Kegel8 V For Men Bedienungsanleitung


2
Symbols on the unit and case
Caution! (electrical output).
Follow operating instructions! Failure to do so could place
the patient or operator at risk.
Neuromuscular Stimulation (STIM) and EMG Triggered
Stimulation (ETS) should not be used by Patients tted with
demand style cardiac pacemakers. Please seek advice from
your health supervisor.
Patient’s shock protection type: BF (Body Floated) Equip-
ment. This equipment is not earthed but contains a battery
within an insulated unit.
Indicates the manufacturer’s catalogue number so that the
medical device can be identied.
Manufacturer's LOT/Batch number. Present it together
with SN number when you report a technical fault or
claim a warranty return.
Manufacturer's serial number of the unit. Present it
together with LOT number when you report a technical
fault or claim a warranty return.
Name and address of Manufacturer.
Date of manufacture.
Conformity indication with the essential health and
safety requirements set out in European Directives.
0088 - Notied Body identication (LRQA Ltd.), see
page 64.
This product should be kept dry.
IP20
on the unit
This is an indication for protection against ingress of water
and particulate matter. The mark IP20 on your unit means:
your unit is protected against solid foreign objects of 12.5mm
dia and greater. Not protected against water.
IP02
on the case
IP02 on the carrying case means: Protected from the ingress
of water droplets from a shower of rain.
Do not dispose in normal dustbin (see page 50 for the
disposal instructions).
REF

3
Welcome to Kegel8® V for Men
Dear Customer,
Thank you for choosing Kegel8®V for Men. You are now one step
closer to a healthier, stronger, and more sensitive pelvic area.
Kegel8 V for Men is a highly effective neuromuscular stimulator
(STIM). It works by producing electrical pulses that stimulate the
muscles and nerve bres in your pelvic area.
If you have never used a neuromuscular stimulator before, you
may be a little apprehensive. That’s quite normal - lots of people
are cautious at rst, but soon realise that there is nothing to worry
about. The slim probe, skin electrode and lead wires are designed
for ease of use, and while you will be able to feel the effects of the
electric pulses, they should not cause any discomfort. If you follow
the instructions in this User Guide, Kegel8 V for Men is a safe and
effective way of improving the condition of your pelvic area.
Indications for use:
●urinary incontinence
Also used for non-medical purposes to:
●maintenance of healthy bowel movement
●achieve more satisfying erection
●improve rectal sensation for enhanced pleasure
●improve pelvic strength
Customer Care We aim to provide our customers with
outstanding products and customer service. So if you have
comments or ideas regarding Kegel8 features, please get in touch
(see Warranty on page 64).

4
Warnings
Please pay careful attention to the following safety warnings:
●Do not insert lead wires into a mains power supply.
●Do not immerse unit into water or any other substance.
●The unit is not protected from the ingress of water droplets from
a shower of rain if used outside the carrying case.
●Type BF equipment, Continuous Operation.
●Do not use the Kegel8®control unit in the presence of a
ammable anaesthetic gas mixture and air or with Oxygen or
Nitrous Oxide.
●If using rechargeable 9 Volt PP3 Nickel Metal Hydride batteries,
be sure to use a CE approved battery charger. Never connect
the Kegel8 control unit directly to a battery charger or to any
other mains powered equipment. We advise not to use Ni-Cad
rechargeable batteries.
●Probes and skin electrodes are for single patient use only.
●Keep out of reach of children.
●Operation in close proximity (e.g. 1m) to a shortwave or
microwave therapy equipment may produce instability in the
stimulator output.
●Application of electrodes near the throax may increase the risk
of cardiac brilation
●This device can deliver current densities in excess of 2mA/cm2
when used at a high intensity with small electrodes.
●Simultaneous connection of a patient to a high frequency
surgical equipment may result in burns at the site of the
stimulator electrodes and possible damage to the stimulator.
●This unit must be used with the guidance of a Physiotherapist
or Doctor.
●No modication of this equipment is allowed!

5
Please read this Kegel8®V for Men User Guide
before using this stimulation device.
STIM (Kegel8) should not be used:
●By patients tted with a demand style cardiac pacemaker
unless so advised by their Doctor.
●By patients who have been diagnosed with, or are receiving
treatment for, pelvic cancer.
●By patients with undiagnosed pain conditions.
●By patients with undiagnosed skin, penile or rectal conditions.
●With patients who have diminished mental capacity or physical
competence who cannot handle the device properly.
●On anaesthetised or desensitised skin.
●When driving a vehicle or operating potentially dangerous
equipment.
When using Kegel8 V for Men:
●Use the Kegel8®V for Men only as directed.
●Do not immerse the control unit in water or any other liquid
●Keep control unit out of reach of children.
●Seek professional advice if you are unsure about using Kegel8
V for Men. Contact your distributor or healthcare professional.
●Only use CE approved skin electrodes, lead wires and probes.
Do not place skin electrodes:
●Over carotid sinus nerves, larynx or trachea
●Inside mouth
●Over the area of the heart unless so advised by your doctor.
●Do not apply stimulation across or through the head, directly
on the eyes, covering the mouth, on the front of the neck
(especially the carotid sinus) or via electrodes placed on the
chest and upper back or crossing over the heart
●Near the thorax may increase the risk of cardiac brillation.
Contra Indications & Precautions

