Viessmann VitoAid Bedienungsanleitung

Operating instructions
for the system user
VIESMANN
Ventilator for continuous respiratory support
VITOAID
6152712 GB 6/2020 Please keep safe.

2
Please follow these safety instruc-
tions closely to prevent accidents
and material losses.
Safety instructions explained
Danger
This symbol warns against the risk
of injury.
!Please note
This symbol warns against the risk
of material losses and environmen-
tal pollution.
Note
Details identified by the word "Note"
contain additional information.
Target group
These operating instructions are inten-
ded for system users.
Before using the device, the operator
must have fully read and understood
these operating instructions.
The ventilator may only be used under
the direction of a physician by trained
specialist medical personnel with knowl-
edge of mechanical ventilation.
The ventilator is not intended for use by
patients, children, individuals with
reduced physical, sensory or mental fac-
ulties, or persons lacking the appropriate
experience and knowledge.
Safety instructions for working with the device
Connecting the device
■The device may only be put into opera-
tion by trained specialist medical per-
sonnel with knowledge of mechanical
ventilation.
■The device may only be connected to
the power supply at permanently instal-
led 230 V standard sockets using the
factory-installed power cable with
standard plug.
■Connection via extension cables or
sockets switched by means of wall or
dimmer switches is not permissible.
■Modifications to the existing installation
may only be carried out by authorised
specialists.
Danger
Incorrectly performed connections
can lead to life threatening condi-
tions for the ventilated patient.
The breathing tube of the VentStar
Oxylog VE300 single-tube system
and the oxygen supply tube may
only be connected by trained spe-
cialist medical personnel with
knowledge of mechanical ventila-
tion.
Safety instructions
For your safety
6152712

3
Working on the device
■All settings and work on the device
must be performed as specified in
these operating instructions.
No other work on or modifications to
the device are permissible.
■Do not open the device.
■Do not remove casings.
■Do not modify or remove attachments.
Safety instructions for operating the device
Danger
Incorrect settings and use of the
wrong components can put the ven-
tilated patient at risk.
■Ventilation must only be provided
via the VentStar Oxylog VE300
single-tube system, 1.5 metres
(standard delivery) and HME filter.
■The device may only be operated
with an externally controllable oxy-
gen supply.
■Regularly check tube connections
during operation. Do not discon-
nect tube connections.
Danger
Contamination and dirt on the
device and on the tubes can harm
the ventilated patient.
■Check connections and tubes
daily for contamination and dirt.
■Immediately remove any fluids
and biological matter.
■Keep the air inlet grille on the
device clear at all times. Before
every use, check the air inlet grille
for contamination and dirt.
Permissible ambient conditions
!Please note
Incorrect ambient conditions can
result in damage to the device and
put safe operation at risk.
■Only use the device in dry,
enclosed spaces and at ambient
temperatures between 5 °C and
35 °C.
■Do not use the device in MRT
environments.
■The device must not be used in
hyperbaric chambers.
General information on the safe use and handling of oxygen
Source: Bundesanstalt für Materialfor-
schung und -prüfung [German Federal
Institute for Materials Testing and
Research], Berlin
Safety instructions
For your safety (cont.)
6152712

4
General safety instructions
■Do not allow oxygen to flow into
enclosed rooms/spaces as this can
lead to an oxygen-enriched atmos-
phere and consequently may present a
fire hazard.
■Keep away from open flames and
sources of ignition.
Safety instructions for medical per-
sonnel and users
■Oxygen must only be handled by
trained or instructed personnel
■Secure oxygen bottles to prevent them
from tipping over or falling down.
■Always open oxygen valves slowly.
■Keep oxygen fittings and system com-
ponents free of oil and grease at all
times.
■On completion of work, close all oxy-
gen fittings and valves.
■Improper use of oxygen is prohibited,
e.g. using it instead of compressed air
or for cooling or cleaning clothing
■Under no circumstances should leak-
ing or faulty oxygen bottles, fittings or
system components be used.
Safety instructions
For your safety (cont.)
6152712

