Rishabh Rish Eine Bedienungsanleitung

DMAN-00IM-0637 Rev_B 12/2017
OPERATING MANUAL
Rish Eine
Size: 48 x 96 Size: 96 x 96

Rish Eine
PROGRAMMABLE DPM V/A
AC Volt Meter (3 ) : Eine - 3VΦ
AC Ammeter (3 ) : Eine - 3AΦ
AC Volt Meter (1 ) : Eine - VΦ
AC Ammeter (1 ) : Eine - AΦ
Installation & Operating Instructions
Section Contents
1. Introduction
2. Measurement Reading Screens
3.
4. Installation
4.1 EMC Installation Requirements
4.2 Case Dimensions and Panel Cut-out
4.3 Wiring
Programming
3.1 Set Up Screens
4.4 Auxiliary Supply
4.5 Fusing
4.6 Earth / Ground Connections
5. Connection Diagrams
5.1 Connection Diagrams for EINE 3V
5.2 Connection Diagrams for EINE 3A
5.3 Connection Diagrams for EINE V
5.4 Connection Diagrams for EINE A
Specifications
6.
1

Available Models :
1. Introduction
The Eine Series is a panel mounted 96 x 96mm Digital Panel
Meters for the measurement of AC Voltage and current in three
The instrument integrates accurate measurement technology.
The parameters are displayed with Ultra high brightness LED
display with 14mm Digit height which enables to take readings
and Programmed at site for the following :
PT Primary, CT Primary, PT Secondary,CT Secondary.The front
panel has two push buttons for user interface to scroll through the
1. Eine 3V
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
X1000
L1
L2
L3 V
X1000
L1
L2
L3 A
2. Eine 3A
RMS
TRUE
X1000
L1
L2
L3 V
RISH Ein
X1000
V
3. Eine V
RMS
TRUE
X1000
L1
L2
L3 V
X1000
A
4. Eine A
RMS
TRUE
phase or single phase systems.
available parameters the two keys has function as follow:
V
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
1. : Scrolls through parameter in upward sequence. Display
2. : Scrolls the parameters in Reverse of above sequence.
V
L3 current, System current and back to L1 current.
sequence Eine 3V : L1 voltage,L2 voltage, L3 voltage, L1-L2
voltage, L2-L3 voltage, L3-L1 Voltage, System voltage and back
to L1 voltage. Display sequence Eine 3A : L1 current, L2 current
RMS
TRUE
from long distance. Programmable DPM can be configured
2

TABLE 1: Parameters Displayed TABLE 2: Parameters Displayed
Measured Parameters Unit of measurement
volt
L1 Voltage volt
L2 Voltage volt
L3 Voltage volt
L1-L2 Voltage
volt
L2-L3 Voltage volt
L3-L1 Voltage volt
System Voltage
Ampere
Ampere
Ampere
Ampere
L1 Current
L2 Current
L3 Current
System Current
with Eine 3V model with Eine 3A models
Measured Parameters Unit of measurement
2 . Measurement Reading Screen
In normal operation the user is presented with the measurement reading screens. These
screens may be scrolled through one at a time in incremental order by pressing the key
and in decrementing order by pressing key.
Measured Parameters Unit of measurement
L1 Voltage volt Ampere
L1 Current
Measured Parameters Unit of measurement
TABLE 3: Parameters Displayed TABLE 4: Parameters Displayed
with Eine V model with Eine A models
3

Screen 1 : Voltage L1
A. Display Screens of Eine 3V Models :
Screen 2 : System Voltage
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
(value displayed after “Sys” flashing on Display)
RMS
TRUE
RMS
TRUE
X1000
L1
L2
L3 V
X1000
L1
L2
L3 A
B. Screens of EINE 3A Models :
Screen 1 : L1 Current Screen 2 : System Current
(Value displayed after “Sys” flashing on Display)
X1000
L1
L2
L3 V
X1000
L1
L2
L3 A
RMS
TRUE
RMS
TRUE
4

