JingYan TLL90S Bedienungsanleitung

2014 Dec 10
0
TLL90S Digital Protractor
User Guide

2014 Dec 10
1
Features:
1) TLL90S: High accuracy ±0.005°, high resolution 0.001°
2) Dual and Single axis with user friendly LCD display angle
3) Laser module installed for alignment
4) Rechargeable
5) V Shape Aluminum metal case for easy to fit at the corner or pipe.
6) 2 Side Magnetic base with milled surface
7) Audible alarming at settable angle range
8) Any angle measurement with gyro
9) USB / Bluetooth to pc connection need adapter
10) PLC Modbus RS485 connection with 6P6C socket, connect up to 255 unit in
one network
Specifications
Gravity Measurement Accuracy: 0 to 20°: ± (0.005°)
20 to 70°: ± (0.01°)
70 to 90°: ± (0.005°)
*After Calibrated
Measuring range: Single axis: 360°, Dual axis: ±40°
Resolution: 0.001°
Gyro Measurement 0.5°
Accuracy:
Gyro Measurement 0.1°
Accuracy:
Gyro Rotation Speed: <50°/s

2014 Dec 10
2
Display Response time: <0.6 second
Audio sound: 60dB @ 30cm
Zero offset drift angle per °C: 0.002° (typical)
Operating temperature: 0 to 50°C
Storage temperature: -10 to 60°C
User Interface: Mono-color LCD with backlight
Supply Power: Rechargeable Li-Polymer 3.7V
Charger port: 5V 500mA Mini type-B USB port
Power Consumption: Standby: 10uA, Operation: 20mA.
Standby Battery Life: 4000 hours
Operating Battery Life: 30 hours
Dimensions (in mm): 93(L) x 66(W) x 28(H)
Magnetic Base: affix at 2 face
Magnetic Force: N35
Weight: 120gram

2014 Dec 10
3
Button Functions
Function Descriptions
Button Normal Mode MENU Mode
This button turns the device ON.
Press and release will turn laser on/ off.
Long Press for 3 seconds will turn off.
Serves as the
escape key at
menu mode
When pressed, the current reading is set to zero;
subsequent measurements are relative to this
reading. The LCD will show the icon to indicate
the device is in zero mode.
Press and hold for 3 seconds to enable or disable
sound. The icon on LCD will be displaced
accordingly. The buzzer alarming could be set at
different accuracy level. Refer to section “Angle
Alarming”.
Serves as the
upward key
for option
selection
When pressed, the current value will freeze; the unit
icon flashes to indicate the reading is on hold.
Press and hold for 3 seconds to switch the
measurement mode Degree -> mm/M -> Gradient
%
Serves as the
downward
key at menu
mode

2014 Dec 10
4
Press and hold for 3 seconds to enter MENU mode,
for set mode options.
Serves as the
Set key.
Any angle measurement start button. Refer to
section “Any Angle”
Press and hold for 3 seconds to enter MENU mode,
for set mode options.
Serves as the
Set key.
LCD Icons Representations
Battery status indication icons
These icons indicate the battery level. There are 3 levels
representing empty, half and full.
In dual-axis mode, the LCD displays the direction of tilt graphically.
It will show as E bubble to display the direction of tilt
Degree mode. Flash when unit is in HOLD mode
mm/M, the height of one end for 1m long plate.
Gradient % mode. Flash when unit is in HOLD mode
Sound notification on.
Blank as off

2014 Dec 10
5
Show this logo as relative value is showing. When the Zero button
is pressed, the unit reset current angle to zero.
Direction of tilt icons, show the tilt angle direction
Dual-Axis Mode. Both the X and Y axis angle will be showed. Dual
axis mode measures inclination up to +-40 degree for each axis
before it automatically switch to single axis mode.
Single Axis Mode. Measure slope up to +-90 degree.
The triangle icon indicates the direction of tilt with respect to the
bottom right corner of the unit.
Battery Charging
It has a built in Lithium Ion rechargeable battery. A standard charger is supplied that
the input voltage is 110V to 240V AC, 50/60Hz, and the output is 5V DC, 500mA.
The charger operating procedure is list below:
1) Plug the Charger into AC socket, the RED indicator on the charger should
turned ON,
2) Plug the USB charging cable to the Charger,
3) Insert the other end of the USB cable to the unit,
4) The battery icon on the LCD blinks to indicate charging in process. Upon
charging complete, the icon stops blinking.
5) The charging time is approx. 3 hours.
The unit could also be charged by connecting the USB cable to the unit and a

