GSi Protek A333 Bedienungsanleitung


Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
2
CONTENTS
INTRODUCTION............................................................................................................... 4
1. SAFETY INSTRUCTIONS .......................................................................................... 5
2. GENERAL OVERVIEW .............................................................................................. 7
2.1 Description ............................................................................................................ 7
2.2 Specifications ........................................................................................................ 7
2.3 Options and Accessories...................................................................................... 15
2.4 Principle of Operation ......................................................................................... 15
3. PREPARATION FOR USE ........................................................................................ 17
3.1 General Information ............................................................................................ 17
3.2 Front Panel .......................................................................................................... 17
3.3 Rear Panel ........................................................................................................... 26
4. GETTING STARTED ................................................................................................. 29
4.1 Analyzer Preparation for Reflection Measurement............................................ 30
4.2 Stimulus Signal Setting ...................................................................................... 31
4.3 IF Bandwidth Setting ......................................................................................... 31
4.4 Number of Traces, Measured Parameter and Display Format Setting............... 32
4.5 Trace Scale Setting............................................................................................. 33
4.6 Analyzer Calibration for Reflection Coefficient Measurement ......................... 34
4.7 SWR and Reflection Coefficient Phase Analysis Using Markers...................... 36
5. MEASUREMENT SYSTEM SETTING .................................................................... 38
5.1 Screen Layout and Functions.............................................................................. 38
5.2 Channel Window Layout and Functions............................................................. 42
5.3 Channel Parameters Setting Using Mouse ......................................................... 50
5.4 Channel and Trace Setting.................................................................................. 56
5.5 Stimulus Setting ................................................................................................. 61
5.6 Trigger Setting.................................................................................................... 66
5.7 Measured Data Setting ....................................................................................... 67
5.8 Data Format Setting ........................................................................................... 69
5.9 Diagram Scale Setting........................................................................................ 75
5.10 IF Bandwidth, Averaging and Smoothing Setting.............................................. 79

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
3
5.11 Initial Condition Setting ..................................................................................... 81
5.12 Reference Frequency Oscillator Selection ......................................................... 81
5.13 User Interface Setting.......................................................................................... 82
6. CALIBRATION AND CALIBRATION KIT ............................................................. 86
6.1 Calibration.......................................................................................................... 86
6.2 Calibration Kit Management............................................................................ 112
7. MEASUREMENT DATA PROCESSING FUNCTIONS ........................................ 117
7.1 Markers............................................................................................................. 117
7.2 Fixture Simulation............................................................................................ 132
7.3 Time Domain Transformation .......................................................................... 137
7.4 S-Parameter Conversion................................................................................... 141
7.5 Limit Test ......................................................................................................... 142
7.6 Memory Trace Function ................................................................................... 145
8. ANALYZER DATA OUTPUT.................................................................................. 149
8.1 Analyzer State Saving and Recalling ............................................................... 149
8.2 Data File Saving ............................................................................................... 151
8.3 Diagram Printout .............................................................................................. 153
9. MAINTENANCE AND STORAGE......................................................................... 155
9.1 Maintenance Procedures ................................................................................. 155
9.2 Storage Instructions.......................................................................................... 156
10. WARRANTY INFORMATION................................................................................ 157
Appendix 1 — Default Settings Table ............................................................................ 159

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
4
INTRODUCTION
This Operating Manual represents design, specifications, overview of functions, and
detailed operation procedure of A333 Network Analyzer, to ensure effective and safe
use of the technical capabilities of the instruments by the user.
Network Analyzer operation and maintenance should be performed by qualified
engineers with initial experience in operating of microwave circuits and PC.
Below you will find the abbreviations used in this Manual:
PC - Personal Computer
DUT - Device Under Test
IF - Intermediate Frequency
CW - Continuous Wave
SWR - Standing Wave Ratio

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
5
1. SAFETY INSTRUCTIONS
Carefully read through the following safety instructions before putting the Analyzer
into operation. Observe all the precautions and warnings provided in this manual for
all the phases of operation, service, and repair of the Analyzer.
The Analyzer must be used only by skilled and specialized staff or thoroughly trained
personnel with the required skills and knowledge of safety precautions.
A333 complies with INSTALLATION CATEGORY II as well as POLLUTION
DEGREE 2 in IEC61010-1.
A333 is MEASUREMENT CATEGORY I (CAT I). Do not use for CAT II, III, or
IV.
A333 is for INDOOR USE only.
A333 is tested with stand-alone condition or with the combination with the
accessories supplied by Protek against the requirement of the standards described in
the Declaration of Conformity. If it is used as a system component, compliance of
related regulations and safety requirements are to be confirmed by the builder of the
system.
Never use the Analyzer without secure grounding. Operation is permitted only with
protective earth connection by the supplied power cable to a 3-pin grounded power
outlet.
Never operate the Analyzer in the environment containing inflammable gasses or
fumes.
Operators must not remove the cover or part of the housing. The Analyzer must not
be repaired by the operator. Component replacement or internal adjustment must be
performed by qualified maintenance personnel only.
Never operate the Analyzer if the power cable is damaged.
Never connect the test terminals to mains.
Observe all the general safety precautions related to operation of equipment powered
by mains.
The definitions of safety symbols used on the instrument or in manuals are listed
below.
Refer to the Manual if the instrument is marked with this
symbol. .
Alternating current.

