BRIGHT OTF6000 Bedienungsanleitung

Instruments
Instruments
Instruments
Instruments
Instruments
Instruments
OTF6000
Cryostat
Manual

Bright Instruments 2
Contents
SAFETY WARNINGS 4
Low Temperature 4
Knives / Blades 5
SAFETY INFORMATION 6
Consumer Protection 6
ELECTRICAL 6
MICROTOME KNIVES & BLADES 6
OPERATION 7
LOW TEMPERATURES 7
ACCESSORIES 7
PRODUCT SAFETY SUGGESTIONS 7
DECONTAMINATION CERTIFICATES 8
WARRANTY 9
1 INTRODUCTION 10
1.1 Receipt of Product 10
1.1.1 Receipt and unpacking 10
1.1.2 Receipt 10
1.1.3 Unpacking 10
1.1.4 Unpacking Procedure 10
1.2 Assembly and Installation 11
1.2.1 Fitting the Hand Wheel 11
1.2.2 Drain Tube and Plug 11
1.2.3 Frost Elimination Cover 11
1.2.4 Positioning 11
1.2.5 Installation and removal of 12
microtome
1.3 Electrical 13
1.3.1 Settling 13
1.3.2 Electrical Requirements 13
1.3.3 Electrical Safety 13
2 OPERATING INSTRUCTIONS 14
2.1 Touchscreen 14
2.1.1 Lock Screen 14
2.1.2 Home Screen Low Spec 15
2.1.3 Home Screen High Spec 16
2.1.4 Temperature Settings: Specimen 17
2.1.5 Temperature Settings: Chamber 18
2.1.6 Motor Drive 19
2.1.7 Motor Drive: Temps Screen 20
2.1.8 Settings: User Account 21
2.1.9 Settings: Admin Account 22
2.1.10 Settings: UV Decontamination 23
2.1.11 Settings: Defrost Control 24
2.1.12 Settings: User Management 25
2.1.13 Settings: Set Clock 28
2.1.14 Settings: Event Log 29
2.2 Cabinet Controls 31
2.2.1 Around Touchscreen 31
2.2.2 Mains Switch 32
2.2.3 Handwheel 32
2.2.4 Quick Freezer Plate 32
2.2.5 Motorised Advanced/Rewind 32
2.2.6 Foot Switch 33
2.3 Microtome Controls Solid Knife 34
2.3.1 Knife Guard 34
2.3.2 Knife Holder 34
2.3.3 Anti-Roll Assembly 34
2.4 Microtome Controls Quick Rel. 35
2.4.1 Blade Guard 35
2.4.2 Anti-Roll Assembly 36
2.5 Common Microtome Controls 37
2.5.1 Knife Block Adjustment 37
2.5.2 Section Thickness Control 37

Bright Instruments 3
Contents Cont.
3 CUTTING AND COLLECTING 38
SECTIONS
3.1 Basic Freezing Techniques 38
3.1.1 Liquid Nitrogen 38
3.1.2 Carbon Dioxide (Gas) 38
3.1.3 Carbon Dioxide (Solid) 38
3.1.4 Freezer Plates 38
3.2 Fitting the Specimen 38
3.3 Trimming the Specimen 40
3.4 Section Cutting and Collection 40
3.5 Hints on Sectioning 40
3.6 Frozen Sectioning 41
3.6.1 Solutions Checklist 41
3.6.2 Specific Problems 42
3.7 Cutting Temperatures 43
3.7.1 Fresh Unfixed Tissue 43
3.7.2 Fixed Tissue 43
4 CARE & MAINTENANCE 44
4.1 Daily Care 44
4.2 Automatic Defrost Cycle 44
4.3 Total Defrosting 44
4.4 Microtome Knives 45
4.5 Operating in extreme conditions 45
4.6 Servicing and Repairs 45
4.6.1 UK 46
4.6.2 Rest of world 46
4.7 Simple Servicing/Repair 46
4.8 Decontamination 46
5 SPECIFICATIONS 48
5.1 Cryostat 48
5.2 5050 Microtome 49
ACCESSORIES 50
Product 51
Low Spec 51
High Spec 51
Equipment Options 52
Feed Screws for /MAR Option 52
Orientation 52
Object Holders 52
Knives & Blades 53
Anti-Roll Plates 53
Ancillary Items 53
MATERIAL SAFETY DATA SHEETS 54
Cryo-M-Bed 53581 55
Cryospray 1234ze 56
Low Temperature Oil 57491 62
REFERENCES & DRAWINGS 64
Figure 1. OTF6000 Cabinet 65
Figure 2. 5050 Microtome Solid Knife 66
Figure 3. 5050 Microtome Solid Knife 67
Anti-Roll Assembly
Figure 4. 5050 Microtome Quick 68
Release Feather Blade
Holder (QRFBH)
Figure 5. 5050 Microtome Quick 69
Release Feather Blade
Holder (QRFBH) Anti-Roll
Assembly
DECONTAMINATION CERTIFICATE 70
QUALITY SURVEY REPORT 71

Bright Instruments 4
SAFETY WARNING
Low temperatures are present in this
equipment. Extreme care should be taken.
DO NOT let bare skin come into contact
with metal surfaces.