6
Contents
Symbols on the unit 2
Welcome to Kegel8® V for Men 3
Indications for Use 3
Warnings 4
Contra Indications & Precautions 5
Kegel8®V for MenUser Instructions 7
Important Safety Information 8
Kegel8® V for Men Check List 10
Electrodes Types 10
Getting to Know Your Kegel8® V for Men 11
How Your Kegel8® V for Men Works 18
What to Expect from Your Kegel8® V for Men 19
Setting Up Your Kegel8® V for Men 20
Selecting a Kegel8® Attachment 21
Inserting a Kegel8® Slim Probe 22
Using a Kegel8® Skin Electrode 24
Start a Kegel8 Training Programme 26
Sensitivity test 28
Urinary incontinence 1 30
Urinary incontinence 2 32
Maintenance of healthy bowel movement 34
Achieving more satisfying erection 36
Urinary incontinence 3 38
Pelvic Floor Strength 40
Pelvic Floor Max Strength 42
Creating a Custom Programme 44
Kegel8® Exercise Tips 48
Taking Care of your Kegel8® 50
Frequently Asked Questions 53
Technical Specications 57
STIM Specications 58
Electromagnetic Compatibility and Interference (EMC) 59
References and Contact Details 63
Warranty 64
Clinical References 65

7
Kegel8®
V for Men
User Instructions

8
Important Safety Information
Who Can Use Kegel8 V for Men®?
Kegel8 V for Men is designed for use by men. When used correctly,
it provides a safe and effective way of strengthening pelvic
muscles, tissue and bres. However, as Kegel8 V for Men works
by using electrical pulses, there are some people that should seek
medical advice before using the device.
Where Can You Use your Kegel8 V for Men?
Kegel8 V for Men is intended for home use, but you can use it in
medical centres and hospitals too. Before you use your Kegel8,
please make sure:
●The room does not contain ammable gases. This is especially
important in hospitals and medical centres as Kegel8’s electrical
pulses can ignite ammable gases, such as those used in
anaesthetics. If in doubt, ask for assistance.
●You are not positioned close (1-2m) to a shortwave or
microwave therapy device, microwave oven, mobile telephone,
wireless device, or radio. These devices can cause instability in
your Kegel8 control unit.
Using your Kegel8 V for Men Safely
●Only use the Kegel8 V for Men on your perineal and rectal
regions as described in this user guide. Never use Kegel8 on
other parts of your body.
●To ensure your personal hygiene, do not allow other people to
use your probe. Always clean your probes thoroughly after use
(see Kegel8 Probe Care on page 50).
●Only use CE approved probes, skin electrodes and lead
wires with your Kegel8. All ofcial Kegel8 accessories are CE
approved.

9
●To power your Kegel8 control unit, use a 9v battery. Never
connect your Kegel8 control unit, skin electrodes, lead wires
or probes to a mains power supply as this could result in
electrocution and death.
●Do not immerse your Kegel8 control unit or its cables or probes
in water, uids or any other substance.
●Please be aware that Kegel8 is Type BF equipment (provides
some protection against electric shock).

10
When you open the Kegel8® V for Men packaging, you should nd:
●1 ×Kegel8®V for Men unit
●1 ×9 Volt Battery
●2 ×Kegel8®Lead Wires
●1 ×Kegel8® Slim Probe (for rectal use)
●4 ×Kegel8® Skin Electrodes
●1 ×KE Gel® for Men
●1 ×Kegel8® V for Men User Guide
●1 ×Kegel8® Product Guarantee Registration Form.
In the unlikely event that any of the items above are missing, please
contact Customer Service (see page 66).
Kegel8®V for Men Check List
Electrodes Types
Shape Code Description
VS.4040 40 x 40 mm, square (* max 53mA)
VS.5050 50 x 50 mm, square
(recommended for general use)
VS.9040 90 x 40 mm, rectangular
VS.9050 90 x 50 mm, rectangular
VS.10050 100 x 50 mm, rectangular
VS.30 30 mm diameter, round
(* max 46mA)
VS.50 50 mm diameter, round
* IMPORTANT : Don’t use VS 4040 at more than 53mA and
VS3030 at more than 46 mA.
Inhaltsverzeichnis
Andere Kegel8 Fitnessgeräte Handbücher