2
PORMDXBI00SE 026607 140316
OVERVIEW
The instrument is a portable digital manometer which has been designed for hydraulic systems.
As well as all the standard manometer functions, instrument also features a digital auto-zero function, hold
function, six different measurement units, overflow and underflow alarms, minimum and peak alarms with
programmable thresholds, date and time display clock, programmable automatic switch off function.
The instrument's distinguishing characteristic is that it carries out UNI 7129-1: 2015 (new systems) and
UNI 11137: 2012 (existing systems) gas tight tests and prints out the results on an infrared ray printer. It also
features a Data Logger which records and prints instantaneous, minimum, mean and peak values at
programmable time intervals.
This instrument has very low energy consumption levels and is powered by six 1.5V AAA batteries.
TECHNICAL SPECIFICATIONS
Supply : 9V, 6x1,5V AAA alkaline batteries
Input: 10 mA
Measuring range: ± 130 hPa
Measurement units: Pa, hPa, kPa, PSI, mmHg, mmH2O
Accuracy : ±1% readout value P > 15 hPa
±0.1 hPa -15 hPa < P < 15 hPa
±1.5% readout value P < -15 hPa
Including linearity, repeatibility and hysteresis
Resolution: According to the unit and measuring range:
Sensor type: semiconductor
Overload: 750 hPa
Display: 2 x LCD 4 digits + sign.
Protection level: IP30
Operating temperature: 0°C .. +40°C
Storage temperature: -10°C .. +50°C
Humidity: 20% .. 80% RH non condensing
Auto switch off: Programmable from 1 to 30 mins
Case: Material: ABS V0 self-extinguishing
Colour: Dark blue Pantone 5534
Size: 87 x 162 x 41mm (L x A x P)
Weight: ~ 264 gr.
START UP
Preliminary operations
Before use please take the following steps:
Insert batteries (included) in the battery housing unit ensuring correct polarity. To remove battery housing
protection panel press gently downwards on the plastic lever (c in fig. 2).
Connect silicon tubes (supplied) to the instrument's input (b in fig. 1) and output (a in fig. 1).
Battery check and replacement
If battery tension goes below the 6.5V threshold the low battery icon ' ' will appear in the display's top
righthand corner. To replace batteries follow the instructions in the paragraph above.
Unit Interval Resolution
Pa ±9999 1
hPa
±99.99 0.01
elsewhere 0.1
kPa
±9.999 0.001
elsewhere 0.01
PSI ±1.885 0.001
mmHg
±9.999 0.008
elsewhere 0.01
mmH2O
±999.9 0.1
elsewhere 1