2014 Dec 10
6
computer’s USB port. This has the same effect when charging the unit with the
provided adaptor.
Note: When the unit is turned OFF, and plug in the USB charge cable, the LCD will
no show anything, it is NORMAL. Once the unit is turned ON, the battery icon should
flash indicating the unit is in charging mode.
Relative/ Absolute Measurement
Relative Measurement
LCD Icon:
Absolute Measurement
LCD Icon: Blank
Relative and Absolute Mode Switching:
1 Press and release the “ZERO” key to set the relative measurement zero point
2 Press and release the “ZERO” key to cancel the relative zero point and back
to absolute measurement mode.
Hold Function
Holding Mode:
LCD Icon blinking:
Holding function:
1. Press and release the “HOLD” key to activate holding function, digit will freeze.
2. Press and release the “HOLD” key to cancel the holding function.

2014 Dec 10
7
Auto Power Off
For no movement in 30 minutes, the unit will power off.
Or we can set to never sleep mode at below instruction.
Power auto off setting:
1. Press and hold “SET” / “ANGLE” key and enter MENU mode
2. Select “POWER” by “ZERO” and “HOLD” key, press “SET” key to enter Power
mode
3. Scroll “NEVER” or “30MIN” by “ON/OFF” and “HOLD” key
4. Press “SET” key to confirm NEVER or 30MIN (30 minutes) sleep
Restore Factory Setting
When you find that the unit is abnormal, you can restore the unit to factory setting.
All tilt mode calibration setting will be restored to factory setting.
*Not suggest restore to factory setting in normal status, for accuracy drift, please
follow Calibration. After factory set, please redo calibration to ensure the accuracy.
Restore to factory setting:
1. Press and hold “SET” / “ANGLE” key and enter MENU mode
2. Select “FACTORY SET” by “ZERO” and “HOLD” key, press “SET” key to enter
FACTORY SET mode
3. Scroll “YES” or “NO” by “ZERO” and “HOLD” key
4. Press “SET” key to confirm
Angle Alarming
Alarming Mode:
LCD Icon:
BUZZER
>SINGLE : 0 0 . 0 0 0
DUAL. X: 0 0 . 0 0 0
DUAL. Y: 0 0 . 0 0 0
RANGE : 0 0 . 5 0 0
ACTIVE : IN OUT

2014 Dec 10
8
Alarming Angle setting:
1. Press and hold “SET” / “ANGLE” key and enter MENU mode
2. Select “BUZZER” by “ZERO” and “HOLD” key, press “SET” or “ANGLE” key to
enter BUZZER setting
Press and hold “ZERO” and “HOLD” key for fast scrolling the digit.
3. Press “SET” or “ANGLE” to enter setting value
SINGLE Vertical / Single axis mode alarming angle (Degree)
DUAL.X Horizontal/ Dual axis mode X axis alarming angle (Degree)
DUAL.Y Horizontal/ Dual axis mode Y axis alarming angle (Degree)
RANGE The range (Degree)that will trigger the audible alarming
For example: SINGLE. set to 20.000
RANGE set to 01.000
While the unit is in +19° to +21°or -19 ° to -21°, the unit will alarm
ACTIVE IN or OUT to set the alarm will alarm in of range or out of range
Calibration
Calibrate the unit, once you found that there is accuracy drift on the unit. You can
verify the accuracy at below step:

2014 Dec 10
9
At Step1, you measured X and Y value, X1 and Y1
At Step2, you measured X2 and Y2, in theory X1=-X2, Y1=-Y2.
If the error is too large, you can enter calibration mode to eliminate the error
Accuracy drift is causing by large ambient temperature change (5 to 10 Degree
Celsius) or the unit has been dropped.
Calibration Procedure:
Step 1: Press and hold the “Angle/Set” key enter Menu mode. Select “Calibration”
mode, press “SET”. Place the unit on a flat table (no need perfect level table)
LCD display “CALIBRATE DUAL AXIS PRESS SET”, press “SET” and buzzer will
beep; wait until the beep sound stop. While the buzzer is beeping keep the unit
stable.
Step 2: Then rotate the unit 180 degree with the other side against the same place.
Press the Set button again, and wait for the beep finished.
Andere Handbücher für TLL90S
1
Inhaltsverzeichnis
Andere JingYan Messgerät Handbücher