Bright Instruments 5
SAFETY WARNING
Extremely sharp knives/blades.
Use knife/blade guards at all times.
Use correct tools for removal and insertion of knives/blades.
DO NOT leave knives/blades laying around.
Place knives/blades not in use, into box/wallet provided.

Bright Instruments 6
Safety Information
CONSUMER PROTECTION
The Consumer Protection Act 1987 Part 1,
refers to Product Liability. This legislation was
issued as a direct result of an EC Directive to
all member states and has been in force with
effect from 01 March 1988.
Bright Instrument Company Limited, ever
mindful of the need to ensure that their
products are not subject to misuse and/or
incorrect handling, have made it their aim to
communicate any possible dangers to their
customers.
Whilst Bright Instrument Company Limited
markets products manufactured to the highest
safety standards, it is in the interest of the
purchaser that he is aware of the resultant
dangers of misuse and/or incorrect handling
of these products.
Your attention is therefore drawn to the
following precautions:
Electrical Microtome Knives
& Blades
a. Warnings - A warning notice is fixed to
the instrument stating that it should be
disconnected from the power supply before
removing the panels. This warning should
be strictly observed. This cryostat is fitted
with an in line mains filter which may affect
portable appliance test results.
b. Fuses - Fuse ratings are clearly indicated
on all fuse panels adjacent to the fuse holder.
If and when replacement is necessary, the
correct fuse rating must be adhered to.
c. Earthing (Grounding) - A protective earth
terminal is fitted, and must be used in all two
wire installations.
NB. It is recommended the OTF6000
is plugged in via an electrical surge
protection unit.
Microtome knives can be hazardous in the
laboratory. Personnel should be made aware
of the dangers and observe the following
warnings:
a. Do not leave the microtome unattended
with an exposed knife in position. Remove the
knife, or cover it with the guards provided.
b. Do not leave knives lying around. Place
knives that are not in use in their boxes.
c. Do not carry knives unless secure in the
box provided.
d. Do not clean the knife along its length.
Wipe from the back edge of the cutting edge.
e. Remember that even used knives are
dangerous – they are still sharp and may
have been used to cut potentially infectious
specimens.
f. Dispose of used knives with the same care
as other sharp objects. On no account should
used knives be placed in waste bins.

Bright Instruments 7
Accessories
Low Temperatures
Operation
Product Safety
Suggestions
Fluids supplied as accessories with Bright
instruments, such as Cryospray 134,
Cryo-M-Bed and microtome oil, are strictly for
laboratory use only. They should not be taken
by mouth and precautions afforded to other
laboratory chemicals should be adhered to.
Parts of this instrument may attain
temperatures as low as -50ºC. It is important
to avoid bare skin touching things such as
cold surfaces. When in doubt, wear gloves.
Avoid touching the knife, it is cold as well as
sharp! Always use knife guards.
When placing object holders in the microtome,
when orientating, manipulating or in any way
placing fingers in a position above the knife
edge - ensure that the handwheel is locked
and knife guards are in position.
All Bright Instrument Company Limited
personnel are encouraged to make
suggestions regarding product safety. We
also welcome such suggestions from our
Customers. They may be submitted by
completing the appropriate (Safety) section
of the Quality Survey Report Form supplied
with all Bright instruments, or alternatively
by letter, telephone or email:
All communications should be direct to our
Warranty Assurance Department and will be
acknowledged.

Bright Instruments 8
Decontamination
Certificates
IMPORTANT
If the instrument or any part of it is to be
returned to Bright Instrument Company
Limited, a decontamination certificate must
accompany it, and please note the following:
a. If the instrument or any part of it has been
exposed to or been in contact with potential
pathogenic or radioactive material, it is
essential that it be decontaminated.
Set procedures are laid down by the Health
and Safety Executive for decontamination.
For the avoidance of doubt, we ask that
instruments or parts returned to us should be
accompanied by a completed decontamination
certificate. A copy of this can be found at
the back of this instruction manual and we
suggest you use a photocopy of this when
returning parts.
b. Alternatively you can download a
decontamination certificate from our website.
c. Should the instrument or any part of it be
received in a condition that Bright Instrument
Company Limited, consider to be a potential
biological hazard, the instrument or part will
be returned un-repaired at the expense of the
customer.
d. Overseas customers declarations must
indicate that the package contains ‘British
Returned Goods’. Failure to do so will involve
customs duty payable by us, which will be
invoiced to the sender.

