Mecon RE 250 Bedienungsanleitung

© MECON GmbH BA RE 250 EN 18-01
Operating Instruction
RE 250 Variable Area Flowmeter

Pa e 2 / 32 www.mecon.de BA RE 250 EN 18-01
Imprint
All ri hts reserved. It is prohibited to reproduce this document, or any parts thereof, without prior written
authorization of MECON GmbH.
Subject to chan e without notice.
Copyri ht 2018 by
MECON GmbH - Rönt enstraße 105 - 50169 Kerpen - Germany

BA RE 250 EN 18-01 www.mecon.de Pa e 3 / 32
Contents
Contents
1 Safety instructions _________________________________________________________ 4
1.1 Intended use _____________________________________________________________ 4
1.2 Certifications ______________________________________________________________ 5
1.3 Safety instructions from the manufacturer _______________________________________ 5
2 Start-up _________________________________________________________________ 6
3.1 Standard device ___________________________________________________________ 6
3.2 Device with electrical output (MEM) ____________________________________________ 7
3.3 Device with limit switches ____________________________________________________ 7
3 Installation ode of operation _______________________________________________ 8
2.1 Notes on installation ________________________________________________________ 8
2.2 Installation _______________________________________________________________ 8
2.3 Operatin principle ________________________________________________________ 10
2.4 Classification accordin the Pressure Equipment Directive 2014/68/EU ________________ 10
4 Service ________________________________________________________________ 11
4.1 Stora e _________________________________________________________________ 11
4.2 Maintenance and cleanin __________________________________________________ 11
4.3 Replacin float ___________________________________________________________ 12
4.4 Replacin float dampin ____________________________________________________ 12
4.5 Returnin the device to the manufacturer ______________________________________ 13
5 Device description _______________________________________________________ 14
5.1 Scope of delivery _________________________________________________________ 14
5.2 Device versions ___________________________________________________________ 14
5.3 Name plate ______________________________________________________________ 16
6 Description code _________________________________________________________ 17
7 Technical Data ___________________________________________________________ 21
7.1 Pressure limits vs. temperature stainless steel ___________________________________ 23
7.2 Dimensions and wei hts ____________________________________________________ 24
8 Electrical connections _____________________________________________________ 28

Pa e 4 / 32 www.mecon.de BA RE 250 EN 18-01
Safety Instructions
1 Safety Instructions
1.1 Intended use
The series RE 250 of flowmeters with a standard len th of 250 mm (9.84 inch) and a completely
metal desi n is suitable for measurin different liquids and ases in closed pipin .
The robust desi n enables the operation in rou h conditions. Different types of flan es, liners and
float materials meet the requirements of the pharmaceutical and chemical industries.
The devices are particularly suitable for the easure ent of:
• Water
• Fluids
• Corrosion protection a ents
• Lubricants
• Saturated and superheated steam
• Food and bevera es
• Industrial ases
Warning!
The operator of these measuring devices is solely responsible for the
suitability, intended use and corrosion resistance of the selected mate
rials. It must be particularly ensured that the materials selected for the
wetted parts of the flowmeter are suitable for the process media to be
measured.
The manufacturer is not liable for any damage resulting from improper or
unintended use of these devices.
No external loads may act upon the meter. The flowmeters are primarily desi ned for static
applications.
Caution!
Hot surfaces resulting from hot process media.
Danger of burns resulting from surface temperatures above 70 °C.
● Take appropriate protective measures, for example contact protection.
● The design of the contact protection must meet the maximum permissible
ambient temperature of the meter.
Caution!
For devices used in hazardous areas, additional safety notes
apply; please refer to the Ex documentation!
The flowmeter may only be operated within the pressure and volta e limits specified on the name
plate.
Before takin the flowmeter out of operations, check that the unit is free of hazardous media and
de-pressurized.

BA RE 250 EN 18-01 www.mecon.de Pa e 5 / 32
Safety Instructions
1.2 Certifications
CE arking
The anufacturer certifies for the device RE 250 eets all statutory require ents of the
following EU directives by applying the CE arking:
• Pressure equipment directive 2014/68/EU
• Low volta e directive 2014/35/EU *
• EMC-directive 2014/30/EU
• NAMUR recommandation NE21 *
• ATEX directive 2014/34/EU **
(*Devices with electrical installations)
(**Devices for use in hazardous areas)
1.3 Safety instructions fro the anufacturer
Disclai er
The manufacturer will not be liable for any dama e resultin from the use of its product, includin ,
but not limited to direct, indirect, incidental, punitive and consequential dama es.
Any product purchased from the manufacturer is warranted in accordance with the relevant product
documentation and our Terms and Conditions of Sale.
The manufacturer reserves the ri ht to revise the content of its documents, includin this disclaimer,
without prior notification, and will not be liable in any way for possible consequences of such chan es.
Product liability and warranty
MECON GmbH assumes no uarantee for repair work carried out by the customer without prior
notification and consultation. Any instruments or parts rejected by a customer must be returned to
us, if no other arran ement has been made.
General infor ation
This manual is intended for the correct installation as well as operation and maintenance of the
devices. Read the instructions carefully before installin the device and placin it into operation.
Specially desi ned or customized models and specialized applications are not included in this
manual.

