Mecmesin MultiTest-i Series Bedienungsanleitung

Assembly and Installation of
MultiTest–i and Vortex-i
System Test Stands
431-393-07
December 2018

ii Mecmesin Assembly and Installation of –i System Test Stands
Important
It is essential that you familiarise yourself with the contents of this Assembly and
Installation Manual, and the separate Guide to Safe Use of Mains Powered Test Systems
before attempting to operate your MultiTest-i or Vortex-i Test System.
Scope
This reference manual covers the following products and their derivatives: (example)
Force testing stands
Single-column stands MultiTest 0.5-i
MultiTest 1-i
MultiTest 2.5-i
MultiTest 5-i
Twin-column stands MultiTest 10-i
MultiTest 25-i
MultiTest 50-i
Torque testing stand
Vortex–i Torque cell capacities:
0.3, 1.5, 3, 6, 10 Nm
2016 © Mecmesin Ltd, supplied with Mecmesin test systems and not for redistribution
Part no. 431-393-07

Assembly and Installation of –i System Test Stands Mecmesin iii
Contents
1. Items Supplied with the Test Stand 1
1.1 MultiTest-i test stand 1
1.2 Vortex-i test stand 1
2. Installation 2
2.1 Unpacking the stand 2
2.2 Lifting the test stand 2
2.3 Locating the stand 2
2.4 Mains power supply 2
3. MultiTest-i Assembly and Installation 3
3.1 Bolting the test stand to the work surface 3
3.2 Fitting the feet to the stand 4
3.3 Fitting the loadcell to the crosshead 4
3.4 Swapping loadcells 6
3.5 Attaching grips and fixtures 6
3.6 Setting the limit stops 6
3.6.1 Limits after connection to the PC, with the software running 8
4. Vortex-i Assembly and Installation 9
4.1 Fitting the crosshead to the Vortex-i 9
4.2 Connecting the intelligent torque cell 9
4.3 Swapping intelligent torque cells 10
5. Connecting the PC (MultiTest-i and Vortex-i) 11
5.1 Connecting the power lead and USB lead 11
5.2 Emergency stop button 11
5.3 Jog buttons 12
6. Installing Emperor Software on Your PC 13
6.1 Minimum system requirements 13
6.2 Access to data folders 13
6.3 Start Emperor 14
System Specifications 15
EC Declaration of Conformance 19

iv Mecmesin Assembly and Installation of –i System Test Stands

Assembly and Installation of –i System Test Stands Mecmesin 1
1. Items Supplied with the Test Stand
1.1 MultiTest-i test stand
1. Test stand
2. Four rubber feet, four attachment screws
3. Four base anchoring brackets, (if applicable)
4. Adjustable handle lever to secure loadcell to crosshead
5. At least one Intelligent Loadcell (ILC)
6. CD with Emperor™ software
7. Mains and data cables
8. Manual: Guide to Safe Use of Mecmesin Mains-Powered Test Stands
1.2 Vortex-i test stand
1. Test stand
2. Intelligent Torque Cell (ITC), as an integral part of the crosshead
3. CD with Emperor™ software
4. Mains and data cables
5. Manual: Guide to Safe Use of Mecmesin Mains-Powered Test Stands
The test stand provides a stable platform with a moving crosshead (force testing) or
rotating platen (torque testing). The Intelligent Loadcell measures the force developed,
and the Intelligent Torque cell measures the torque applied. A desktop PC or laptop
computer is used to run the Emperor™ software that provides the control of the moving
part of the stand, and records, stores and displays the information provided by the
measuring cell. Emperor™ is used to create the program for the test, and can make
calculations on the received data. The results of the calculations and the raw data can be
printed out or saved in files, and may be exported for remote, secure storage, or may be
used for further analysis by other software applications.
The stand communicates with the controlling computer via a USB or serial connection. If
other connections are available, then peripheral devices such as printers and network
connections can be added.

2 Mecmesin Assembly and Installation of –i System Test Stands
2. Installation
2.1 Unpacking the stand
When you first receive the stand please check that there is no obvious damage to the
packaging. If there is any sign that the packaging or the test stand itself has been
damaged, please contact Mecmesin or your authorised distributor immediately. Do not
use the stand until you have done so.
We strongly recommend that the packaging is kept, as this can be useful if the machine
needs to be returned for calibration.
Section 1 lists items that should be included with your test stand. Please contact
Mecmesin or your authorised distributor if any items are missing or damaged.
2.2 Lifting the test stand
The unpackaged weight of the test stand is given in the Specification table at the back of
this manual. Do not try to lift heavy loads unaided. Use suitable lifting equipment if
necessary.
2.3 Locating the stand
The test stand should be positioned on a suitable, level, stable work surface.
2.4 Mains power supply
Mecmesin –i stands can be used on 110–120 or 220–240 V ac 50-60 Hz supplies. The rear
fuse carrier will be set for your local requirement, but is reversible, so should you replace
a fuse, the correct local voltage must be selected. The voltage that is selected is the one
where the arrows meet (the power inlet is inverted for some test stands):
Fuse carrier Carrier removal

