Hitachi EUP-O732T Bedienungsanleitung

0123
Intraoperative Probe
EUP-O732T
INSTRUCTION MANUAL
Notes for operators and responsible maintenance personnel
★ Please read through this Instruction Manual carefully prior to use.
★ Keep this Instruction Manual together with the system with care to
make it available anytime.
Tokyo , Japan
Q1E-EP1234-7
© Hitachi, Ltd. 2013,2017. All rights reserved

(1) Q1E-EP1234
Manufacturer:
Hitachi,Ltd
2-16-1,Higashi-Ueno,Taito-ku,Tokyo,110-0015,Japan
+81-3-6284-3668
http://www.hitachi.com/businesses/healthcare/index.html
European
Representative:
Hitachi Medical Systems GmbH
Otto-von-Guericke-Ring 3 D-65205 Wiesbaden, Germany
EU Importer:
Address:
Hitachi Medical Systems Europe Holding AG
Sumpfstrasse 13 CH-6300 Zug, Switzerland
Local
Distributor:

(2) Q1E-EP1234
About this manual
This instruction manual contains safety precautions, the inspection,
the operation procedure and the reprocessing procedure of EUP-O732T.
Please read this manual thoroughly to ensure the safety operation.
If you have any questions concerning the operation of the probe, please
contact a service support.
The following conventions are used throughout the manual to denote
information of special emphasis.
WARNING: “Warning” indicates the presence of a hazard which may
result in severe personal injury, substantial property
damage, or death if the warning is ignored.
CAUTION: “Caution” indicates the presence of a hazard which will or
can cause minor personal injury or property damage if the
caution is ignored.
NOTICE: “Notice” indicates information of installation, operation,
or maintenance, which is important, but not hazard related.

(3) Q1E-EP1234
Graphical Symbols for Use in Labeling of Hitachi Ultrasound Probes
Some graphical symbols that are used in labeling of Hitachi Ultrasound Probes are
compliant with EN980:2008 standard. Refer to the following table about the meanings
of them.
Explanation of Symbol Symbol Descriptive Content
Manufacturer Company
Name and Address
Hitachi, Ltd.
2-16-1, Higashi-Ueno, Taito-ku,
Tokyo,110-0015, Japan
+81-3-6284-3668
http://www.hitachi.com/businesses/
healthcare/index.html
Authorized
Representative
in The European
Community
Hitachi Medical Systems GmbH
Otto-von-Guericke-Ring 3 D-65205
Wiesbaden, Germany
Keep away
from Sunlight
Store the probe in a cool place and keep
away from high temperature, high humidity,
or direct sunlight.
Contains or presence
o
f
natural rubber latex
Contains or presence of natural rubber
latex
Do not re-sterilize
Do not re-sterilize
Do not reuse
Do not reuse

(4) Q1E-EP1234
Definition of symbol
The following symbol is also used for HITACHI Ultrasound Probes.
Location Symbol Definition
Probe connector
This instrument complies with
Directive 93/42/EEC relating to
Medical Device and Directive
2011/65/EU relating to RoHS
Probe connector IPX7 IPX7 mark
See section 1.6.
Probe connector Type BF APPLIED PART
Probe connector General warning sign
Probe connector Warning; dangerous voltage
Probe connector Caution; Biohazard
Probe connector
Follow the instruction manual to
operate this instrument. If not
avoided, may result in injury, property
damage, or the equipment trouble.
Probe connector STERRAD sterilization compatibility
mark
Probe connector
Upper Limit of Temperature;
The probes that are applicable to
Ethylene Oxide Gas Sterilization use
symbol of “Upper Limit of Temperature:
55 degrees”.
Probe connector
Do not waste the instrument as general
waste. Comply with a local regulation.
Probe connector Rx Only
By prescription only. U.S. Federal Law
restricts this device to sale on order
of a physician only.

(5) Q1E-EP1234
CONTENTS
Page
1. Introduction ··········································· 1
1.1 Features············································ 1
1.2 Principles of Operation····························· 1
1.3 Intended Use········································ 2
1.4 Compositions········································ 2
1.5 Accessories (Option)································ 2
1.6 External View······································· 3
2. Inspection before Use ·································· 4
2.1 Inspection for Appropriate Connection··············· 4
2.2 Inspection for Material Surface····················· 4
3. Operation Procedure ···································· 5
4. Option of EUP-O732T ···································· 7
4.1 Magnetic Sensor (EZU-RV2S)·························· 7
4.2 Magnetic Sensor (EZU-RV3S)·························· 9
5. Cleaning and Disinfection and Sterilization ············ 12
5.1 Point of use (Pre-cleaning)························· 15
5.2 Containment and transportation······················ 15
5.3 Manual Cleaning and disinfection···················· 15
5.4 Drying·············································· 18
5.5 Inspection·········································· 18
5.6 Packaging··········································· 18
5.7 Sterilization······································· 18
5.8 Storage············································· 20
6. Maintenance and Safety Inspection ······················ 21
6.1 Daily Inspection···································· 21
6.2 Store··············································· 21
7. Safety Precautions ····································· 22
8. Specifications ········································· 24
8.1 Probe··············································· 24
8.2 Supplier’s List of the Probe························ 25
8.3 Supplier’s List of the Magnetic Sensor
Attachment and the Spacer for EZU-RV2S·············· 25
9. Disposal of the Probe ·································· 26