Pa e 6 / 32 www.mecon.de BA RE 250 EN 18-01
Start up
2 Start-up
2.1 Standard device
When startin up the device, the followin points must be observed:
● Ensure that the actual operation conditions (pressure, temperature) do not exceed the limits
which are iven on the nameplate of the device
● Avoid float impacts!
Therefore it is recommended to start up with a closed shut-off valve to be opened slowly. The use
of solenoid valves is not recommended.
● When measurin liquids, the pipes must be vented slowly to prevent shock pressure due to as
bubbles.
● When measurin ases, the pressure must be increased slowly in order to prevent hi h shock
pressure.
● Durin start-up of new systems, residual materials may be carried in the medium and could
adhere to the float. In such cases it is recommended to clean the instrument after a relatively
short period of operation.
● When operatin the flowmeter at low flowrates, the device has to start-up with a hi h flowrate to
allow the float to level out. This will ensure measurements within the specified accuracy class.
Special issues for easuring gas flow:
● Valves have to be installed downstream of the device if p
abs
>1.013 bar and upstream of the
device if p
abs
=1.013 bar (free outflow).
● Install a throttle just downstream of the device to prevent compression oscillations durin
measurement.
● Ensure that the operatin pressure for the flowmeter corresponds to the calibration reference
pressure to avoid erroneous measurements.
Float da ping
A float dampin is recommended
● in eneral for measurin ases
● where pipin vibrations can not be avoided
It is also stron ly recommended in the case of:
● applications where air bubbles in the medium cannot be avoided
● possibility of shock pressure in the pipin , e. . due to rapid throttlin or process shut-off
● turbulence, pulsations, etc. which could lead to vibrations of the flowmeter
● rapid pressure build-up in the pipin

BA RE 250 EN 18-01 www.mecon.de Pa e 7 / 32
Start up
2.2 Device with current output (MEM)
The ma neto-electrical measurin transducer (MEM) is factory-set when it is delivered to the
customer.
After applyin the supply volta e to the device, initially the current output will be about 3.5 mA
to 4 mA for a few seconds. After that a current correspondin to the pointer deflection will flow.
Important!
Due to the influence of the float magnet, the MEM transmitter will only
output the correct current if the pointer position is caused by the float.
Turning the pointer manually will cause incorrect values, but it is suitable
for testing the unit.
2.3 Device with li it switches
The limit switches can be set over the entire measurin ran e by varyin the position of the limit
pointers.
Important!
Unless otherwise specified by the customer, the factory set value for the
switching point on versions with one limit switch is about 0% of URV
and for devices with two limit switches it is about 20% / 80% of URV.
To set the limit switches at the desired position the followin steps have to be observed:
● Unloosen the two small fixin screws (2) before movin the pointer (1).
● Move the pointer to the desired position.
● Ti hten the screws to fix the pointer.
Fig. 2 Adjustment of the switchin point (limit switch)
1 Limit pointer
2 Fixin screw

