Integra CALEC ST II Bedienungsanleitung

Installation and operation manual
CALEC®ST II
Table of contents
1 Safety 2
2 Structure of a measuring point 4
3 Scope of delivery, installation accessories 5
4 Installation 6
5 Controls and operation 32
6 Maintenance and repair 49
7 Disposal 49
8 Technical specifications 50
9 CE Declaration of Conformity 54
10 Appendix 55
VD 3-126e 01.2018

1 Safety
1.1 Intended use
CALEC®ST II is a high precision instrument designed for the collection, analysis, presentation and transmission
of information. Improper or non-intended use of the device may compromise operational safety. We accept no
liability for any resulting damages.
1.2 Notes on safety instructions and symbols
The device has been designed to fulfil modern safety requirements. It has been tested and delivered in a condition
that ensures safe operation. However, improper or non-intended use of the device may result in it becoming
dangerous. Please always pay attention to the safety instructions in this manual which are accompanied by the
following symbols:
2CALEC®ST II
WARNING!
WARNING indicates an action or measure which, if performed incorrectly, can cause
potentially life-threatening injuries and lead to a high safety risk. Always follow the
instructions and proceed with caution.
NOTE!
NOTE indicates an action or measure which, if performed incorrectly, may have an
indirect effect on the operation of the device, or trigger an unexpected response.
COMMENT!
COMMENT provides information and recommendations for efficient and trouble-free
operation.
CAUTION!
CAUTION indicates an action or measure which, if performed incorrectly, can cause
minor injuries and/or incorrect operation or destruction of the device. Always follow the
instructions.
REFERENCE!
REFERENCE refers to additional documents.

3
CALEC®ST II
1.3 Installation, startup and operation
General hazards and warnings
This device is intended for permanent installation with a fixed electrical connection. The installation, connection
to the electricity supply, startup and maintenance must be carried out by trained, qualified personnel who are
authorised to perform this type of work. The respective specialist personnel must have read and understood this
installation and operation manual, and follow the instructions contained therein. The operator must ensure that
the measuring system is wired correctly in accordance with the wiring diagrams. Contact protection is deactivated
when the cover is removed (risk of electric shock). The housing may only be opened by qualified personnel.
Disconnect the power supply prior to electrical installation, and ensure that it cannot be reconnected
without your consent.
Pay attention to the following points during installation:
• Voltage, operating data
• Maximum transmission length
• Cable cross-section and length
• Ambient temperature and installation position
1.4 Technical progress
We reserve the right to modify the technical specifications without prior notice.
WARNING!
Danger of electrocution!
Touching or gripping live electrical parts can cause an electric shock, which may
result in burns, paralysis or death.
• The device should only be opened, installed or repaired when the power has
been switched off.
• Only specialist technicians are authorised to work on or with mains voltage.
They must comply with the applicable regulations.
• Only the intended terminals should be used to connect the device to the mains
supply.
REFERENCE!
Latest version of this manual!
The latest version of this manual is available at:
http://www.aquametro.ch/qr/prod/calec-st/11111.html

4CALEC®ST II
2 Structure of a measuring point
A complete measuring point for thermal energy consists of the CALEC®ST II energy calculator, paired temperature
sensors and flow sensor.
The applicable regulations and recommendations for installation and operation must also be complied with,
e.g. Section 6 of EN 1434, and the recommendations of professional associations such as the „Merkblätter der
Fernwärmeversorgung“ [fact sheets for district heating] issued by the German Energy Efficiency Association for
District Heating, Cooling and CHP (AGFW). The installation of control immersion sleeves is mandatory in some
countries (e.g. in France after FDE 39-007).
NOTE!
Type of temperature sensor, pulse value, installation side!
Check whether the temperature sensor type (e.g. Pt100), pulse value and installation
side (cold, hot) of the flow sensor match the nameplate on the CALEC®ST II.
NOTE!
Calibrated device!
Depending on the version, the CALEC®ST II is a calibrated measuring device
(see calibration markings on the nameplate).
Calibration is rendered invalid if any changes are made to the calibration parameters.
The parameters can only be accessed once the verification seal has been destroyed.
Parameters related to the calibration of calibrated devices can only be adjusted by
the manufacturer or by a designated calibration centre. The commissioning of a calibrated
measuring point must be made by an authorized organisation according to the valid regu-
lations.
After commissioning of the meter, the components (the calculator, the flow meter and two
temperature sensors) must be sealed.

5
CALEC®ST II
3 Scope of delivery,
installation accessories
The supplied assembly and connection accessories
are located on the top of the housing:
The yellow piercing awl 1) can be used to pass the
cables through the membrane seals easily
The strain on the wire can be eased using the
white strain-relieving discs 2).
The two red plugs 3) are used to secure the two screws in the front cover.
Unauthorised opening of the device can be detected if these safety caps have been installed.
Removing the protective caps:
Pierce with a sharp tool and lever them out. This damages the cap, which needs to be replaced.
NOTE!
Calibrated device!
The unit can be damaged if not stored correctly!
This precision measuring device can be damaged by heat, moisture, dirt and vibration,
which can cause malfunctions.
The device must be stored in accordance with the specifications and only removed from
the packaging immediately prior to installation.

6CALEC®ST II
4 Installation
The CALEC®ST II can be mounted on a mounting rail or on a flat wall. Suitable mounting rails are available as
an accessory (article number 19838). Please refer to the last page of this document for a hole template for both
installation types.
Device dimensions
Hole dimensions for wall mounting Clear height
(for clipping on the housing cover)

7
CALEC®ST II
Tools, assembly materials
(Not included in the scope of delivery)
Opening the housing
3 x Ø
43 x Ø6
Torx T15
Ø6
3 x 4.3 x 12
3.5 x 0.6

8CALEC®ST II
Unit design
ÀThe front cover can be locked into the lower section of the housing
ÁThe connection diagram is located on the inside of the front cover
ÂThe motherboard can be accessed by removing the plug-in totaliser
ÃMotherboard with terminals and microswitches for configuring inputs and outputs
ÄClip-on holder for mounting rail
À
Á
Â
Ã
Ä

9
CALEC®ST II
4.1 Installation
Installation instructions
Select the installation point:
• protected from moisture, heat, direct sunlight and damage
• easily accessible for reading, operation and installation
• at a sufficient distance from sources of electromagnetic interference
Mounting on rail DIN-EN 50222
ÀDrill holes Ø6
ÁScrew in the mounting rail
ÂClip the device onto the
mounting rail
Wall mounting
Only mount the device on
a flat surface!
ÀDrill holes Ø6
ÁScrew in the device

10 CALEC®ST II
4.2 Electrical connection
Hazard warning
Connection diagram
The connection diagram is located on the inside of the front cover.
Sample connection diagram:
Analogue outputs in socket #1 and M-Bus
in socket #2.
WARNING!
Danger of electrocution!
Touching or gripping live electrical parts can cause an electric shock, which may
result in burns, paralysis or death.
• The device should only be opened, installed or repaired when the power has been
switched off.
• Only specialist technicians are authorised to work on or with mains voltage.
They must comply with the applicable regulations.
• Only the screw-type terminals in the separated area on the left in the terminal
compartment should be used to connect the device to the mains supply.