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
6
Direct current.
On (Supply).
Off (Supply).
A chassis terminal; a connection to the instrument’s chassis,
which includes all exposed metal structure.
WARNING
This sign denotes a hazard. It calls attention to a procedure,
practice, or condition that, if not correctly performed or
adhered to, could result in injury or death to personnel.
CAUTION
This sign denotes a hazard. It calls attention to a procedure,
practice, or condition that, if not correctly performed or adhered
to, could result in damage to or destruction of part or all of the
instrument.
Note
This sign denotes important information. It calls attention to a
procedure, practice, or condition that is essential for the user to
understand.

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
7
2. GENERAL OVERVIEW
2.1. Description
A333 Network Analyzer has been designed for use in the process of development,
adjustment and testing of various electronic devices in industrial and laboratory
facilities, including operation as a component of an automated measurement system.
A333 is designed for operation with external PC, which is not supplied with the
Analyzer.
2.2. Specifications
2.2.1. Basic Specifications
The basic specifications of the Analyzer detailed in table 1.1 apply over the
temperature range of 23 °C ±5 °C (unless otherwise specified) after 40 minutes of
warming-up, with less than 1 °C deviation from the full two-port calibration
temperature, at output power of -5 dBm, and IF bandwidth 1 Hz. The calibration
requires the use of the calibration kits specified in table 6.3.
Table 1.1 Basic Specifications
Frequency range 300 kHz to 3.2 GHz
CW frequency accuracy ±5×10
–6
Harmonic distortion -30 dB
Non harmonic spurious -30 dB
Output power level -45 dBm to +10 dBm
Output power level accuracy ±1.0 dB

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
8
Table 1.1 (continued)
Magnitude transmission measurement accuracy, if S
11
andS
22
of the DUT are less than -32 dB, and S
21
and
S
12
values are as follows:
+5 dB to +15 dB
-50 dB to +5 dB
-70 dB to -50 dB
-90 dB to -70 dB
0.2 dB
0.1 dB
0.2 dB
1.0 dB
Phrase transmission measurement accuracy, if S
11
andS
22
of the DUT are less than -32 dB, and S
21
and
S
12
values are as follows:
+5 dB to +15 dB
-50 dB to +5 dB
-70 dB to -50 dB
-90 dB to -70 dB
2
o
1
o
2
o
6
o
Magnitude reflection measurement accuracy, if S
11
and
S
22
values are as follows:
-15 dB to 0 dB
-25 dB to -15 dB
-35 dB to -25 dB
0.4 dB
1.5 dB
4.0 dB
Phase reflection measurement accuracy, if S
11
and
S
22
values are as follows:
-15 dB to 0 dB
-25 dB to -15 dB
-35 dB to -25 dB
4
o
7
o
22
o
Receiver noise floor (IF bandwidth 10 Hz) -120 dBm
Trace noise (IF bandwidth 3 kHz) 0.001 dB rms
Uncorrected directivity 25 dB
Uncorrected source match -15 dB
Uncorrected load match -25 dB

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
9
Table 1.1 (continued)
AC mains power 220 ± 22 V
Power consumption 30 W
Dimensions LxWxH 275x415x97 mm
Weight 5 kg
Operating conditions:
- Temperature
- Humidity at 25 °C
- Altitude
- Error-Corrected Temperature Range
(with <1°C deviation from calibration temperature)
5 °C to 40 °C
90%
0 to 2,000m
23°C ± 5°C
Non- Operating conditions:
- Temperature
- Humidity at 25 °C
- Altitude
-10 °C to 60 °C
90%
0 to 4,572m

Version 1.0 ▪ June 2008
A333
Network Analyzer
© 2007, 2008 GS Instru ents, Co., Ltd.
10
2.2.2. Supplemental Specifications
Source stability within operating temperature range ±5×10
–6
.
Frequency resolution 1 Hz.
Measurement time per test point 125 µs.
Number of test points per sweep from 2 to 10001.
Source to receiver port switchover time less than 10 ms.
Power level resolution 0.05 dB.
IF bandwidth settings from 1 Hz to 30 kHz with step of 1/1.5/2/3/5/7.
Measurement of conversion coefficient |С21| for frequency converting
DUT. |С21| measurement accuracy for |С21| values from -60 dB to +10 dB
is 1.5 dB.
External reference frequency is 10 MHz, input level is 2 dBm ± 2 dB, and
input impedance at «10 MHz» input is 50 Ω. Connector type is BNC
female.
Output reference signal level is 3 dBm ± 2 dB at 50 Ωimpedance. «OUT 10
MHz» connector type is BNC female.
Connector for external trigger source is BNC female. TTL compatible
inputs of 3 V to 5 V magnitudes have up to 1 µs pulse width. Input
impedance at least 10 kΩ.
Effective directivity 45 dB.
Effective source match -40 dB.
Effective load match -45 dB.
Dependence ofS21orS11 parameter of 0 dB per 1 degree variation of
environment temperature is less than 0.02 dB.
Warm-up time 40 min.