UFM UFM-50 Handbuch

UFM-50
Ultrasonic Doppler Flowmeter
MANUAL
Installation & Operating Instructions
Ultrasonic Doppler Flowmeter
Model: UFM-50
T +31 (0) 165 855 655
www.u-f-m.nl
U-F-M b.v.
Argon 24
4751 XC Oud Gastel
The Netherlands

Note: This page has been left blank intentionally.

P. 3
UFM-50 Doppler Flowmeter
INDEX
CONNECTIONS ................................................................................................................ 4
KEYPAD SYSTEM............................................................................................................ 6
CALIBRATION MENU..................................................................................................... 7
ICONS................................................................................................................................. 8
MESSAGE ICON............................................................................................................... 9
STATUS ............................................................................................................................. 9
PASSWORD..................................................................................................................... 10
MENU SELECTIONS...................................................................................................... 10
UNITS/MODE.................................................................................................................. 11
CALIBRATION................................................................................................................ 13
CALIBRATION (CONT.)................................................................................................ 14
RELAY PARAMETERS.................................................................................................. 15
DATA LOGGING ............................................................................................................ 16
COMMUNICATION (OPTIONAL) ................................................................................ 18
SPECIAL FUNCTIONS................................................................................................... 20
SENSOR MOUNTING..................................................................................................... 23
ENCLOSURE INSTALLATION..................................................................................... 27
FIELD TROUBLESHOOTING ....................................................................................... 28
COMMON QUESTIONS AND ANSWERS ................................................................... 31
WARRANTY.................................................................................................................... 33
OPTIONS ..........................................................................................................................34
MODBUS® COMMUNICATION ....................................................................................40
HART® COMMUNICATION ..........................................................................................51
SPECIFICATIONS ...........................................................................................................63
PIPE CHARTS ..................................................................................................................65
IMPORTANT NOTE: This instrument is manufactured and calibrated to meet product specifications.
Please read this manual carefully before installation and operation. Any unauthorized repairs or
modifications may result in a suspension of the warranty.
If this product is not used as specified by the manufacturer, protection may be impaired.
Available in Adobe Acrobat pdf format

P. 4
UFM-50 Doppler Flowmeter
CONNECTIONS:
POWER INPUT: The standard model requires AC power input between 100 to 240 VAC 50/60Hz. No
adjustments are necessary for voltages within this range. Connect L (Live) N (Neutral) and AC Ground.
Optional DC input model requires 9-32 VDC/10 Watts. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CSA/UL electrical safety standards, AC power input and relay
connection wires must have conduit entry to the instrument enclosure. Installation requires a switch,
overcurrent fuse or circuit breaker in the building (in close proximity to the equipment) that is marked
as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
QUICK BENCH TEST:
Connect Sensor as shown on next page, then Power. Test operation of the UFM-50 by holding the
sensor in one hand and rubbing your thumb or fingers briskly across the face (plastic surface) of the
sensor. Allow 15 seconds for the UFM-50 to process the signal and display a flow value.

P. 5
UFM-50 Doppler Flowmeter
CONNECTIONS
AC
L N
NO
C
NC
NO
C
NC
–
+
4-20mA
RLY2
RLY1
NC
C
NO
NC
C
NO
NC
C
NO
NC
C
NO
RLY3 RLY4 RLY5 RLY6
EXTRA RELAYS OPTION
POWER
INPUT
HEATER OPTION
SENSOR
GND
SERIAL COMMUNICATION OPTION
AC
GND
SENSOR
RCVR
GND
GND
TMTR
RS-485
Output
+ G

P. 6
UFM-50 Doppler Flowmeter
KEYPAD SYSTEM
The diagram on page 7 shows the UFM-50 menu system. Arrows show the four directions to leave a
menu box. Pressing a corresponding keypad arrow will move to the next item in the direction shown.
Move the cursor (highlighted) under numerals and increase or decrease numerals with the and
keys.
To store calibration values permanently (even through power interruptions), press the button.