Pa e 8 / 32 www.mecon.de BA RE 250 EN 18-01
Installation
3 Installation and ode of operation
3.1 Notes on installation
Information !
All instruments are carefully checked for functional capability before
shipment. Upon receipt of the device please carry out a visual inspection
of the outer packing for damage or improper handling.
Please contact the delivery carrier if you discover any defects. In such
cases a description of the defect, the type and the serial number of the
device are needed.
Information !
Unpack the equipment with care to prevent damage.
Information !
All instruments are carefully checked for order conformity.
Please check the delivery for completeness using the packing list. Please
examine the flowmeter nameplate to verify that the device was built
according to your order. Particularly check devices with electrical
components for the correct supply voltage.
3.2 Installation
Below the essential points concerning installation are described and have to be observed:
● Remove the transport securin device from the fittin .
● Prior to the installation verify that the float slides smoothly in the fittin without cantin or
deadlockin . The pointer must follow the float movement smoothly.
● Verify that all accessories such as sprin stop, as/fluid dampin are still correctly fixed in the
fittin .
● In the zero position (no flow) the pointer must be at the marked reference point (first line on
the scale). When movin the float manually to the end position, the pointer must be above the
final value of the scale.
● The flowmeter must be installed vertically – flow direction from bottom to top. For installation
recommendations refer to uideline VDI/VDE 3513 Part 3.
● The installation in the pipin must be tension-free – therefore the pipin must be positioned
parallel and ali ned with each other.
● Avoid a corrosive atmosphere –provide ventilation where necessary.
● Ensure a sufficient distance of at least 200 mm to ma net-influencin parts such as solenoid
valves or ferroma netic components e. . steel brackets.
● Observe a lateral distance of at least 300 mm between two devices mounted next to one
another. The distance can be reduced by mountin the devices sta ered by the len th of one
device. To eliminate doubt the influence can be tested by movin the instrument up and down

BA RE 250 EN 18-01 www.mecon.de Pa e 9 / 32
Installation
ca. 200 mm at a selected distance while observin whether the position of the pointer on the
display chan es.
● Specify the installation zone re ardin a reliable readin of the displayed values and adequate
spacin for servicin .
● An inlet run upstream and an outlet run downstream the device is not necessary in case of a
linear media flow profile. In cases of hi hly asymmetrical flow profiles, however, additional
measures (e. . inlet tracks, flow rectifiers) with a len th of at least 250 mm could be
appropriate to ensure the measurin accuracy.
● Avoid the installation of unilaterally constrictin fittin s upstream of the device.
● Ensure that the pipin is installed securely to avoid vibration or oscillation of the devices.
● The nominal width of the flowmeter and the connected pipin must be identically.
● Use connectors connectin bolts suitable for the particular device version (dimension, pressure
ratin ).
● Ensure that the clearance between the flan es of the pipin corresponds to the device
dimension in addition to two askets.
● Before connectin the device, blow or flush out the pipes leadin to the device.
● The surface rou hness of the flan e sealin areas must be appropriate for the used askets.
● The askets and bolts of the prescribed dimensions must be appropriate for the operatin
pressure, the temperature and the media.
● Do not use carbon steel ounting brackets for the device!
Additionally for devices with PTFE liner:
● Use askets which conform to the inner and outer dimensions of the raised face of the
flowmeter.
● Ti hten the screws crosswise to ensure the ti htness of the process connections. For ti htenin
torques please refer to Tab. “2.2 Ti htenin torques”.
● If the instrument is absolutely calibrated for a pressure exceedin 1.013 bar any valve is
usually installed at the rear of the flowmeter, at 1.013 bar absolute (free emanation) in front
of the instrument.
Special issues for easuring gas flow:
● Valves have to be installed downstream of the device if p
abs
>1.013 bar and upstream of the
device if p
abs
=1.013 bar (free outflow).
● Install a throttle just downstream of the device to prevent compression oscillations durin
measurement.
● Ensure that the operatin pressure for the flowmeter corresponds to the calibration reference
pressure to avoid erroneous measurements.
Tightening torques
The flan e bolts for devices with PTFE liner may be ti htened with the followin torques (acc.
uideline VDI/VDE 3513):

Pa e 10 / 32 www.mecon.de BA RE 250 EN 18-01
Installation
Process connection ax. torque
EN1092-1 ANSI B16.5 N ftlbf
15 ½” 14 10
25 1” 14 10
40 1 ½” 25 18
50 2” 25 18
80 3” 35 25
100 4” 42 30
Tab 2.2 Ti htenin torques for devices with PTFE liner
Magnetic filters
To prevent a malfunction of the flowmeter caused by ferroma netic substances contained in the
medium (such as weld beads) a ma netic filter should be mounted in flow direction upstream of
the instrument.
This kind of filter is also recommended if such particles cannot be avoided in standard operation.
3.3 Operating principle
Like other devices of this series the variable area flowmeter RE 250 operates accordin to the
principle of flotation:
The measurin unit consists of a metal tube with a measurin rin in which a float can move up
and down. The media is flowin upwards throu h the tube and raises the conical float. Here the
annular ap enlar es until the equilibrium between the buoyancy force F
A
, the dra force F
S
and
the wei ht of the float F
G
is reached. The adjusted hei ht of the float is directly proportional to
the flowrate. The movement of the float is transmitted via a ma net inside the float to a
correspondin ma net inside the display unit outside of the measurin pipe.
Fig. 1 Operatin principle of the RE 250