3. Programming
The following sections comprise step by step procedures for configuring the EINE 3A/ A
and EINE 3V/V for individual user requirements. To access the set-up screens press
and hold the “ ” and “ ” Keys Simultaneously.
This will take the User into the Sys Type Screen (in case of 3A/3V) Followed by “Sys” on
Display (Section 3.1) or directly into the CT/PT Primary Screen (in case of A / V respectively) .
3.1 Set Up Screens
3.1.1. System type(for Eine 3V/3A)
This screen is used to set the system type. System type “3” for
3 phase 3 wire & “4” for 3 phase 4 wire system.
advances to the system type confirmation menu.
Pressing the “ down” key accepts the present value and
case of 3A) value Edit” menu (see section 3.1.2 for PT
advances to the “PT primary (in case of 3V) or CT primary (in
primary and 3.1.3. for CT primary). Pressing the “ up” key
will enter to the system type selection menu and scroll the values
through will enter values available. Pressing the “ down” key
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
X1000
V
X1000
V
RMS
TRUE
C. Display Screens of Eine V/A Models :
X1000
V
X1000
A
RMS
TRUE
Note: In Case 3P3W accuracy of I2 is 2% and Sys current is 1%.
5

System Type Confirmation
This screen will only appear following the edit of system type.
If system type is to be edit again,
Pressing the “ up” key will scroll between “3“ for 3 phase
3 Wire and “4” for 3 Phase 4 Wire.
Pressing the “ down” key sets the displayed value and
will advance to “Potential Transformer Primary Value Edit”
menu. (See section 3.1.2 for 3V & V) and “Current
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
transformer primary value (see section 3.1.3 for 3A & A)
3.1.2. Potential Transformer Primary Value (for 3V & V)
This screen displays “PtPr” message followed by previously set PT primary value on display.
For Eine 3V (3 phase) user can set PT primary value from 100VLL to 999kVLL & for
Eine V (single phase), user can set PT primary from 57.5(VLN) to 999 (kVLN).
Pressing the “ ” key accepts the present value and advances
to the “PT Secondary ” selection menu(See Section 3.1.4) for
Eine -3V and V.
Pressing the “ ” key will enter the “Potential transformer
Primary Value Multiplier Selection.
Eine 3V: Initially the “multiplier must be selected. Pressing
the “ ” Key will move the decimal point position to the sider ight
and show # # #. , after which it will again Shift to #.# #, # #.#,
Pressing the “ ” key accepts the present multiplier (Decimal Point position) and advances
to the “PT Primary value Edit” menu with decimal flashing to indicate cursor position.
# # #. with Annunciation of “x1000”, which indicates the value in kV.
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
Eine -V: Initially the “multiplier must be selected. Pressing the “ ” Key will move the decimal
point position to the Sideright and show # # .# #, # # #.# after which it will again shift
to #.# # #, # #.# #, # # #.# With annunciation of “x1000”, which indicates the value in kV.
PT Primary value Edit
Pressing the “ ” key accepts the present value at the cursor position and advances the
Pressing the “ ” key will scroll the value of the most significant digit.
cursor to the next Less significant digit.When the least significant digit has been set,
6

Potential Transformer Primary Value Confirmation
This screen will only appear following an edit of the Potential
Transformer Primary Value If the set value is to be corrected,
pressing the “ ” key will return to the “Potential Transformer
Primary Value Edit” stage. Pressing the “ ” key sets the
value and then advance to the “PT secondary screen”
Selection menu.
3.1.3 Current Transformer Primary Value (for 3A & A)
This screen displays “CtPr” message followed by previously set CT primary value on display.
This screen enables user to set CT primary from 1A to 999kA.
Pressing the “ ” key accepts the present value and
advances to the “CT Secondary Selection screen”
Pressing the “ ” key will shift decimal point position to right
ides and show #.# #, # #.#, # # #. , after which it will again shift
Pressing the “ ” key accepts the decimal point position and
enters into Current Transformer Primary value edit.
to #. # #, # #.#, # # #. with Annunciation of “x1000”, It indicates
the value in kA.
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
X1000
L1
L2
L3 V
X1000
L1
L2
L3 A
RMS
TRUE
Note : the flashing decimal point indicates the cursor position,
a steady decimal point will be present to identify the scaling of the number until the cursor
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
pressing the “ ” key will advance to the “Potential transformer
Primary Value Confirmation” screen. For Eine 3 V: When PT
primary is set < 100VLL then meter shows”Err” & again goes to PT
primary edit stage with the minimum PT primary value i.e. 100VLL
position coincides with the steady decimal point position. At this stage the decimal point will
Eine -V: When PT primary is set less than 57.5 VLN then meter
shows “Err” & again goes to PT primary edit stage with the
minimum PT primary value i.e. 57.5VLN.
be flashing.
Note: For Eine V(600VLN ) it will exit from set up menu.
7

