4
ENG
USER MANUAL
Centrifuge CM-7S Plus
General safety
•Use only as specied in the operating instruc-
tions provided.
•The unit should not be used if it has encoun-
tered a physical shock or has been dropped.
•The unit must be stored and transported in a
horizontal position (see package label).
•After transport or storage, allow the unit to dry
out (2-3 hrs) before connecting to mains power.
•It is necessary to observe the safety area of 300
mm around the centrifuge in accordance with
EN-61010-2-20. Persons and hazardous materi-
als must not be located in the safety area whilst
the centrifuge is in operation.
•Use only original accessories (rotors, adaptors,
etc.) provided by the manufacturer and ordered
specically for this model.
Electrical safety
•Connect to the mains only with a voltage cor-
responding to that on the serial number label.
•Ensure that the switch and plug are easily ac-
cessible during use.
•Do not plug the unit into the main outlet with-
out grounding, and do not use an extension
lead without grounding.
•Before moving the unit, disconnect it from the
mains. To turn o the unit, disconnect the pow-
er plug from the mains outlet.
•It is the user’s responsibility to carry out appro-
priate decontamination if hazardous material is
spilt on or inside the equipment. If liquid is split
inside the unit, disconnect it from the mains
and have it checked by a competent person.
During operation
•Do not centrifuge ammable or chemically vig-
orously reactive materials. If such liquids are
spilled in the rotor or rotor chamber, the centri-
fuge must be cleaned with a moist cloth and a
mild soap solution.
•Do not use rotors with visible signs of corrosion,
wear or mechanical damage.
•Do not ll in the containers after they are insert-
ed in the rotor.
•Do not leave the operating unit unattended.
•Do not operate the unit in environments with
aggressive or explosive chemical mixtures.
•Do not operate the unit if it is faulty or been in-
correctly installed.
•For indoor use only.
•Do not use outside laboratory rooms.
•Before using any cleaning or decontamination
method except those recommended by the
manufacturer, check with the manufacturer
that the proposed method will not damage the
equipment.
•Do not make modications to the unit.
Biological safety
•Without a bioseal, the centrifuge is not a biosafe-
ty system in accordance with EN61010-2-20 and
cannot be used for centrifuging hazardous ma-
terials contaminated with toxic, radioactive or
pathogenic microorganisms.
•It is the user’s responsibility to carry out appro-
priate decontamination if hazardous material is
spilt on or inside the equipment.
Preparation for use
•Unpack the centrifuge.
•If the machine was stored at a temperature be-
low 0°C, make sure that it is kept at room tem-
perature for at least two hours before turning
it on.
•Pleace the centrifuge onto a level, horizontal
surface.
•Examine the power plug, power cord and ap-
pearance of centrifuge for damadge.
•Connect the power cord to the centrifuge, in-
sert the plug in to the socket and press network
switch at ( I ) symbol. Display will light up on the
control panel.
•Open the lid by pressing the OPEN button.