5
1. Introductory information Symbols ................................................................................................. 6
Standard delivery ................................................................................... 6
Intended use .......................................................................................... 6
Your device is preset at the factory ........................................................ 7
2. Device overview Overview of connections and controls ................................................... 8
3. Connecting VitoAid Connecting the oxygen supply and breathing tube ............................... 9
Adjusting the oxygen supply .................................................................. 10
4. Switching on and off Switching on VitoAid .............................................................................. 11
Switching off VitoAid .............................................................................. 11
5. Where to find the controls Operating VitoAid ................................................................................... 12
■Screen displays .................................................................................. 12
6. Switching ventilation on
and off
Switching on ventilation ......................................................................... 14
Switching off ventilation ......................................................................... 14
7. Setting the ventilation
parameters
Selecting the ventilation mode ............................................................... 15
■Pressure-controlled ventilation (PCV) ................................................ 15
■High-flow CPAP (continuous positive airway pressure) ...................... 15
Setting the positive end-expiratory pressure (PEEP) ............................ 15
Setting the inspiratory pressure ............................................................. 15
Setting the respiratory rate .................................................................... 16
Setting the ratio of inspiratory time to expiratory time ............................ 16
8. Setting alarm limits Setting the lower limit for respiratory volume ......................................... 18
Setting the upper limit for respiratory volume ........................................ 18
9. Further settings Restoring factory settings ...................................................................... 19
10. Checks Checking values .................................................................................... 20
11. Alarms and fault mes-
sages
Acknowledging alarms and fault messages ........................................... 21
12. What to do if... is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
and a numeric code > are shown on the display. ........................... 23
13. Maintenance Cleaning the device surface .................................................................. 24
Cleaning the air inlet grille ..................................................................... 24
14. Service VitoAid Hotline ....................................................................................... 25
Index
Index
6152712

6
Symbol Meaning
Reference to other document containing
further information
1.
Step in a diagram:
The numbers correspond to the order in
which the steps are carried out.
Warning of material losses and environ-
mental pollution
Live electrical area
Pay particular attention.
■Component must audibly click into place.
or
■Acoustic signal
■Fit new component.
or
■In conjunction with a tool: Clean the sur-
face.
Dispose of component correctly.
Dispose of component at a suitable collec-
tion point. Do not dispose of component in
domestic waste.
Standard delivery
■VitoAid ventilator
■VentStar Oxylog VE300 single-tube system, length:
1.5 metres
Intended use
The VitoAid ventilator is designed to provide continu-
ous respiratory support to adult patients who require
mechanical ventilation.
The VitoAid ventilator may only be used in emergency
situations where the capacities of the existing supplies
of approved ventilators are not sufficient.
The ventilator is intended solely for use in care facili-
ties and hospitals for invasive and non-invasive posi-
tive pressure ventilation.
Intended use requires the following:
■Ventilation with external oxygen supply:
The ventilator is connected via a silicone tube to a
supply of medical oxygen with pressure control.
■Ventilation without external oxygen supply:
The tube connection for the oxygen supply to the
ventilator is closed (delivered condition).
Introductory information
Symbols
6152712