Current Transformer Primary value Edit
Pressing the “ ” key will scroll the value of the most significant digit.
position and advances the cursor to the next Less significant digit.
When the least significant digit has been set, pressing the “ ”
key will advance to the “Current transformer Primary Value
Pressing the “ ” key accepts the present value at the cursor
confirmation” screen. When CT primary is set less than 1A
X1000
L1
L2
L3 V
X1000
L1
L2
L3 A
RMS
TRUE
, then meter shows “Err” and CT primary is set to 1A.
minimum CT primary value that user can set is 1A.
Note : the flashing decimal point indicates the cursor position, a steady decimal point will be
present to identify the scaling of the number until the cursor position coincides with the steady
decimal point position. At this stage the decimal point will be flashing.
Current Transformer Primary Value Confirmation
This screen will only appear following an edit of the Current
Transformer Primary Value.
If the set value is to be corrected, pressing the “ ” key will
return to the “Current Transformer Primary Value Edit” stage.
Pressing the “ ” key sets the value and then advance to the
“CT secondary” Selection menu.
X1000
L1
L2
L3 V
X1000
L1
L2
L3 A
RMS
TRUE
3.1.4. Potential Transformer Secondary Value (for 3V & V)
This screen displays “Pt-S” message followed by previously set PT secondary value on display.
For Eine 3V (3 phase) user can set PT secondary value from 100VLL to 500VLL & for
Eine V (single phase), user can set PT primary from 57.5(VLN) to 300VLN.
8

9
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
Pressing the “ ” key accepts the present value and advances
to the “Auto scroll” selection menu(See Section 3.1.6) for 3V.
In case of Eine V it will exit from set up menu.
Pressing the “ ” key will enter the “Potential transformer
Secondary Value Most significant digit Selection.
Pressing the “ ” key will scroll the value of the most significant digit.
position and advances the cursor to the next Less significant digit.
When the least significant digit has been set, pressing the “ ” key will advance to the
“Potential transformer secondary value confirmation” screen.For Eine 3V minimum PT secondary
Pressing the “ ” key accepts the present value at the cursor
can set is 100V(L-L).For Eine V If Pt secondary is less than 57.5 V(L-N) then meter shows
“Err” and PT secondary is set to 57.5 V(L-N).
Note: For Eine V decimal point is fixed at third digit position.
Potential Transformer Secondary Value Confirmation
This screen will only appear following an edit of the Potential
Transformer Secondary Value.
If the set value is to be corrected, pressing the “ ” key will
return to the “Potential Transformer Secondary Value Edit” stage.
Pressing the “ ” key sets the value and then advance to the
“Auto scroll” Selection menu in case of Eine -3V and in case of
Eine -1V it will exit from set up menu.
X1000
L1
L2
L3 V
X1000
L1
L2
L3 V
RMS
TRUE
L1
L2
L3
L1
L2
L3
X1000
A
X1000
A
3.1.5 Current Transformer Secondary Value (for 3A & A)
This screen is used to set Current transformer secondary
value. "1"for CT secondary 1A and "5" for CT secondary 5A.
pressing " " down key accepts the present value and
advances to the Auto Scroll option in case of Eine -3A and
for Eine -A it will exit from setup and goes to measurement
screen.