Bright Instruments 9
Warranty
1. The Seller ‘s products are carefully
inspected and submitted to its standard tests.
2. The Seller warrants all its products to be
free from defects in workmanship and
materials under normal conditions of use and
service provided always:
a. That if any of the goods so manufactured
is alleged to be defective in workmanship and
material and is returned carriage paid, and
protected against damage in transit to the
Seller’s works at Luton within 12 months from
the date of despatch and if after examination
by the Seller that goods or part of them are
found to be so defective then the Seller will
repair or replace them free of charge and will
return them to the Buyer, carriage paid.
b. Where any part of the goods manufactured
by the Seller is repaired or replaced under
the terms of the foregoing warranty, such
warranty shall thereafter be limited to a
period of six months from the date when
the goods shall have been re-delivered
to the Buyer.
c. This warranty does not apply to any defects
caused by wear and tear, incorrect installation
abnormal conditions of working, accident,
misuse or neglect.
d. That save as in this clause herein before
expressed, the Seller shall not be under any
liability for negligence or otherwise in respect
of defects in goods delivered or for any injury,
damage or loss resulting from such defects
and the Seller’s liability under this clause shall
be in lieu of any warranty or condition implied
by law as to the quality or fitness for any
particular purpose of such goods.
e. This warranty is expressly in lieu of all
other warranties, guarantees or liabilities
expressed or implied by any of the Seller’s
Representatives or Agents.
Please see our separate Product Warranty
sheet for deliveries to the mainland UK.
WARNING: Before proceeding to Operating
Instructions, ensure you are familiar with
the contents of the pages marked ‘Safety
Information’. This instrument must only be
used by competent persons.

Bright Instruments 10
Based on the long-established and reliable
OTF cryostat range, the new OTF6000 brings
Bright cryostats completely up to date. New
styling with improved user ergonomics, the
latest blade systems in the ever-reliable and
powerful 5050 microtome, a huge choice of
options plus money-saving package deals
make these cryostats absolutely unique.
They are suitable for an endless range of
applications and in the correct configuration
are capable of cutting a wide diversity of
specimens from undecalcified bone to brain,
resins, plastics and plant tissue as well as
more usual soft tissue. Allowing full anti-roll
plate adjustment for perfect results coupled
with long lasting temperature stability, the
OTF6000 can truly be described as complete.
1.1 RECEIPT OF PRODUCT
1.1.1 RECEIPT AND UNPACKING
This instrument received a final test and
inspection prior to despatch from the factory.
The following instructions are given for the
re-assembly of the instrument, adjustments
and its correct use. If the instrument is
received before preparations for installation
are completed. It should be stored in a clean,
dry place and not exposed to dirty or damp
conditions.
1.1.2 RECEIPT
Immediately upon receipt of the instrument,
make a careful examination for evidence of
damage encountered in transit. If any damage
is found or suspected, notify both the carrier
and Bright Instrument Company Limited
immediately.
1.1.3 UNPACKING
All packing must be carefully removed and
parts checked against the enclosed packing
list. If any damage or discrepancy is noted,
please inform our agent/distributor or Bright
Instrument Company Limited immediately.
1.1.4 UNPACKING PROCEDURE
The following should be followed when the
instrument is received:
a. At least two people are required during the
unpacking.
b. Check that the outer packing is in good
order and does not show signs of serious
damage.
c. Remove the microtome from the case and
unpack it.
d. Remove and unpack the accessories.
1. Introduction
Inhaltsverzeichnis
Andere BRIGHT Laborausrüstung Handbücher
Beliebte Laborausrüstung Handbücher anderer Marken

Agilent Technologies
Agilent Technologies 5800 ICP-OES Bedienungsanleitung

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Bedienungsanleitung

NI
NI PXI-5422 Handbuch

Collomix
Collomix Aqix Bedienungsanleitung

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Bedienungsanleitung

Ocean Insight
Ocean Insight FLAME-NIR+ Bedienungsanleitung

Parker
Parker ALIGN-MG-NA Bedienungsanleitung

BD
BD 644787 Bedienungsanleitung

DENTAURUM
DENTAURUM Compact Megaplus Bedienungsanleitung

Biuged Laboratory Instruments
Biuged Laboratory Instruments BGD 626 Bedienungsanleitung

VWR
VWR SAS Super IAQ Bedienungsanleitung

illumina
illumina MiSeqDx Bedienungsanleitung