7
■The patient is supplied via the VentStar Oxylog
VE300 single-tube system, 1.5 metres (standard
delivery) and HME filter. The HME filter must be
positioned close to the patient.
■The supply tubes comply with the specification for
ventilators.
Usage for a purpose other than invasive and non-inva-
sive positive pressure ventilation shall be deemed
inappropriate.
Incorrect usage or operation of the device (e.g. open-
ing of the device) is prohibited and will result in an
exclusion of liability. Incorrect usage also occurs if the
components of the ventilator or the tubes connected to
it are modified from their intended function.
Your device is preset at the factory
The ventilator is preset at the factory for pressure-con-
trolled ventilation (PCV) with the following parameters:
You can change these settings individually at any time:
See page 15 onwards.
Ventilation parameters
■The positive end-expiratory pressure (PEEP) is set
to 10 millibars.
■The inspiratory pressure is set to 25 millibars.
■The respiratory rate is set to 15 breaths per minute.
■The ratio of inspiratory time to expiratory time is set
to 1 : 2.
Alarm limits
■The alarm limit for expiratory pressure equates to the
setting for positive end-expiratory pressure (PEEP).
■The alarm limit for inspiratory pressure is 45 milli-
bars.
■The alarm limit for the lower respiratory volume is set
to 0. This means that monitoring for this value is
deactivated.
■The alarm limit for the upper respiratory volume is
set to 999. This means that monitoring for this value
is deactivated.
Note
The alarm limits are only active if PCV (pressure-con-
trolled ventilation) is set: See page 15.
Power failure
All settings are retained if there is a power failure.
Introductory information
Intended use (cont.)
6152712

8
A
B
C
D
G
F
H
K
E
Fig. 1
AConnection for breathing tube of the VentStar Oxy-
log VE300 single-tube system (standard delivery)
■Internal 7 of breathing tube: 22 mm
■Tube length: 1.5 metres
BTube connection for oxygen supply
■Internal 7: 6 mm
CTransport handle
DON/OFF switch
EBuzzer for acoustic signals
FConnection for factory use only: Do not connect
anything here!
GStatus LED
HOperating unit with 2-part display and operating
buttons
KAir inlet grille: It is essential to keep this grille clear.
Danger
If it is not possible to draw in a sufficient amount
of air through the air inlet grille, the ventilation
parameters may deviate from the set values.
This could potentially harm the patient.
■Keep the air inlet grille clear at all times.
■Before using the device, always check the air
inlet grille for contamination and dirt.
■If required, clean the air inlet grille with a vac-
uum cleaner.
Device overview
Overview of connections and controls
6152712

9
2.
3.
C
B
D
A
Fig. 2
AConnection for breathing tube
BTube connection for oxygen supply
COxygen supply tube
■Internal 7: 6 millimetres
DVentStar Oxylog VE300 single-tube system with
HME filter
■Internal 7: 22 millimetres
■Tube length: 1.5 metres
1. Pull the protective caps off tube connections A
and B.
2. Push oxygen supply tube C onto tube connection
B.
3. Push the breathing tube of the VentStar Oxylog
VE300 single-tube system with expiration valve
and HME filter D onto tube connection A.
4. Secure both tubes, e.g. with tube clips, to prevent
accidental disconnection.
5. Secure the ventilator to prevent it from accidentally
tipping over.
Connecting VitoAid
Connecting the oxygen supply and breathing tube
6152712

10
The inspiratory oxygen concentration is adjusted via
the flow rate of the oxygen supply according to the
curve shown below.
Note
To measure the oxygen flow rate, a flow meter needs
to be integrated into the oxygen supply.
80
60
20
0
100
Oxygen flow rate in l/min
oxygen concentration in %
5 10 15 20 25
40
Inspiratory
Fig. 3 Curve for adjusting the oxygen supply
Connecting VitoAid
Adjusting the oxygen supply
6152712
Dieses Handbuch passt für folgende Modelle
1
Inhaltsverzeichnis
Andere Viessmann Lüfter Handbücher

Viessmann
Viessmann Tecto Refrigo CMC1 Bedienungsanleitung

Viessmann
Viessmann VITOVENT 300 Betriebsanleitung

Viessmann
Viessmann Vitovent 300-W Bedienungsanleitung

Viessmann
Viessmann Vitovent 300-W Betriebsanleitung

Viessmann
Viessmann EVO-COOL CS 900 Bedienungsanleitung

Viessmann
Viessmann Vitovent 300-W Bedienungsanleitung

Viessmann
Viessmann VITOVENT 300 Bedienungsanleitung

Viessmann
Viessmann VITOVENT 300 Kurzanleitung

Viessmann
Viessmann Vitovent 300-W Kurzanleitung

Viessmann
Viessmann Vitovent 300-W Kurzanleitung