Nippon Hi-Lutor Bedienungsanleitung

Option for High Concentration Mercury Measurement
Hi-Lutor
INSTRUCTION MANUAL
NIC-TD-0000115-02
NIPPON INSTRUMENTS CORPORATION

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation
CONTENTS
1. OUTLINE..........................................................................................................................1
SAFETY PRECAUTIONS ......................................................................................... 1
1.1.1 MEANINGS AND EXPLANATIONS OF SYMBOLS ........................................ 1
1.1.2 CAUTIONS CONCERNING INSTALLATION................................................... 1
1.1.3 CAUTIONS CONCERNING HANDLING ..........................................................1
1.1.4 CAUTIONS CONCERNING DISPOSAL ............................................................ 2
2. OUTLINE OF INSTRUMENT........................................................................................2
SIMPLIFIED EXPLANATION ..................................................................................2
HI-LUTOR UNIT CONTENTS.................................................................................. 2
EXTERNALAPPEARANCE.....................................................................................3
3. MEASUREMENT............................................................................................................. 4
ATTACHING PROCEDURE......................................................................................4
UNLOCK SOFTWARE..............................................................................................5
SETTING PROCEDURE ...........................................................................................6
MEASUREMENT PROCEDURE..............................................................................8
4. MAINTENANCE..............................................................................................................9
DAILY INSPECTION................................................................................................. 9
PERIOD OF REPLACEMENT .................................................................................. 9
MAINTENANCE WORK ........................................................................................ 10
4.3.1 REPLACEMENT OF HI-LUTOR....................................................................... 10
4.3.2 CHECKING OF FACTOR................................................................................... 10
5. DETAILED SYSTEM SPECIFICATION.................................................................... 11
SPECIFICATION OF HI-LUTOR............................................................................ 11
DIMENSIONS AND WEIGHT................................................................................ 11
6. CAUTIONS CONCERNING DISPOSAL OF INSTRUMENT.................................. 11
7. CONTACTADDRESS ...................................................................................................12

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 1
1. OUTLINE
SAFETY PRECAUTIONS
The Hi-Lutor is an option unit of the EMP-3.
Before using this unit, read the instruction manual of the EMP-3 carefully.
To use this product safely, observe the following precautions.
Nippon Instruments Corporation will not take any responsibility for accidents caused
by a user’s carelessness.
In the case of mishandling, a user may be killed or suffer a serious
injury immediately.
In the case of mishandling, a user may be killed or suffer a serious
injury.
In the case of mishandling, a user may suffer an injury or material
damage is expected.
Attach the unit to the EMP-3 main unit securely. If the unit is
not screwed, it may fall and be damaged while it is used.
If the high concentration of mercury is expected, only enter the
probe inside the contaminated area and the measurer should
measure from the non-contaminated area. It is to prevent the
contamination by mercury.
Do not measure the gas with the hydrocarbon concentration of
0.5% or higher.
If any time-change component is not replaced, original
performance may not be demonstrated.
When the pressure loss is large, the Factor changes. When the
probe filter is dirty, replace it soon.
Do not disassemble or modify the product without permission.
When it is necessary to replace a component, replace the whole
unit. It is impossible to replace a component alone.
Danger
Warning
Caution
Caution
Caution

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 2
The inside of the instrument has adsorbed mercury in the
environmental atmosphere. If the instrument is discarded
inadequately, environmental pollution or personal damage may
result.
Dispose of it according to the regulations or rules of the local
government.
2. OUTLINE OF INSTRUMENT
SIMPLIFIED EXPLANATION
The Hi-Lutor is an option unit for measuring high concentration mercury.
By connecting a Hi-Lutor, the measuring range is expanded to 0.01 to 20.0 mg/m3.
HI-LUTOR UNIT CONTENTS
Hi-Lutor 1unit
Connection Tube 2set (1 for maintenance)
Caution

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 3
EXTERNAL APPEARANCE
No.
Name
Description
1
Inlet
Connected to the probe
2
Outlet
Connected to the inlet of the EMP-3
3
Connection Tube
Tube to connect the Hi-Lutor and the inlet of the EMP-3
4
Fixing Metal
Component to connect the Hi-Lutor and EMP-3
5
Probe
Probe of the EMP-3
1
2
3
5
4

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 4
3. MEASUREMENT
ATTACHING PROCEDURE
Attach the unit to the EMP-3 main unit securely. If the unit is
not screwed, it may fall and be damaged while it is used.
(1) Remove the screws of the fixing metal of the belt on the right side of the EMP-3 main
unit.
(2) Fix the fixing metal of the Hi-
Lutor together using the
removed screws.
(3) Connect the probe to the inlet
of the Hi-Lutor, and connect
the outlet of the Hi-Lutor and
the inlet of the EMP-3 main
unit using the connection tube.
*Please confirm that the connection tube isn’t come out.
Caution
Remove
Fix together

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 5
UNLOCK SOFTWARE
When using for the first time, it is necessary to unlock the software.
When purchasing at the same time as EMP-3, it will be shipped after unlocking.
Once unlocking is completed, input after the second time is unnecessary.
(1) Start up the EMP-3.
(2) Select “MENU” - “Configuration”.
(3) Select “Hi-Lutor”.
(4) The password screen shown below appears. Press the number button, enter the key code
written on the inspection report using the displayed numeric keypad, and press “Enter”.
* If you enter a wrong key code, you can’t unlock it. Please check the password and
re-enter it.

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 6
SETTING PROCEDURE
(1) Start up the EMP-3.
(2) Select “MENU” - “Configuration”.
(3) Select “Hi-Lutor”.
(4) Select “Exit”. The screen shown below appears. Select the “Hi-Lutor Set”icon.

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 7
(5) The Factor screen shown below appears. Press the number button, enter the Factor
written on the Hi-Lutor using the displayed numeric keypad, and press “Enter”.
(6) If other items have been set, push “Exit”to return to the “START”screen.

Instruction Manual: Hi-Lutor NIC-TD-0000115-02
Nippon Instruments Corporation 8
MEASUREMENT PROCEDURE
If the high concentration of mercury is expected, only enter the
probe inside the contaminated area and the measurer should
measure from the non-contaminated area. It is to prevent the
contamination by mercury.
Do not measure the gas with the hydrocarbon concentration of
0.5% or higher.
Insert the probe from outside of the mercury
contaminated area as shown on the right to
measure mercury.
*Please use the long probe (optional) if necessary.
The measurement procedure is the same as that for the EMP-3.
When the “START”or “STOP”button on the main screen is pressed, a measurement starts
or stops, respectively.
*When the alarm is set to OFF (enter 0) and the measurement result exceed 20mg/m3, the
measurement result is displayed in gray.
* For details, see the instruction manual of the EMP-3.
Caution
Contaminated
area
Non-
contaminated area
Inhaltsverzeichnis
Andere Nippon Messgerät Handbücher

Nippon
Nippon MA-3000 RD-5+SC-5 Bedienungsanleitung

Nippon
Nippon WA-5A Bedienungsanleitung

Nippon
Nippon MA-3 Solo Bedienungsanleitung

Nippon
Nippon MA-3000 Bedienungsanleitung

Nippon
Nippon RA-4500 Bedienungsanleitung

Nippon
Nippon AQUA Bedienungsanleitung

Nippon
Nippon EMP-3 Bedienungsanleitung

Nippon
Nippon Mercury/PE-1 Bedienungsanleitung

Nippon
Nippon MA-3000 Bedienungsanleitung

Nippon
Nippon EMP-3 Bedienungsanleitung