Metal Samples MS2510 Bedienungsanleitung

MS2510 Current Loop Receiver
User Manual
Metal Samples Company
A Division of Alabama Specialty Products, Inc.
152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358-4202 Fax: (256) 358-4515
E-mail: [email protected] Internet: www.metalsamples.com

.

_______________________________________________________________________________________________
2
Table of Contents
SECTION 1 – INTRODUCTION
Receiving the Instrument.................................................................................................................................................3
Unpacking .......................................................................................................................................................................3
Cleaning ..........................................................................................................................................................................3
Installation.......................................................................................................................................................................3
Mounting.........................................................................................................................................................................4
Electrical .........................................................................................................................................................................5
Main Power.....................................................................................................................................................................6
Transmitter Connections .................................................................................................................................................6
Ethernet Connections ......................................................................................................................................................6
Adjustments.....................................................................................................................................................................6
Storage and Installation Environment .............................................................................................................................6
Physical Dimensions.......................................................................................................................................................7
Specifications..................................................................................................................................................................8
SECTION 2 – SAFETY
Potential Electrical Hazards ............................................................................................................................................9
SECTION 3 – START-UP AND OPERATION
MS2510 Receiver Start-Up.................................................................................................................................11
Operation..........................................................................................................................................................12
Controls Operation.............................................................................................................................................14
Title Screen ....................................................................................................................................................................14
Main Screen ...................................................................................................................................................................15
Setup Screen ...................................................................................................................................................................17
Data Chart Screen............................................................................................................................................................19
About Screen ..................................................................................................................................................................20
SECTION 4 – MAINTENANCE
Fuse Replacement ...........................................................................................................................................................21
SECTION 5 – TROUBLESHOOTING
Troubleshooting Procedures............................................................................................................................................22
SECTION 6 – RETURNING THE INSTRUMENT FOR REPAIR
Return Instructions .........................................................................................................................................................23
SECTION 7 – WARRANTY
Warranty .........................................................................................................................................................................25
SECTION 8 – WIRING DIAGRAM
Wiring Diagram ..............................................................................................................................................................26

.

_______________________________________________________________________________________________
3
Section 1 - Introduction
RECEIVING THE INSTRUMENT
The MS2510 Receiver is carefully prepared for shipment to protect it during transit and to enable the
user to install it with a minimum amount of reassembly.
Upon receipt, the instrument should be carefully inspected for transit damage. If damage is apparent,
notify Metal Samples and the carrier for claims inspection.
UNPACKING
Open the package and remove the instrument carefully. The instrument contains a display screen.
While removing, utmost care should be given to the display.
CLEANING
Clean the instrument thoroughly to remove any foreign material that may have accumulated during
shipment.
INSTALLATION
Place the receiver instrument in a convenient location, preferably away from heat producing sources or
direct sunlight.
Use the mounting brackets supplied with the instrument as shown in Figure 1.

_______________________________________________________________________________________________
4
MOUNTING
Figure 1 – Instrument Mounting

_______________________________________________________________________________________________
5
ELECTRICAL
Figure 2 – Connections
Transmitter Connections
1. Connect ‘+Ve’ terminal to TB1
2. Connect ‘–Ve’ terminal to TB2 Main Power Connection
1 Phase, 100 -240V AC, 50 /60 Hz, 2A
Current Loop Fuse
100mA 250V Ethernet
Port

_______________________________________________________________________________________________
6
MAIN POWER
Connect the single phase main voltage 100 V to 240 V by means of supplied main cord. Local
regulations must be considered before connecting it.
Note: In some cases, the instrument is shipped without a connector on the power cord. Each locality
should inspect the power cord and install the proper connector if necessary. The connector should be
installed in a properly grounded outlet for proper over current and earth fault protection.
TRANSMITTER CONNECTIONS
Connect the transmitter connections as below.
Positive Terminal - 1 (+)
Negative Terminal - 2 (-)
(See wiring diagram – Page 26.)
ETHERNET CONNECTIONS
If you are connecting the MS2510 to an Ethernet switch or Router, use a Straight RJ45 cable. If you
are connecting to a PC, use a Crossover RJ45 cable. Before connecting the MS2510 to your local
network please contact your network administrator for proper IP address configuration of the
instrument.
Warning – A duplicated IP address may cause network failure at the installation site. Consult
Network Administrator.
ADJUSTMENTS
The MS2510 Receiver functions must be adjusted and operated only by those authorized personnel
who have read and who thoroughly understand the descriptions of the various control functions
presented in this manual.
STORAGE AND INSTALLATION ENVIRONMENT
Protect the receiver instrument from mechanical stress, humidity, dust and thermal damage. Storage
temperature is -20º C to +60º C. Operating temperature is 0o C to 50oC.

_______________________________________________________________________________________________
7
PHYSICAL DIMENSIONS
Figure 3 – Physical Dimensions
Overall Dimensions: 9.85” H x 11.73” W x 6.50” D

_______________________________________________________________________________________________
8
SPECIFICATIONS
WEIGHT - ~4 lbs (~1.9 Kg)
DIMENSIONS - 9.85” x 11.73” x 6.50” (25.02cm x 29.79cm x 16.51cm)
MOUNTING - Panel Mount / Rack Mount
OPERATING TEMP. - 32F to 122o F (0o C to 50oC)
POWER SUPPLY
Voltage - 100 – 240 V, 1 phase, 50/60 Hz
Current - < 2 Amps
SIGNAL
Input - 4 – 20 mA current loop
Voltage - 24V DC
Input Impedance - 250 Ohms
Fuse - 100mA 250V
NETWORK
Use Crossover cable to connect the instrument directly to a PC.
Use Straight Cable to connect to the network switch, router…