulse input
Alarm output
*B-513 (sold separately) cable is required for external input/output.
(For logic/pulse input, alarm output, trigger input
<Connection diagram>
Grounding cable
B-513
Lead wire resistance should be 10 Ω or below per wire, and
equivalent among the three wires.
Shunt resistance
Ex: for current in the 4 to 20 mA range,
apply a resistance of 250 Ω (±0.1%) and
perform measurement in the 1 to 5 V range.

• Avoid electrical shock and short circuit accidents
・Do not apply voltage of AC33 Vrms or 60 VDC or above between the
analog input section and main unit (GND terminal), or to between each
analog channel.
・ Do not apply voltage of AC33 Vrms or 60 VDC or above to analog input
section (between + and - terminals).
・ Do not apply radio-frequency signals.
• Warming-up
The GL800 should be allowed to warm up with the power turned on for
approximately 30 minutes to achieve operation according to the specified
performance.
• Unused channels
Capacitors have been incorporated into the analog input circuit to increase the noise
elimination capability.
Therefore, if input terminal is open, or signals are not input to the terminal,
measured results may be influenced by signals from other channels. In such a case,
set input setting to “Off” or short circuit the + and – terminals.
If signals are input correctly, measured results are not influenced by other channels.
Precautions to Observe When Performing Measurement
10 channels or less
11 to 20 channels
21 to 50 channels
51 to 100 channels
101 to 200 channels
500 msec or above
1 sec or above
2 sec or above
5 sec or above
10 sec or above
• Noise countermeasures
If measured values fluctuate due to extraneous noise, conduct the following
countermeasures.
(Results may differ according to noise type.)
Ex 1: Connect the GL800’s GND to ground.
Ex 2: Connect GL800’s GND to measurement object's GND.
Ex 3: Operate GL800 with batteries (B-517).
Ex 4: In the AMP settings menu, set filter to any setting other than “OFF”.
Ex 5:
Set the sampling interval which enables GL800’s digital filter (see table below).
Number of Measuring
Channels* Sampling Interval which enables Digital Filter
*”Number of Measuring Channels” is the number of channels in which
input settings are NOT set to “OFF”.
4
50Ω
50Ω
500kΩ
500kΩ
0.05μF
0.05μF
+
-
Channel switching relay

CH GROUP SPAN/TRACE
POSITION
TIME/DIV
QUIT MENU
KEY LOCK
USB DRIVEALM CLR
FILE
NAVI
CURSOR DISPLAY
REVIEW
START
STOP
ENTER
GL800 Descriptions of the Control Panel Keys
1. SPAN/TRACE/POSITION
2 TIME/DIV
Press the TIME/DIV key to change the time axis display range on the waveform screen.
This key enables SPAN, POSITION, and TRACE settings to be made independently for each
channel. Each time this key is pressed, the mode displayed in the waveform operation display
area changes in the sequence shown below. Use the and keys to select the channel, and
the and keys to change the setting values.
Press this key to switch to the next group consisting of 10 channels.
Press the side to switch to the group consisting of the next 10 channels with a smaller number.
Press the side to switch to the group consisting of the next 10 channels with a larger number.
MONITOR
SPAN
POSITION
TRACE
2
1. CH GROUP
1
3
(1)CH GROUP (2)SPAN/TRACE/POSITION
(3)TIME/DIV
(4)MENU
(5)QUIT
(6)DIRECTION KEYS
(7)ENTER
(8)FAST FORWARD KEY (KEY LOCK)
(9)START/STOP(USB DRIVE)
(10)REVIEW
(11)DISPLAY
(12)CURSOR (ALARM CLEAR)
(13)FILE
(14)NAVI
1 to 10 191 to 20011 to 20
5
Displays digital values (default).
Used to make span settings (change the waveform
amplitude).
Used to make position settings (adjust the upper and
lower values of the waveform).
Used to make trace settings (set the waveform display to
On or Off).
Note: If the QUIT key is pressed when the GL800 is in the SPAN, POSITION, or
TRACE mode, the display returns to MONITOR mode.

