FireSting®-PRO | User Manual
© PyroScience GmbH 6
3QUICK START
Step 1: Download the correct software and manual from the downloads tab of your
purchased device on www.pyroscience.com
Step 2: Connect the FireSting®-PRO meter with the micro-USB cable to the Windows
PC/laptop (Windows 7, 8, 10).
Step 3: Connect appropriate PyroScience sensor(s) to the optical sensor connectors 1to
4at the device (see 4).
Step 4: Connect an external temperature sensor (item no. TDIP15 or TSUB21) to the
Pt100 connector or, alternatively, an optical temperature sensor (see 5) to one of the
optical sensor connectors 1to 4 for automatic temperature compensation.
Step 5: Prepare appropriate calibration standards, as described in the respective sensor
manuals (see 8).
Step 6: Start the logger software by clicking on the short-cut "Pyro Workbench" on your
desktop.
Step 7: Open the Settings Wizard by clicking on the FireSting®-PRO picture. Select the
respective analyte and enter all sensor settings for each sensor, including an appropriate
mode of temperature compensation.
Step 8: Open the Calibration Wizard and follow the calibration instructions for each
sensor. Measurements with the respective sensors will start automatically after all
required sensor calibrations have been performed.
Step 9: Configure the Graphs according to your preferences.
Step 10: Activate Data Logging.