P. 7
UFM-50 Doppler Flowmeter
CALIBRATION MENU
--Password----------
Password 0000
--Messages----------
LData og Logging
dLog Use 0.00000 %
Sensor Good
--Status------------
Velocity 0.00ft/s
Flow 0.00 USG/M
Min Flow 0.00USG/M
Signal Strength 0%
Signal Cutoff 10%
--Menu--------------
Units / Mode
Calibration
Relay Parameters
Data Logging
Communication
Special Functions
Simulation
Configuration
OPTIONAL FEATURES
--24 hr log----------
Date Feb. 12/2010
50138 USGTotal
34.82 USG/mAverage
52.20 USG/mMaximum
11:08:00Max Time
0.000 USG/mMinimum
9:15:00Min Time
USG/m
Tot 20130 USG
Relays 1 2 3 4 5 6
0.00
*
* Menu only appears if "New Password" has been
changed from 0000 in "Special Functions" menu.
--Data Logging-------
Log Site ID 0
Mode Flow
File Format .LG2
Date May 18/2018
Time 11:27:40
Interval 10sec
Data Log Logging
--Relay Parameters--
Relay 1
Function Flow
/mOn 1000 USG
/mOff 0.000 USG
--Simulation--------
Test Actual
Flow /m250USG
Flow4-20mA 5.60
Relays 1 2 3 4 5 6
--Configuration-----
Serial# 12345
Utility 2 0.51.2 .
Doppler 1. 6.11
CommBoard 1.18.0
6Relays
Analog Out 1
--Communication-----
Protocol Modbus
Address 001
BPS 9600
Parity Even
Stop Bits 1
--Calibration-------
Mode Flow
20mA 500 USG/m
4mA 0 USG/m
5.00 USG/mMin Flow
Signal Cutoff 10%
Damping 10%
Mode FIR
Percent 10%
Window 1.0 ft/s
Pipe ID 4.03 in
Cal Constant 1.000
--Special Functions-
Language English
Analog Out 4-20mA
Backlight High
Reset Totalizer NO
Neg. Totals No
Rev. Flow No
Capture Par No
Capture WF No
Restore Defaults No
New Password 0000
--Units/Mode--------
Mode Flow
Linear in
Volume USG
Multiplier x1
Decimal 0s
Velocity ft/s
Flow USG/m
Decimal 2s

P. 8
UFM-50 Doppler Flowmeter
ICONS

P. 9
UFM-50 Doppler Flowmeter
MAIN DISPLAY
The main display shows the units selected from the Units/Mode menu, flow or
velocity rate being measured, totalizer and relay states. The UFM-50 will start-up
with this display.
MESSAGE ICON
Press from the main display to view status of the data logger and error/
warning messages provided by the instrument. The Message Icon will appear on
the main display if error messages are being generated by the instrument. Press
to return to the main display.
Sensor Displays the status of the sensor hardware. Even with
the sensor disconnected or shorted, the flowmeter will
display Good. To check sensor integrity, use the sensor
resistance test procedure on page 28. If HardwareFault
is detected, navigate to the Configuration page in the
Main Menu to determine which circuit board is not
communicating.
STATUS
Press from the main display to view instrument status. Velocity will be
displayed in ft/sec or m/sec.
Velocity Displays the measured flow velocity in units selected in
the Units/Mode menu.
Flow Displays the measured flow rate in units selected in the
Units/Mode menu.
Min Flow Displays the minimum flow rate setting. This value is
read-only. The Min Flow can be changed in the
Calibration programming menu. A measured Flow
value below the Min Flow value will display as 0 flow
on the meter’s LCD display.
Signal
Strength Displays the strength of the received Doppler signal on
a 0-100% scale.
Signal Cutoff Displays the signal cutoff. This value is read-only. The
Signal Cutoff can be changed in the Calibration
programming menu. If Signal Strength is less than
Signal Cutoff, the meter will report 0 velocity and flow
on the LCD display. Setting may need to be adjusted in
the case of unstable flow near zero, or when high levels
of industrial noise are present.
USG/m
Tot 20130 USG
Relays 1 2 3 4 5 6
0.00
--Messages----------
LData og Logging
dLog Use 0.00000 %
Sensor Good
--Status------------
Velocity 0.00ft/s
Flow 0.00 USG/m
Min Flow 0.00USG/m
Signal Strength 0%
Signal Cutoff 10%

P. 10
UFM-50 Doppler Flowmeter
24 HR LOG
Press from the main display to view a formatted flow report from instruments
with a built-in data logger. Press to scroll down one day or repeatedly to scroll
to a specific date. Up to 365 days can be stored. Newest date will overwrite the
oldest. Press to return to the main display.
IMPORTANT: Inserting a USB drive into the meter while on this screen will
transfer 24 HR Log data to the USB drive in .csv format.
PASSWORD
The Password (a number from 0000 to 9999) prevents unauthorized access to the
Calibration menu.
From the Main Display press the key to get to Password. Factory default
password is 0000 and if it has not been changed, this screen will be bypassed
completely.
A new password can be stored by going to the Special Functions New
Password menu.
If a user password is required, press to place the cursor under the first digit and
or to set the number, then to the second digit, etc. Press or to proceed
to the Menu Selections screen.
MENU SELECTIONS
The Menu selections page is used to navigate to specific menus which are
described in more detail on the following pages.
Press or to navigate to different menus, and to enter the selected menu.
--Menu--------------
Units / Mode
Calibration
Relay Parameters
Data Logging
Communication
Special Functions
Simulation
Configuration