3. MENU
Press the MENU key to open a setup menu. Each time this key is pressed, the setup screen
tabs change in the sequence shown below.
AMP
DATA
TRG
USER
I/F
OTHR
4
4 QUIT (LOCAL)
5
6 ENTER
7
keys (DIRECTION KEYS)
5.
6
6.
keys (KEY LOCK)
8
6
• AMP Settings
Used to make the input, range, filter and other settings.
• Data Capture Settings
Used to make settings such as the sampling interval, data capture
destination, and calculations during data capture.
• Trigger Settings
Used to specify data capture start and stop conditions, and alarm conditions.
• User Settings
Used to set the names of the users of this device, and to change from one
user to another.
• Interface Settings
Used to set USB ID numbers and LAN IP addresses.
• Other Settings
Used to make settings such as screen brightness and AC frequency.
Press the QUIT key to cancel the settings and return them to their default status. If the device
is in the Remote (Key Lock) status, namely the external operating status via the interface, press this key
to return the device to the normal operating status (Local).
These keys are used to select menu setup items, to make span settings in the digital display area, to
move the cursors during a data replay operation, and so forth.
Press the ENTER key to enter the settings made in the setup menus, and to confirm your settings.
These keys are used to move the cursor at high speed during a data replay operation, and to
change the operation mode in the file settings box. Hold down both keys simultaneously for at
least two seconds to enable key lock status. To cancel key lock status, press them again for at
least two seconds. The key lock status can be confirmed by the status of the key lock lamp on
the monitor.

8. REVIEW key
9 DISPLAY key
1 CURSOR (ALM CLR) key
FILE key
NAVI key
Digital + Calc.
Waveform + Digital
Expanded Waveform
7. START/STOP (USB DRIVE) key
9
10
11
12
13
14
7
Press the START/STOP key to perform a data capture start operation while the GL800 is
in the Free Running status, and a data capture stop operation when data capture has
ended. If this key is held down while the power to the GL800 is turned on, the GL800
goes into USB Drive Mode.
Press the REVIEW key to perform a data capture start operation while the GL800 is in
the Free Running status, and a data capture stop operation when data capture has
ended. If this key is held down while the power to the GL800 is turned on, the GL800
goes into USB Drive Mode.
Note: A data replay operation will not be performed if data has not been captured.
Press the DISPLAY key
Press the CURSOR key to switch between the A and B cursors during a data replay
operation. If the Alarm setting has been specified as "Alarm Hold", press this key to clear
the alarm. The alarm settings are made in the "TRIG" menu.
Press the FILE key to save data to the GL800's internal memory and to a USB memory
device.
Press the NAVI key to display operational descriptions during the Free Running status,
and during data capture and data replay operations.
• Waveform + Digital
This is the default screen when the GL800 is turned on, and both
waveforms and digital values are displayed. The screen settings
can also be changed by using the SPAN/POSITION/TRACE key.
• Expanded Waveform
Displays waveforms only.
• Digital + Calc.
Displays large-size digital values and two types of
calculation processing results. The calculation settings are
made in the "DATA " menu.

GL800
Descriptions of the Menu Screens
Start side
Rising
trigger
Falling
trigger
Within
the range
Outside
the range
Alarm rangeTrigger position
Stop side
8
1.Simplified message display area
2.Time/DIV display area
3.Device access lamp
4.Key lock lamp
5.Remote lamp
6.Clock display
7.AC/Battery status indicator
8.Waveform operation display area
9.Digital display area
10.Quick settings
11.Alarm display area
12.Pen display13.File name display area
14.Scale lower limit
15.Waveform display area
16.Scale upper limit
17.Data capture bar
1.Simplified message display area :Displays the operating status.
2.Time/DIV display area :Displays the current time scale.
3.Device access lamp :Turns red when USB memory is accessed.
When the GL800’s internal memory is being accessed, the MEM lamp turns red.
4.Key lock lamp :Displays the key lock status. (Yellow = keys locked, white = not locked)
5.Remote lamp: :Displays the remote status. (Yellow = Remote status, white = Local status)
6.Clock display :Displays the current date and time.
7.AC/Battery status indicator :Displays the following icons to indicate the operating status of the AC power supply and
the battery.
8.Waveform operation display area :Displays the mode selected by the SPAN/POSITION/TRACE key.
9.Digital display area :Displays the input values for each channel. The and keys can be used to select the
active channel (enlarged display). Moreover, the selected active channel is displayed at the
very top of the waveform display.
10.Quick settings :Displays items that can be easily set. The and keys can be used to make a Quick
settings item active, and the and keys to change the values.
11.Alarm display area :Displays the status of the alarm output terminal. (Red = alarm generated, white = alarm
not generated)
12.Pen display :Displays the signal positions, trigger positions, and alarm ranges for each channel.
13.File name display area: :Displays the data capture file name during the data capture operation. During a data
replay operation, the name of the data replay file is displayed.
14.Scale lower limit: :Displays the lower limit of the scale of the currently active channel.
15.Waveform display area: :The input signal waveforms are displayed here.
16.Scale upper limit: :Displays the upper limit of the scale of the currently active channel.
17.Data capture bar :During a data capture operation, this bar displays the remaining memory capacity of
the device used for data capture. When data is being replayed, the display position
information is displayed here.
AC/Battery Indicator
Battery power: Full
When the AC power supply is being used
Battery power: Medium
Battery power: Low Battery power: Very low