Assembly and Installation of –i System Test Stands Mecmesin 3
3. MultiTest-i Assembly and Installation
3.1 Bolting the test stand to the work surface
In order to comply with European regulation and safe use of the equipment, single
column stands should be secured to the bench as follows:
Test stand Height (mm) Feet/fixing supplied Bolting recommended?
0.5-i 1710 Anchor brackets Yes
1-i 1510 Anchor brackets Yes
2.5-i 941 Rubber feet No
5-i 1082 Rubber feet /locating eyes Yes
The extended-length test stands MultiTest 0.5-i and MultiTest 1-i are supplied with base
anchoring brackets to allow the test stands to be bolted to a bench. Screw the anchoring
brackets to the four positions on the base plate of the MultiTest 0.5-i or 1-i using the M6
screws provided. Secure the test stand to the bench using suitable fastenings.
MultiTest 0.5-i and 1-i are supplied with anchoring brackets
For additional stability the MultiTest 5-i is fitted with two ‘locating eyes’ on the base of
the MultiTest 5-i to allow the stand to be bolted to a bench.
Mecmesin twin-column force testing stands and Vortex torque testing stands require no
further stability fixing than a flat, secure and stable working surface.

4 Mecmesin Assembly and Installation of –i System Test Stands
Locating eyes on MultiTest 5-i
3.2 Fitting the feet to the stand
Fitting rubber feet to the base of the test stand
The MultiTest 2.5-i and MultiTest 5-i are supplied with rubber feet. Support the stand and
fit the four rubber feet to the base of the stand.
3.3 Fitting the loadcell to the crosshead
Single-column stands
Ensure the stand is switched off.
Screw the adjustable handle lever with the red button into the crosshead dovetail. This
lever is designed to tighten without full rotation. Hold the button on the top to raise and
disengage the lever, and turn it away from the direction required. Release and turn, and
repeat as required.
Do not tighten the lever without a loadcell attached, or the dovetail can become
distorted.

Assembly and Installation of –i System Test Stands Mecmesin 5
Inserting the adjustable handle lever
On single-column test stands, slide the loadcell (ILC) sideways onto the dovetail bracket at
the front of the crosshead. The threaded stud must always be on the underside. Secure the
loadcell using the handle lever. Some lateral adjustment is available if needed to align the
ILC and a sample. Slacken the lever to make the adjustment, and then re-tighten.
Note: There is some additional adjustment available by moving the anvil plate. Use a
3 mm Allen key to slacken the four retaining screws, reposition the anvil plate and re-
tighten the screws.
Align the electrical connector of the ILC with the socket on the test stand. Gently push the
connector to locate, then tighten the knurled locking ring in a clockwise direction.
Slide the loadcell onto the dovetail … tighten securely … connect the loadcell
Twin-column stands
The ILC is attached to a twin-column test stand using a cap-head bolt passed through the
central hole in the moving crosshead, and secured using the Allen key provided.
Attaching a loadcell on a twin-column system

6 Mecmesin Assembly and Installation of –i System Test Stands
3.4 Swapping loadcells
You can swap loadcells by simply disconnecting one cell and fitting another. First, return
to the Logon Screen and switch off the stand before unplugging the loadcell. When the
new loadcell has been connected, switch the stand on again, and log on again,. After a
few seconds the new loadcell will be automatically recognised. Emperor will read in the
new cell’s range, serial number and calibration status.
3.5 Attaching grips and fixtures
Grips and other holding fixtures are often paired, with one being attached to the anvil
plate, and the other to the underside of the loadcell. Some fixtures have the QC Quick-
interchange system which allows for very rapid changing of the holding accessory. With
the Quick-interchange system, the grip is attached to a mounting using an 8 mm diameter
pin. The grip can be fitted and removed without the need for additional tools. Be sure to
fit the locking spring to secure the accessory in place.
Attaching a grip with the QC Quick Interchange fittings
Take extra care when fitting or removing heavy grips to the underside of the loadcell.
Support the accessory while the securing device is removed so that it does not fall. Before
fitting a sample, check that both grips and plates are secure.
3.6 Setting the limit stops
Limit stops are provided to prevent damage to the loadcell and grips. A dual safety
system provides protection with the first stage being software controlled, followed, if
necessary, by a second hardware limit that removes power from the motor. Upper and
lower limit stops can be set to restrict movement of the crosshead.
Limit stops should be adjusted after grips or holding accessories have been fitted so that
the limit positions will prevent unwanted contact between moving and static parts.
To set either limit stop, slacken the knurled knob by turning it anti-clockwise, then slide
the stop to the desired position and re-tighten the knob again.
Dieses Handbuch passt für folgende Modelle
7
Inhaltsverzeichnis
Andere Mecmesin Laborausrüstung Handbücher
Beliebte Laborausrüstung Handbücher anderer Marken

Agilent Technologies
Agilent Technologies 5800 ICP-OES Bedienungsanleitung

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Bedienungsanleitung

NI
NI PXI-5422 Handbuch

Collomix
Collomix Aqix Bedienungsanleitung

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Bedienungsanleitung

Ocean Insight
Ocean Insight FLAME-NIR+ Bedienungsanleitung

Parker
Parker ALIGN-MG-NA Bedienungsanleitung

BD
BD 644787 Bedienungsanleitung

DENTAURUM
DENTAURUM Compact Megaplus Bedienungsanleitung

Biuged Laboratory Instruments
Biuged Laboratory Instruments BGD 626 Bedienungsanleitung

VWR
VWR SAS Super IAQ Bedienungsanleitung

illumina
illumina MiSeqDx Bedienungsanleitung