5.3 Replacing the Optional Recorder and Probe(s)
1. Open the incubator door, and locate the probe mounting
plate attached to the center of the right interior wall.
Remove the mounting plate.
2. Single pen recorders have only one long stainless steel
probe, attached to the lower end of the back of the
mounting plate. Remove the probe by carefully sliding it
out of the housing.
IMPORTANT! Use extreme care when installing the new
recorder to avoid severely bending the probe capillary tubes.
Dual pen recorders have the stainless steel probe plus a wet
bulb probe attached to the back of the mounting plate. Remove
the screws securing the mounting plate and slide the wet bulb
probe out of its housing.
3. Remove the screws securing the ceiling of the incubator
and remove the ceiling.
4. Remove the top three screws on both edges of the right
duct sheet.
5. Lean the duct sheet out in order to remove the
Permagum seal from around the probe access hole.
6. Open the control panel door, and remove any Permagum
from around the access hole.
7. Pull the probe(s) carefully up through the hole.
8. Follow the probe cable(s) to the back of the recorder,
and carefully clip any plastic ties holding the cable(s) to
other wiring.
9. Remove the three screws securing the recorder and pull
it carefully out from the front of the control panel.
10. Replace the recorder with the correct Forma part.
Note: When replacing the recorder and probe(s), retie the
probe cable(s) to the existing wires.
IMPORTANT! Use extreme care when the new recorder to
avoid severely bending the probe capillary tubes.
5.4 Removing the Top Section
If it becomes necessary to remove the top of the incubator
for moving it through low doorways, use the following proce-
dure:
• Read the instructions completely before starting the
removal process.
• Provide adequate space and sufficient lighting to perform
the work.
• Disconnect the inlet water supply and the drain.
• Before lifting the top section from the incubator, provide
carpenter’s horses or another suitable support arrangement
so that the underside of the top is suspended.
Section 5 - Service
Servicing must be performed by qualified service
personnel only!
De-energize all potential sources of
energy to this unit and lockout/tagout
their controls. (O.S.H.A. Regulation,
Section 1910-147.)
5.1 Access to the Electrical Components
To gain access to the electrical components, open the con-
trol panel and grasp the left side of the control panel housing.
Pull gently, but firmly, up and out.
5.2 Replacing the Overtemp/Undertemp Probe and
Thermostat
1. Remove the incubator ceiling by remove the screws
holding it in place.
2. Remove the top three screws from the top of the right
duct cover.
3. Lean the duct sheet out, and remove the Permagum seal
from around the probe access hole.
4. Remove the two plastic clips that hold the probe in place
and remove the probe.
5. Open the control panel.
6. Pull the overtemp probe up through the access hole and
into the control panel.
7. Follow the wires from the probe to the thermostat
mounted on the control panel. Clip the plastic ties hold-
ing the overtemp cable to the existing wiring.
8. Pull the overtemp knob on the control panel off.
9. Remove the two screws that hold the overtemp assembly
to the control panel.
10. Disconnect the two wires from the back of the thermo-
stat assembly.
11. Pull the entire assembly out of the panel, and remove the
unit.
12. Replace the thermostat and probe.
Note: Reseal the probe access hole with Permagum, and retie
the overtemp cable to the existing wires after replacing the
probe.
Model 3920 _______________________________________________________________________________Service
5 - 1