-1- Q1E-EP1234
WARNING
WARNING
1. Introduction
1.1 Features
The Intraoperative Probe EUP-O732T is a probe for Convex Array
electronic scanning.
The acoustic output of this probe when connected to ultrasound
scanner was measured according to IEC 60601-2-37 standard. The
table of measured acoustic output data is contained in the
operation manual of each ultrasound scanner. This probe is
categorized in class IIa according to Directive 93/42/EEC.
According to IEC 60601-1 the probe is classified as type BF.
Never use the probe for following applications.
Direct contact to the heart.
Biopsy to the heart.
Never use the probe with (HF) surgical equipment.
1.2 Principles of Operation
This probe and the ultrasound diagnostic scanner enable image
diagnosis using ultrasonic waves. This system operates under the
principles described below.
1) When an electric pulse signal is applied from the transmitter
to the transducer of the probe, the transducer converts electric
signals into mechanical vibration energy for emitting
pulse-shaped ultrasonic waves into the body part, liquid or
other medium contacting the transducer.
2) The emitted ultrasonic waves are reflected by boundaries with
different acoustic characteristics (acoustic impedance)
within the body.
3) The transducer is also used to receive reflected ultrasonic
waves. The transducer vibrates mechanically due to the received
ultrasonic waves and converts mechanical vibrations into
electric energy. Electric signals are converted to shades of
brightness by brightness modulation to obtain an image.

-2- Q1E-EP1234
1.3 Intended Use
The Intraoperative Probe EUP-O732T is designed for observation
and diagnosis of the following regions mainly by connecting with
the HITACHI ultrasound scanner.
Intraoperative
Never use the probe for following regions.
1) The heart (Do not contact directly.)
2) The eyeball
1.4 Compositions
Components of the EUP-O732T are as follows:
1) Probe······················ 1 piece
2) Instruction Manual········· 1 copy
Sterilization has not been made to the probe shipped from
the factory. Prior to use of the probe, be sure to clean,
disinfect and sterilize them following this instruction
manual.
1.5 Accessories (Option)
1) Magnetic Sensor Attachment
Magnetic sensor attachment is used for Real-time Virtual
Sonography (RVS). It is used to fix the magnetic sensor to the
probe.
2) Spacer for EZU-RV2S
Spacer for EZU-RV2S is the spacer needed to fix the magnetic sensor
(EZU-RV2S) to the Magnetic sensor attachment.
The Magnetic sensor attachment and the Spacer for EZU-RV2S
are not sterilized when shipped from the factory. Prior to
use, be sure to clean, disinfect and sterilize them.
WARNING
CAUTION
CAUTION

-3- Q1E-EP1234
1.6 External View
The external view of EUP-O732T is shown in Fig.1.
Immersible
part:
This part can be immersed
i
n disinfectant
solution and also can be cleaned by water.
Un-immersible
part:
This part should not be immersed in disinfectant
solution and also can not be cleaned by water.
Immersible part (IPX
7
)
C
able
Connector
Un-immersible
part
Fig.1 External View
Scan
head
Probe
housing
Applied part
Spacer for EZU-RV2S
(Option)
Magnetic sensor
attachment(Option)

-4- Q1E-EP1234
2. Inspection before Use
Prior to use, the probe and accessories must be carefully inspected
so that they are appropriate for use. If you find any damage, do
not use them and contact a service support immediately.
2.1 Inspection for Appropriate Connection
1) Confirm that the system is correctly operating. Please refer
to the instruction manual of the Main unit for the operation.
2) Confirm that any unauthorized devices or instruments such as
an unauthorized biopsy attachment is not attached or connected
to the probe.
2.2 Inspection for Material Surface
1) Visually inspect the surface of the probe head, housing and
cable for any crack, scratch or denaturalization.
2) Visually inspect the Magnetic sensor attachment and the Spacer
for EZU-RV2S for any crack, deformation or denaturalization.
Beware of the Spacer for EZU-RV2S since it is so small.
CAUTION
Inhaltsverzeichnis
Andere Hitachi Zubehör Handbücher

Hitachi
Hitachi HBU-2015 Bedienungsanleitung

Hitachi
Hitachi UST-9120 Bedienungsanleitung

Hitachi
Hitachi UJ 18DSL Bedienungsanleitung

Hitachi
Hitachi UST-547 Bedienungsanleitung

Hitachi
Hitachi THM-R2AE Bedienungsanleitung

Hitachi
Hitachi EUP-S50A Bedienungsanleitung

Hitachi
Hitachi Super drive series Bedienungsanleitung

Hitachi
Hitachi S21 Bedienungsanleitung

Hitachi
Hitachi SPX-RTH1 Bedienungsanleitung